LIFEPAK® 20 and LIFEPAK 20e defibrillator/monitors 99507 (REF 99507-000010)Recalls & reported problems
Physio-Control, Inc. · LP20E-CONFIGURED, AHA 2005,PCNG, SPO2,ENGLISH · Class 3
All models of LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors (23)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885074515281), received between Sep 19, 2022 and Sep 19, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Electrical /Electronic Property Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 19, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER CONTACTED STRYKER TO REPORT A NON-CRITICAL ISSUE WITH THEIR DEVICE. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED EVENT. UPON EVALUATION OF THE CUSTOMER'S DEVICE, STRYKER OBSERVED THAT THE DEVICE LOGGED A CRITICAL EVENT THAT IS ASSOCIATED WITH A DEVICE LOCK UP AND WAS INTERMITTENTLY BOOTING UP TO A WHITE DISPLAY. A WHITE DISPLAY CAN BE INDICATIVE OF A DEVICE LOCK UP CONDITION. AS A RESULT, DEFIBRILLATION THERAPY MAY BE UNAVAILABLE, IF IT WAS NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15452096 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page