Terumo® Advanced Perfusion System 1 5785Recalls & reported problems

Terumo Cardiovascular Systems Corporation · The air sensor is a non-invasive, ultrasonic sensor that can detect air bubbles in the perfusion circuit. This is the 1/4"x1/16" sensor. Includes cable and holder. · Class 2

All models of Terumo Advanced Perfusion System 1 (31)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00886799000045), received between Mar 26, 2024 and Oct 29, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Alarm System” (2 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 25, 2020Class IITerminated

    Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.

    Z-2960-2020 · Root cause: Employee error · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Alarm System2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 29, 2024Device Alarm System; Device Alarm System

    PER CLINICAL REVIEW: ON (B)(6) 2024, THE TEAM EXPERIENCED A PROBLEM WITH THEIR HEART LUNG MACHINE (HLM) DURING CARDIOPULMONARY BYPASS (CPB) WHEREBY AN AIR BUBBLE DETECTOR (ABD) ALARM OCCURRED EVEN WHEN IT WAS TURNED OFF. ACCORDING TO THE USER, WHEN THE AIR SENSOR WAS ATTACHED TO THE CEREBRAL PERFUSION LINE, THE AIR ALARM OCCURRED EVEN WHEN IT WAS OFF. IT OCCURRED WHEN THE CEREBRAL PERFUSION LINE WAS CONNECTED TO THE OPERATING FIELD CIRCUIT AND PRIMING WAS PERFORMED. THEY FINISHED THE CASE WITHOUT CHANGING OUT THE ABD. THERE WAS NO DELAY TO THE PROCEDURE, NO BLOOD LOSS, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20561918 ↗
  • MalfunctionMar 26, 2024Device Alarm System; Device Alarm System

    PER CLINICAL REVIEW: THE USER REPORTED AN ISSUE WITH THEIR HEART LUNG MACHINE (HLM) DURING CARDIOPULMONARY BYPASS (CPB). SPECIFICALLY, THE USER NOTED THAT THE AIR BUBBLE DETECTOR (ABD) TURNED OFF, BUT AN ALARM CONTINUED TO SOUND. THIS OCCURRED DURING DEEP HYPOTHERMIC CIRCULATORY ARREST (DHCA), WHILE THE USER WAS DELIVERING SELECTIVE CEREBRAL PERFUSION TO THE PATIENT. THERE WAS NO CASE DELAY, SURGERY WAS COMPLETED SUCCESSFULLY WITH NO PATIENT HARM AND BLOOD LOSS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18979322 ↗