Terumo® Advanced Perfusion System 1 890500Recalls & reported problems
Terumo Cardiovascular Systems Corporation · Central Control Monitor (CCM) combines an embedded computer, a touch screen, and a Graphical User Interface (GUI) to provide a centralized display of system information and onscreen control of all system devices · Class 2
All models of Terumo Advanced Perfusion System 1 (31)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00886799001813), received between Apr 8, 2024 and Aug 28, 2026 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Output Problem” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Output Problem6
- Inadequate User Interface4
- Display or Visual Feedback Problem3
- Crack1
- Power Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
- Insufficient Information4
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 28, 2026Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50499058 ↗MalfunctionMar 30, 2026Inadequate User Interface
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE CENTRAL CONTROL MONITOR (CCM) TOUCHSCREEN HAD AN ISSUE WHEN ATTEMPTING TO ADJUST THE SLIDER FOR THE OCCLUDER. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24729788 ↗MalfunctionDec 12, 2025Output Problem; Output Problem
ADDITIONAL INFORMATION WAS RECEIVED THAT THE ISSUE WAS INTERMITTENT, OCCURRING FIRST DURING SET-UP THEN MULTIPLE TIMES DURING TROUBLESHOOTING. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23790531 ↗MalfunctionOct 28, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE CENTRAL CONTROL MONITOR (CCM) DISPLAY WOULD NOT POWER ON. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23409183 ↗MalfunctionAug 5, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING PRIMING OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE CENTRAL CONTROL MONITOR (CCM) WENT BLANK AND WAS NOT WORKING. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22723656 ↗MalfunctionJul 24, 2025Output Problem; Output Problem
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE CENTRAL CONTROL MONITOR (CCM) DISPLAYED A 'SYSTEM COMPUTER NEEDS SERVICE' MESSAGE. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22592222 ↗MalfunctionJul 16, 2025Output Problem; Output Problem
IT WAS REPORTED THAT THE CENTRAL CONTROL MONITOR (CCM) DISPLAYED A 'SYSTEM COMPUTER NEEDS SERVICE' MESSAGE. THERE WERE NO REPORTED CONSEQUENCES TO THE PATIENT. NO OTHER DETAILS REGARDING THE NATURE OF THIS EVENT WERE PROVIDED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 22521836 ↗MalfunctionMay 29, 2025Output Problem; Output Problem
IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR A NON-CLINICAL ACTIVITY, THE CENTRAL CONTROL MONITOR (CCM) DISPLAYED A 'SYSTEM COMPUTER NEEDS SERVICE' MESSAGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22107642 ↗MalfunctionMay 1, 2025Power Problem; Power Problem
THE FIELD SERVICE REPRESENTATIVE (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, THE CENTRAL CONTROL MONITOR (CCM) WOULD NOT POWER ON. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21946622 ↗MalfunctionDec 3, 2024Inadequate User Interface; Inadequate User Interface
ADDITIONAL INFORMATION WAS RECEIVED THAT AS A RESULT OF THE REPORTED ISSUE, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20836843 ↗MalfunctionOct 28, 2024Output Problem; Output Problem
IT WAS REPORTED THAT THE CENTRAL CONTROL MONITOR (CCM) DISPLAYED A 'SYSTEM COMPUTER NEEDS SERVICE' MESSAGE MULTIPLE TIMES. THERE WAS NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT. NO OTHER DETAILS REGARDING THE NATURE OF THIS EVENT WERE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20549099 ↗MalfunctionSep 18, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT THE CENTRAL CONTROL MONITOR (CCM) WAS BLANK UPON BOOT UP. THERE WERE NO REPORTED ADVERSE CONSEQUENCES TO THE PATIENT. NO OTHER DETAILS REGARDING THE NATURE OF THIS EVENT WERE PROVIDED.
Patient problems: Insufficient Information
MAUDE report 20251516 ↗MalfunctionJun 14, 2024Crack; Inadequate User Interface; Crack; Inadequate User Interface
IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR A NON-CLINICAL ACTIVITY, THE CENTRAL CONTROL MONITOR (CCM) WAS FOUND TO BE CRACKED WHICH CAUSED THE CURSOR SELECTIONS TO BE INACCURATE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19537114 ↗MalfunctionApr 8, 2024Inadequate User Interface; Inadequate User Interface
IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR A NON-CLINICAL ACTIVITY, THE CENTRAL CONTROL MONITOR (CCM) TOUCHSCREEN WAS OUT OF CALIBRATION. THE USER FACILITY ATTEMPTED A RECALIBRATION AND THE ARROW ON THE DISPLAY WENT MISSING CAUSING THE CCM TO BE UNUSABLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19067640 ↗
Other Console, Heart-Lung Machine, Cardiopulmonary Bypass devices
- Terumo® Advanced Perfusion System 1 801188Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass3 recalls447 reports
- MAQUET CARDIOHELP CARDIOHELP-iMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 recalls2 deaths64 reports
- MAQUET CARDIOHELP CARDIOHELP SystemMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 recalls18 deaths107 reports
- Terumo® Advanced Perfusion System 1 816300Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass300 reports
- Terumo® Advanced Perfusion System 1 801763Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 death230 reports
- Terumo® Advanced Perfusion System 1 816571Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass230 reports
- ESSENZ HLMLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall1 death110 reports
- Terumo® Advanced Perfusion System 1 5773Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 recall78 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page