Skip to content

Ingenia Ingenia Elition X (REF 782198)Recalls & reported problems

Philips Medical Systems Nederland B.V. · Ingenia Elition X · Class 2

All models of Ingenia (54)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 28, 2026Class IIOpen, Classified

    When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

    Z-3311-2026 · Root cause: Software design · Recall details · FDA record ↗

Other System, Nuclear Magnetic Resonance Imaging devices

All System, Nuclear Magnetic Resonance Imaging devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page