Skip to content

FAST-FIX 72203986Recalls & reported problems

Smith & Nephew, Inc. · TFCC FAST-FIX KIT INCLUDES FAST-FIX 360 DELIVERY DEVICE, DISPOSABLE SLOTTED CANNULA, AND KNOT PUSHER/SUTURE CUTTER · Class 2 · Implant

Adverse event reports
7
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00885554032154), received between Mar 26, 2018 and Aug 25, 2025 — classified as 4 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K132079); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Activation, Positioning or Separation Problem1
  • Break1
  • Failure to Fire1
  • Premature Activation1
  • Retraction Problem1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • Bone Fracture(s)1
  • Foreign Body In Patient1
  • Injury1
  • Loss of Range of Motion1
  • Muscular Rigidity1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 25, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE T2 OF THE FAST FIX DID NOT DEPLOYED WHEN THE TRIGGER WAS PUSHED. T1 WAS DEPLOYED WITH NO ISSUES. ALL ANCHORS WERE REMOVED USING FORCEPS THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SURGICAL DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22890206 ↗
  • InjuryDec 7, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER AN ARTHROSCOPY WITH TFCC FAST-FIX KIT, THE PATIENT HAD STIFFNESS OF RIGHT DIGITS 2-4 PIP JOINTS ON (B)(6) OF 2019 . THE EVENT WAS TREATED WITH MANIPULATION UNDER ANESTHESIA. THE OUTCOME OF THE PATIENT IS RESOLVED.

    Patient problems: Loss of Range of Motion; Muscular Rigidity

    MAUDE report 12951832 ↗
  • InjurySep 9, 2021Break; Break

    IT WAS REPORTED THAT, AFTER ARTHROSCOPY, A TFCC FAST FIX KIT HAS A FIBER WIRE GRINDING OVER DISTAL ULNAR AND RIGHT WRIST DURING MOTION. PROCEDURE HAD BEEN COMPLETED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: Foreign Body In Patient

    MAUDE report 12451573 ↗
  • MalfunctionMar 22, 2019Premature Activation; Retraction Problem

    IT WAS REPORTED THAT DURING A WRIST SURGERY AFTER T1 WAS DEPLOYED, THE INTERNAL ROD DID NOT RETRACT, CAUSING A SIMULTANEOUS DEPLOYMENT. AN OPEN SURGERY WAS PERFORMED TO RETRIEVE T2 AND THE SUTURE MATERIAL WAS REMOVED FROM THE JOINT SPACE. NO SIGNIFICANT DELAY WAS REPORTED. A BACK UP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8445720 ↗
  • InjuryFeb 28, 2019Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT RE-INJURED THE WRIST AND HAD A SURGICAL INTERVENTION.

    Patient problems: Injury

    MAUDE report 8377473 ↗
  • InjuryFeb 14, 2019Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT HAD A LEFT ELBOW DISLOCATION THAT WAS TREATED WITH ULNAR NERVE RELEASE. PATIENT WAS RECOVERED.

    Patient problems: Bone Fracture(s)

    MAUDE report 8340748 ↗
  • MalfunctionMar 26, 2018Activation, Positioning or Separation Problem

    IT WAS REPORTED THAT THE SECOND ANCHOR PRE-DEPLOYED DURING THE SURGERY. A BACKUP DEVICE WAS USED TO COMPLETE THE SURGERY. NO SIGNIFICANT DELAY OR PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7372178 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K132079. One submission often covers many models, so check the product and lot details in each record.

  • Z-1822-2015 · Apr 30, 2015 · Terminated
    ULTRA-FAST FIX Knot Pusher Suture Cutters Product Number: 72201537 Intended for Meniscal Repair.

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.