EVOS 71174910Recalls & reported problems
Smith & Nephew, Inc. · 2.0MM DRILL SHORT WITH AO QUICK CONNECT · Class 1
All models of EVOS (191)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00885556386798), received between May 21, 2019 and Apr 6, 2026 — classified as 4 injuries, 9 malfunctions.
The most frequently reported specific device problem is “Break” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break7
- Fracture6
- Physical Resistance/Sticking2
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Device Embedded In Tissue or Plaque2
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 6, 2026Break
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, ONE (1) 2.0MM DRILL SHORT W/AO QC BROKE WHILE FINISHING THE DRILLING. THERE WAS NO FRACTURE TO THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24789475 ↗MalfunctionNov 30, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A CLAVICLE INTERNAL FIXATION, THE 2.0MM DRILL SHORT W/AO QC BROKE IN THE 2.0/2.7MM DOUBLE-ENDED DRILL GUIDE AND COULD NOT SEPARATE. ALL PIECES WERE ACCOUNTED FOR. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18237271 ↗MalfunctionSep 13, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING A OSTEOSYNTHESIS OF THE TIBIA, WHEN THE DOCTOR WENT TO DRILL USING THE 2.0MM DRILL SHORT W/AO QC IT BROKE BECAUSE IT HIT THE NAIL. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. THE DOCTOR REMOVED THE TIP OF THE DRILL WITH KELLY FORCEPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17741920 ↗MalfunctionAug 28, 2023Fracture; Physical Resistance/Sticking; Fracture; Physical Resistance/Sticking
IT WAS REPORTED THAT, DURING A CLAVICLE ORIF, A 2.0MM DRILL SHORT BROKE INSIDE A 2.0/2.7MM DOUBLE-ENDED DRILL GUIDE AND COULD NOT BE REMOVED FROM THE MENTIONED GUIDE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO PIECES FELL INTO THE PATIENT, THE DRILL BROKE OFF CLEAN. IT WAS CONFIRMED THAT THE END OF THE DRILL BIT WAS STILL STICKING OUT OF THE OTHER SIDE OF THE GUIDE AND NO OTHER PIECES WERE MISSING. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17638415 ↗InjuryApr 12, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING AN OSTEOSYNTHESIS FOR MULTIPLE FRACTURES OF THE DISTAL HUMERUS OF A CHILD, THE 2.0MM DRILL SHORT W/AO QC BROKE OFF IN ONE OF THE MANY PERFORATIONS FOR PLACING THE SCREWS. IT WAS ATTEMPTED TO REMOVE THE BROKEN PIECE FROM THE BONE BY USING THE RESPECTIVE TECHNIQUE WITH THE USE OF A VARIABLE ANGLE DRILL GUIDE, BUT IT WAS NOT POSSIBLE, SO THE SURGEON DECIDED TO LEAVE IT THERE. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN S+N BACKUP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED AS A CONSEQUENCE OF THIS PROBLEM, THE PATIENT'S HEALTH CONDITION IS STABLE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 16727983 ↗MalfunctionDec 12, 2022Fracture; Fracture
IT WAS REPORTED THAT, DURING SURGERY, THE 2.0MM DRILL SHORT W/AO QC BECAME INCARCERATED INTO THE 2.0/2.7MM DOUBLE-ENDED DRILL GUIDE. THE DRILL WAS COMPLETED STUCK AND ATTEMPTS TO REMOVE RESULTED IN THE DRILL BREAKING AT BOTH ENDS. NOTHING FELL INTO THE PATIENT. THE DRILL GUIDE WAS NEW, IT HAS NEVER BEEN USED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15964459 ↗InjuryOct 26, 2022Fracture; Fracture
IT WAS REPORTED THAT, DURING AN OSTEOSYSNTESIS OF THE LEFT HUMERUS, A 2.0MM DRILL SHORT W/AO QC WAS USED WHICH BROKE IN THE PATIENT'S BONE WITHOUT BEING ABLE TO EXTRACT IT. IT IS UNKNOWN HOW SURGERY WAS COMPLETED AND IF THERE WAS A DELAY. THE PATIENT'S HEALTH WAS NOT AFFECTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 15674476 ↗MalfunctionMay 18, 2022Break; Separation Failure; Break; Separation Failure
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, A 2.0MM DRILL SHORT WITH AO QUICK CONNECT GOT STUCK IN THE 2.0/2.7MM DOUBLE-ENDED DRILL GUIDE, AND BROKE. NO PIECES FELL INSIDE OF PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14431858 ↗MalfunctionJan 24, 2022Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, WHEN PLACING THE 2.0MM DRILL SHORT W/AO QC IN THE 2.0 GUIDE, THE DRILL GOT STUCK IN THE GUIDE AND WHEN TRYING TO TAKE IT OUT, THE DRILL BROKE. INCIDENT OCCURRED OUTSIDE THE PATIENT. THERE WAS NO DELAY AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13339510 ↗InjuryJan 8, 2021Break; Break
IT WAS REPORTED THAT DURING A REVISION SURGERY THE 2.0MM DRILL SHORT W/AO QC BIT BROKE WHILE BEING USED TO DRILL A PILOT HOLE FOR A SCREW. THE INSTRUMENT BROKE INSIDE PATIENT. S&N BACKUP DEVICE AVAILABLE. NO SURGICAL DELAY OR PATIENT HARM REPORTED DUE THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11142853 ↗InjuryJul 17, 2020Break; Break
IT WAS REPORTED THAT DURING A REVISION SURGERY THE 2.0MM DRILL SHORT W/AO QC BROKE WHILE BEING USED INSIDE THE PATIENT TO DRILL A PILOT HOLE FOR A SCREW, ALL PIECES WERE RETRIEVED. REASON FOR REVISION: PATIENT DIDN'T HEAL AFTER HAVING COMPETITIVE DEVICES PUT IN. NO DELAY. PROCEDURE WAS FINISHED USING A S&N BACK UP DEVICE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10292072 ↗MalfunctionDec 6, 2019Break; Break
IT WAS REPORTED THAT DURING SURGERY, WHILE THE DOCTOR WAS DRILLING WITH THE 2.0 DRILL, THE DRILL GUIDE SNAPPED OFF AND WAS ON THE DRILL BIT (NO PARTS FELL INTO THE PATIENT). THIS DID NOT ADD ANY SUBSTANTIAL TIME TO THE CASE BECAUSE AN S&N BACK UP WAS AVAILABLE. THERE WAS NO HARM TO THE PATIENT OR STAFF.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9429276 ↗MalfunctionMay 21, 2019Break; Break
IT WAS REPORTED THAT 2.0 DRILL BIT BROKE INSIDE THE TISSUE PROTECTOR DURING SURGERY. NO DELAY REPORTED. UNKNOWN IF SNN BACKUP WAS USED OR REQUIRED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8628569 ↗
FDA clearance and classification
The FDA regulates product code HTW (bit, drill) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the EVOS 71174910
Has the EVOS 71174910 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00885556386798), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the EVOS 71174910?
The FDA's MAUDE database lists 13 adverse event reports for this device, received between May 21, 2019 and Apr 6, 2026. In those reports, the most common specific device problems are “Break” (7), “Fracture” (6) and “Physical Resistance/Sticking” (2); the most common patient problem is “Device Embedded In Tissue or Plaque” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the EVOS 71174910?
None of the reports are classified as deaths, 4 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the EVOS 71174910 FDA approved?
Devices of this type (product code HTW) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Smith & Nephew, Inc.
Who makes the EVOS 71174910, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand EVOS, model 71174910). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BIT, DRILL devices
- QUICKSET 2274-25-500Depuy Orthopaedics, Inc. · BIT, DRILL1,978 reports
- QUICKSET 2274-35-500Depuy Orthopaedics, Inc. · BIT, DRILL769 reports
- QUICKSET 2274-30-500Depuy Orthopaedics, Inc. · BIT, DRILL761 reports
- NA 03.037.022Synthes GmbH · BIT, DRILL335 reports
- NA 323.062Synthes GmbH · BIT, DRILL308 reports
- QUICKSET 2274-56-000Depuy (Ireland) · BIT, DRILL275 reports
- QUICKSET 2274-45-500Depuy Orthopaedics, Inc. · BIT, DRILL210 reports
- LCS 2366-83-000Depuy (Ireland) · BIT, DRILL204 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page