Intuitive Surgical, Inc. · CAUTERY SPATUAL TIP · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00886874111031), received between Feb 17, 2023 and Oct 18, 2023 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Fire1
  • Melted1
  • Thermal Decomposition of Device1

Patient problems

  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 18, 2023Thermal Decomposition of Device; Thermal Decomposition of Device

    INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE CAUTERY SPATULA TIP (TIP) AS A DISCREPANT RMA ALONG WITH THE PRODUCT RETURNED UNDER PATIENT IDENTIFIER (B)(6) . THERE WAS NO ALLEGED MALFUNCTION OR PATIENT INJURY REPORTED AGAINST THIS PRODUCT. ISI MADE MULTIPLE FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.

    Patient problems: Insufficient Information

    MAUDE report 17955164 ↗
  • InjuryFeb 17, 2023Fire; Melted; Detachment of Device or Device Component; Fire; Melted; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED ORAL LARYNGECTOMY PROCEDURE, THE SP CAUTERY SPATULA TIP ACCESSORY, WHILE INSTALLED ON THE MONOPOLAR CAUTERY INSTRUMENT, CAUGHT ON FIRE WHILE IN THE ORAL CAVITY. THE INSTRUMENT WAS REMOVED AND THERE WAS REPORTEDLY NO ADDITIONAL PATIENT HARM. INTUITIVE SURGICAL, INC. (ISI) CONTACTED THE ROBOTICS COORDINATOR WHO WAS NOT PRESENT WHEN THIS EVENT OCCURRED BUT ARRIVED TO THE ROOM ABOUT FIVE MINUTES AFTERWARDS. THE SURGEON AND SURGICAL STAFF INFORMED THE ROBOTICS COORDINATOR THAT THE MONOPOLAR CAUTERY INSTRUMENT WAS BEING USED WITH THE SPATULA TIP. THE PATIENT WAS INTUBATED NASALLY. THE INSTRUMENT REPORTEDLY APPEARED NORMAL DURING INSPECTION AND USE, UNTIL THE SURGEON ACTIVATED CAUTERY WITH IT FOR THE FIRST TIME, THEN THE INSTRUMENT CAUGHT ON FIRE. THE ENERGY SETTINGS WERE UNKNOWN. THE SURGICAL STAFF SAID THEY REMOVED THE INSTRUMENT IMMEDIATELY UPON OBSERVING THE FLAME. THE SURGEON REPORTED TO THE ROBOTICS COORDINATOR THAT THERE WAS NO PATIENT INJURY, BUT THAT TWO PIECES OF THE SPATULA TIP'S WHITE SLEEVE FELL INTO THE PATIENT. THE SURGEON REPORTED RETRIEVING BOTH FRAGMENTS, AND THAT THERE WAS NO ADDITIONAL PATIENT INJURY OR EFFECT ON THE PROC

    Patient problems: Foreign Body In Patient

    MAUDE report 16397942 ↗