Skip to content

Skater™ All-Purpose and Nephrostomy Set 756608020Recalls & reported problems

Argon Medical Devices, Inc. · Skater™ All-Purpose and Nephrostomy Set – Locking 8Fr x 20cm · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00886333217779), received between Sep 16, 2024 and Sep 16, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Physical Resistance/Sticking1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 16, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    REPORTER INDICATED "DURING AN ULTRASOUND-GUIDED PUNCTURE OF THE GALLBLADDER USING A DRAIN, AN INCIDENT OCCURRED. THE DRAIN AND THE TROCAR COULD NOT BE SEPARATED, PREVENTING THE TROCAR FROM BEING REMOVED. THE EQUIPMENT WAS THEN REMOVED, THE EQUIPMENT WAS CHANGED AND THE PUNCTURE WAS PERFORMED. MEDICAL INTERVENTION WAS REQUIRED."

    Patient problems: Insufficient Information

    MAUDE report 20226162 ↗

Other Catheter, Nephrostomy, General & Plastic Surgery devices

All Catheter, Nephrostomy, General & Plastic Surgery devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page