ULTRABUTTON 72205602Recalls & reported problems
Smith & Nephew, Inc. · ULTRABUTTON TIB ADJUSTABLE FIXATION DEVICE, LARGE · Class 2 · Implant
All models of ULTRABUTTON (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885556817223), received between Dec 2, 2024 and Feb 7, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Laceration(s)1
- Perforation1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 7, 2026Material Deformation
IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION, THE LOOP OF ONE (1) ULTRABUTTON TIB GOT ENLARGED WHILE CYCLING AFTER THE DEPLOYMENT. THE DOCTOR REMOVED THE BUTTON BY CUTTING THE LOOP AND PUT A NEW BUTTON WITH A SCREW. THE PROCEDURE WAS RESUMED, AFTER A DELAY GREATER THAN 30 MINUTES, USING A SIMILAR S+N BACK-UP PRODUCT. ADDITIONAL ANESTHESIA WAS NOT REQUIRED AS CONSEQUENCE OF THE SURGICAL PROLONGATION. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24286000 ↗MalfunctionMay 29, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE SUTURE OF THE ULTRABUTTON TIB ADJUSTABLE FIXATION DEVICE BROKE WHILE TENSIONING AFTER FIXING IT ON THE TIBIAL SIDE. THE PROCEDURE WAS COMPLETED WITH THE SAME FAULTY DEVICE AND USING AN ULTRATAPE, THERE WAS DELAY GREATER THAT 30 MINUTES. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22110387 ↗MalfunctionDec 2, 2024Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING A QUAD ST ACL RECONSTRUCTION SURGERY, THE LOOP OF THE ULTRABUTTON TIB MEDIUM ADJUSTABLE FIXATION DID NOT REDUCE ON THE FIRST ATTEMPT WITH THE DEVICE, THE SURGEON EXPERIENCED EXTREME STIFFNESS, MAKING IT DIFFICULT TO MOVE, RESULTING IN THE BUTTON TEARING OFF THE GRAFT, WHICH COMPROMISED THE ULTRABUTTON QUAD ADJUSTABLE FIXATION FEMORAL ATTACHMENT, IT WAS UNABLE TO BE USED ON THE REMAINING GRAFT. ON THE SECOND ATTEMPT, WITH THE SECOND ULTRABUTTON TIB MEDIUM ADJUSTABLE FIXATION, THE TIBIAL BUTTON WAS TESTED OUTSIDE THE KNEE JOINT, BUT IT WOULD NOT REDUCE DESPITE FULL-FORCE PULLING, DIFFERENT ANGLES, AND MULTIPLE REDUCTION ATTEMPTS. WHEN A THIRD DEVICE WAS OPENED, A LARGE ULTRABUTTON TIB ADJUSTABLE FIXATION DEVICE, IT WAS FOUND THAT THE BLUE SUTURE WENT THROUGH THE REDUCING SUTURES, THEREFORE, THE SURGEON REMOVED AND ADDED THEIR OWN SUTURE TO BE ABLE TO REDUCE THAT ONE. THE PROCEDURE WAS COMPLETED USING A QUAD BUTTON AND TIB BUTTON LARGE BACKUP DEVICES. THE SURGERY TOOK 6 AND A HALF HOURS, THEREFORE, THERE WAS A DELAY GREATER THAN 30 MINUTES. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Laceration(s); Perforation
MAUDE report 20824722 ↗
Other Fastener, fixation, nondegradable, soft tissue devices
- TRUESPAN ORTHOCORD 228151Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,686 reports
- TRUESPAN ORTHOCORD 228152Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,398 reports
- OMNISPAN ORTHOCORD 228142Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE932 reports
- FOOTPRINT 72202901Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE721 reports
- RIGIDLOOP 232447Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE633 reports
- Arthrex® AR-3636Arthrex, Inc. · Fastener, fixation, nondegradable, soft tissue624 reports
- OMNISPAN ORTHOCORD 228141Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE609 reports
- TRUESPAN ORTHOCORD 228162Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE460 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗