EPIQ Elite Diagnostic Ultrasound System 795234 (UDI 00884838124912)Recalls & reported problems
Philips Ultrasound LLC · EPIQ Elite Diagnostic Ultrasound System VM 12.0. · Class 2
All models of EPIQ Elite Diagnostic Ultrasound System (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00884838124912), received between Jul 16, 2025 and Mar 23, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 7, 2026Class IIOpen, Classified
Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.
Z-3249-2026 · Root cause: Software Design Change · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture1
- Unintended Application Program Shut Down1
- Unintended System Motion1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 23, 2026Unintended System Motion
IT WAS REPORTED THE SWIVEL MECHANISM OF THE EPIQ ELITE ULTRASOUND SYSTEM'S CONTROL PANEL DID NOT LOCK PROPERLY WHILE TRANSPORTING THE SYSTEM WITHIN THE USER FACILITY. THE FAILURE OCCURRED OUTSIDE OF CLINICAL USE AND NO PATIENT OF USER WAS HARMED. THE SERVICE ENGINEER RESEATED THE ARTICULATION ARM BUSHING TO REPAIR THE SYSTEM. THE ENGINEER TEAM DETERMINED THE FAILURE WAS CAUSED BY A HARDWARE ISSUE IN THE ARTICULATING ARM LATCHING MECHANISM PREVENTING THE LOCKING SOLENOID FROM FULLY ENGAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24663488 ↗MalfunctionDec 24, 2025Fracture; Fracture
IT WAS REPORTED THE C10-3V TRANSDUCER HAD A HOLE IN THE LENS WITH EXPOSED ELECTRONICS. THIS WAS ASSOCIATED WITH AN EPIQ ELITE ULTRASOUND SYSTEM. THE ISSUE WAS FOUND OUTSIDE OF CLINICAL USE AND NO PATIENT OR USER WAS HARMED. INFORMATION WAS PROVIDED TO THE CUSTOMER FOR A REPLACEMENT PART TO ADDRESS THE CUSTOMER'S IMMEDIATE CONCERNS. PHILIPS HAS STARTED AN INVESTIGATION, AND UPON COMPLETION, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23888228 ↗MalfunctionJul 16, 2025Unintended Application Program Shut Down; Unintended Application Program Shut Down
IT WAS REPORTED THAT THE EPIQ ELITE ULTRASOUND SYSTEM WAS NOT AVAILABLE DURING A CRITICAL OB (OBSTETRICS) EXAMINATION. THE SYSTEM DISPLAYED AN ERROR CODE THREE TIMES INDICATING RESTART OR SHUT DOWN WAS NEEDED AFTER ACQUIRING CINEMATIC LOOPS. THE PROCEDURE WAS ABLE TO BE COMPLETED AFTER MULTIPLE REBOOTS. THERE WAS NO PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. THE PHILIPS SERVICE ENGINEER ADVISED THE CUSTOMER TO WAIT 6-10 SECONDS AFTER ACQUIRING THE CINEMATIC LOOP BEFORE ENTERING REVIEW, AS A WORKAROUND, TO ADDRESS THE CUSTOMER¿S IMMEDIATE CONCERNS. A THOROUGH INVESTIGATION WAS PERFORMED TO IDENTIFY THE CAUSE OF THE REPORTED ISSUE. IT WAS DETERMINED THAT THE REPORTED PROBLEM WAS SOFTWARE RELATED. ANY ACTIONS NEEDED TO RESOLVE THIS ISSUE WILL BE CONSIDERED FOR DEVELOPMENT AND IMPLEMENTED IN FUTURE SOFTWARE UPDATES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22516814 ↗
Other System, Imaging, Pulsed Doppler, Ultrasonic devices
- Transducer C5-1 989605412042Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic11 recalls0 reports
- Transducer X5-1 989605400802Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic10 recalls0 reports
- EPIQ 7 Diagnostic Ultrasound System 989605386721Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls2 deaths146 reports
- Transducer S7-3t 989605406772Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic5 recalls1 death18 reports
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE OLYMPUS GF TYPE UCT180Olympus Medical Systems Corp. · System, imaging, pulsed doppler, ultrasonic199 reports
- Transducer S7-3t 989605439232Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls4 reports
- Transducer 3D9-3v 989605395622Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic3 recalls1 report
- ACUSON Redwood Diagnostic Ultrasound System 11503314Siemens Medical Solutions USA, Inc. · System, imaging, pulsed doppler, ultrasonic3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page