EPIQ Elite Diagnostic Ultrasound System 795234 (UDI 00884838124912)Recalls & reported problems

Philips Ultrasound LLC · EPIQ Elite Diagnostic Ultrasound System VM 12.0. · Class 2

All models of EPIQ Elite Diagnostic Ultrasound System (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00884838124912), received between Jul 16, 2025 and Mar 23, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 7, 2026Class IIOpen, Classified

    Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.

    Z-3249-2026 · Root cause: Software Design Change · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fracture1
  • Unintended Application Program Shut Down1
  • Unintended System Motion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 23, 2026Unintended System Motion

    IT WAS REPORTED THE SWIVEL MECHANISM OF THE EPIQ ELITE ULTRASOUND SYSTEM'S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE SYSTEM WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OF USER WAS HARMED. THE SERVICE ENGINEER RESEATED THE ARTICULATION ARM BUSHING TO REPAIR THE SYSTEM. THE ENGINEER TEAM DETERMINED THE FAILURE WAS CAUSED BY A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24663488 ↗
  • MalfunctionDec 24, 2025Fracture; Fracture

    IT WAS REPORTED THE C10-3V TRANSDUCER HAD A HOLE IN THE LENS WITH EXPOSED ELECTRONICS. THIS WAS ASSOCIATED WITH AN EPIQ ELITE ULTRASOUND SYSTEM. THE ISSUE WAS FOUND OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED. INFORMATION WAS PROVIDED TO THE CUSTOMER FOR A REPLACEMENT PART TO ADDRESS THE CUSTOMER'S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23888228 ↗
  • MalfunctionJul 16, 2025Unintended Application Program Shut Down; Unintended Application Program Shut Down

    IT WAS REPORTED THAT THE EPIQ ELITE ULTRASOUND SYSTEM WAS NOT AVAILABLE DURING A CRITICAL OB (OBSTETRICS) EXAMINATION. THE SYSTEM DISPLAYED AN ERROR CODE THREE TIMES INDICATING RESTART OR SHUT DOWN WAS NEEDED AFTER ACQUIRING CINEMATIC LOOPS. THE PROCEDURE WAS ABLE TO BE COMPLETED AFTER MULTIPLE REBOOTS. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE PHILIPS SERVICE ENGINEER ADVISED THE CUSTOMER TO WAIT 6-10 SECONDS AFTER ACQUIRING THE CINEMATIC LOOP BEFORE ENTERING REVIEW, AS A WORKAROUND, TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERNS. A THOROUGH INVESTIGATION WAS PERFORMED TO IDENTIFY THE CAUSE OF THE REPORTED ISSUE. IT WAS DETERMINED THAT THE REPORTED PROBLEM WAS SOFTWARE RELATED. ANY ACTIONS NEEDED TO RESOLVE THIS ISSUE WILL BE CONSIDERED FOR DEVELOPMENT AND IMPLEMENTED IN FUTURE SOFTWARE UPDATES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22516814 ↗

Other System, Imaging, Pulsed Doppler, Ultrasonic devices

All System, Imaging, Pulsed Doppler, Ultrasonic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page