Terumo Cardiovascular Systems Corporation · Roller Pump (4" Blue). The small roller pump is a peristaltic pump with a 4 inch (10 cm) raceway that can be used for arterial, cardioplegia, vent, and suction applications. A universal tub clamp mechanism accommodates a variety of tubing sizes, including dual tube sets. · Class 2

All models of Terumo Advanced Perfusion System 1 (31)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jan 10, 2020Class IITerminated

    These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.

    Z-1108-2020 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  2. Jan 10, 2020Class IITerminated

    These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.

    Z-1109-2020 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗