DYONICS SYNOVATOR 72203523Recalls & reported problems
Smith & Nephew, Inc. · DYONICS SYNOVATOR PLATINUM 4.5 MM BLADES · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00885554027839), received between Jun 12, 2020 and Sep 11, 2022 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Flaked” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Flaked3
- Corroded2
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Foreign Body In Patient1
- Injury1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjurySep 11, 2022Flaked; Flaked
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY, THE SURGEON OBSERVED THAT DEBRIS HAD COME FROM THE WORKING END OF THE SYNOVATOR SHAVER BLADE INTO THE JOINT; IT WAS STATED THE SHAVER BLADE WAS "UNDER NO PRESSURE". THE DEBRIS WAS REMOVED BY IRRIGATION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NON-SIGNIFICANT SURGICAL DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 15398014 ↗MalfunctionJun 13, 2021Corroded; Corroded
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY PROCEDURE, THERE WAS RUST ON THE TIP OF THE SYNOVATOR PLATINUM BLADE. NO SIGNIFICANT DELAY AND PROCEDURE WAS COMPLETED WITH A BACKUP DEVICE WITH NO FURTHER COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11989120 ↗InjuryJan 18, 2021Flaked; Flaked
IT WAS REPORTED THAT DURING SHOULDER SCOPE, THERE WAS SHEDDING METAL IN THE PATIENT, IT POP OUT FROM DISTAL TIP OF SHAVER. THE MALFUNCTION WAS SOLVED WITH NO SIGNIFICANT DELAYS OR PATIENT HARM USING A BACK UP DEVICE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11189419 ↗InjuryJan 18, 2021Flaked; Flaked
IT WAS REPORTED THAT DURING SHOULDER SCOPE, THERE WAS SHEDDING METAL IN THE PATIENT, IT POP OUT FROM DISTAL TIP OF SHAVER. THE MALFUNCTION WAS SOLVED WITH NO SIGNIFICANT DELAYS OR PATIENT HARM USING A BACK UP DEVICE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11189418 ↗MalfunctionJan 6, 2021Corroded; Corroded
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE 4.5MM SYNOVATOR PLATINUM SERIES HAD RUST AT THE TIP OF THE BLADE. THE PRODUCT WAS NOT USED INSIDE THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DELAY OF 30 MINUTES OR LESS USING A SMITH AND NEPHEW BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11127825 ↗InjuryJun 12, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT DURING KNEE PROCEDURE, THE SYNOVATOR PLATINUM BLADE WAS NOT WORKING, A BLACK RESIDUE WAS LEFT BEHIND WHILE USING. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO DELAY. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Injury
MAUDE report 10149278 ↗
FDA clearance and classification
The FDA regulates product code HAB (saw, powered, and accessories) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the DYONICS SYNOVATOR 72203523
Has the DYONICS SYNOVATOR 72203523 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00885554027839), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the DYONICS SYNOVATOR 72203523?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Jun 12, 2020 and Sep 11, 2022. In those reports, the most common specific device problems are “Flaked” (3), “Corroded” (2) and “Fluid/Blood Leak” (1); the most common patient problems are “Foreign Body In Patient” (1) and “Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the DYONICS SYNOVATOR 72203523?
None of the reports are classified as deaths, 4 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the DYONICS SYNOVATOR 72203523 FDA approved?
Devices of this type (product code HAB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Smith & Nephew, Inc.
Who makes the DYONICS SYNOVATOR 72203523, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand DYONICS SYNOVATOR, model 72203523). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SAW, POWERED, AND ACCESSORIES devices
- DYONICS POWERMAX ELITE 72200616Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES4,288 reports
- NA 72200616SSmith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES486 reports
- NA 72201500Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES136 reports
- PLATINUM 72205321Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES117 reports
- Zimmer Biomet™ 89-8509-450-60Zimmer Surgical SA · Saw, Powered, And Accessories83 reports
- NA 72200617Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES77 reports
- NA 72202530Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES53 reports
- DYONICS INCISOR 72203013Smith & Nephew, Inc. · SAW, POWERED, AND ACCESSORIES46 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page