Intuitive Surgical, Inc. · TIP COVER ACCESSORY · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This da Vinci tip cover accessory is a protective cover fitted to surgical instrument tips. The 443 reports often describe covers tearing, splitting or coming loose during surgery, and some describe pieces ending up inside patients. There are 146 injuries, 297 malfunctions and no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
440
Deaths reported
0
Injuries reported
144
Recalls naming it
0

Summary

The FDA's MAUDE database contains 440 adverse event reports that list this exact device (DI 00886874111048), received between Dec 13, 2017 and Sep 23, 2026 — classified as 144 injuries, 295 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (196 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 4 reportsFeb 2024: 4 reportsMar 2024: 3 reportsApr 2024: 2 reportsMay 2024: 3 reportsJun 2024: 15 reportsJul 2024: 8 reportsAug 2024: 12 reportsSep 2024: 5 reportsOct 2024: 9 reportsNov 2024: 4 reportsDec 2024: 7 reportsJan 2025: 6 reportsFeb 2025: 1 reportsMar 2025: 8 reportsApr 2025: 4 reportsMay 2025: 18 reportsJun 2025: 8 reportsJul 2025: 15 reportsAug 2025: 9 reportsSep 2025: 10 reportsOct 2025: 7 reportsNov 2025: 10 reportsDec 2025: 21 reportsJan 2026: 13 reportsFeb 2026: 12 reportsMar 2026: 8 reportsApr 2026: 12 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 13 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202621
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn196
  • Detachment of Device or Device Component87
  • Material Separation64
  • Thermal Decomposition of Device40
  • Arcing25
  • Break21
  • Crack10
  • Material Fragmentation7

Patient problems

  • No Clinical Signs, Symptoms or Conditions320
  • Foreign Body In Patient35
  • Insufficient Information30
  • No Known Impact Or Consequence To Patient23
  • Device Embedded In Tissue or Plaque14
  • Burn(s)10
  • Hemorrhage/Blood Loss/Bleeding4
  • Unspecified Tissue Injury4

Recent adverse event reports

The 25 most recent of 440 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026Difficult to Open or Close

    It was reported that during a Vinci-assisted surgical procedure, the Monopolar Curved Scissors (MCS) instrument jaws would not close. Customer undocked the instrument and re-attached it, but it was not recognized. Customer removed the Trocar and forcibly bent the joint to remove the instrument. The procedure was completed with no patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50865247 ↗
  • MalfunctionAug 25, 2026Material Split, Cut or Torn

    It was reported that during a da Vinci-assisted surgical procedure, the tip cover accessory was torn intraoperative. The procedure was completed with no reported injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50424670 ↗
  • MalfunctionAug 25, 2026Material Split, Cut or Torn

    It was reported that during a da Vinci-assisted surgical procedure, the tip cover accessory was torn intraoperative. The procedure was completed with no reported injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50424667 ↗
  • MalfunctionAug 25, 2026Material Split, Cut or Torn

    It was reported that prior to starting-post anesthesia of a da Vinci-assisted surgical procedure, the Monopolar Curved Scissors (MCS) tip cover was brittle and torn. The procedure was completed with no reported injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50424666 ↗
  • MalfunctionAug 24, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50420157 ↗
  • MalfunctionAug 24, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50420156 ↗
  • MalfunctionAug 24, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50420155 ↗
  • MalfunctionAug 24, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50420154 ↗
  • MalfunctionAug 12, 2026Material Split, Cut or Torn

    It was reported that during a da Vinci-assisted surgical procedure, the scrub nurse reported that the silicone tip of the Tip Cover accessory was ruptured and proceeded to exchange it for a new one. The robotic coordinator reported that this was the fourth Tip Cover from the same lot to exhibit similar damage.\\\The procedure was completed with no reported injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50229490 ↗
  • MalfunctionAug 12, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50229489 ↗
  • MalfunctionAug 12, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50229357 ↗
  • MalfunctionAug 12, 2026Material Split, Cut or Torn

    It was reported that during a da Vinci-assisted surgical procedure, the scrub nurse reported that the silicone tip of the tip cover accessory had been ruptured during installation. The robotic coordinator indicated them to replace it with a new one and complete the surgery as planned.\\\The procedure was completed with no reported injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50228303 ↗
  • MalfunctionAug 11, 2026Material Separation

    It was reported that during a da Vinci-assisted surgical procedure, the Monopolar Curved Scissor (MCS) Tip Cover Accessory was lost during the removal of the forceps/gripper. A search was conducted in robotic mode, but the MCS Tip Cover Accessory could not be located. The da Vinci system was then undocked and the surgical procedure was converted to conventional laparoscopy, however the MCS Tip Cover Accessory was still not found. Preparations were then made for a laparotomy conversion. Prior to the initiation of the laparotomy conversion, the disposable MCS Tip Cover Accessory was discovered on the floor, which allowed the laparotomy conversion to be avoided. The procedure was completed with no reported injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50203074 ↗
  • InjuryAug 6, 2026Material Separation

    Refer to H11 for Follow-Up information.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50155739 ↗
  • OtherJul 30, 2026Material Split, Cut or Torn

    Patient problems: Burn(s)

    MAUDE report 50091269 ↗
  • MalfunctionJul 28, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50062236 ↗
  • InjuryMay 26, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED BENIGN HYSTERECTOMY PROCEDURE, THE MONOPOLAR CURVED SCISSORS (MCS) TIP COVER ACCESSORY, INSTALLED ON AN MCS INSTRUMENT, FELL OFF INSIDE THE PATIENT¿S BODY. THE MCS TIP COVER ACCESSORY WAS RETRIEVED WITHOUT COMPLICATIONS DURING THE SAME PROCEDURE. THE CUSTOMER INDICATED THAT EXCESSIVE LUBRICATION MAY HAVE CONTRIBUTED TO THE MCS TIP COVER ACCESSORY DISLODGING. THE PROCEDURE WAS COMPLETED AS PLANNED. INTUITIVE SURGICAL, INC. (ISI) HAS ATTEMPTED TO GATHER ADDITIONAL INFORMATION REGARDING THE REPORTED EVENT; HOWEVER, NO RESPONSE HAS BEEN RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25294303 ↗
  • MalfunctionMay 18, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE SURGEON NOTICED A SMALL PART OF THE MONOPOLAR CURVED SCISSORS (MCS) TIP COVER ACCESSORY WAS TORN. IT WAS IMMEDIATELY REMOVED FROM THE CAVITY AND REPLACED WITH ANOTHER ONE TO PREVENT BURNS TO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25196967 ↗
  • InjuryMay 5, 2026Material Separation

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, AN MONOPOLAR CURVED SCISSORS (MCS) TIP COVER ACCESSORY FELL INSIDE THE PATIENT. A PATIENT WHO HAD A CASE ON (B)(6) 2025, RECENTLY TOOK A COMPUTED TOMOGRAPHY (CT) IMAGE AND FOUND THAT SOMETHING LIKE A TIP COVER WAS REFLECTED. IT COULD NOT BE CONFIRMED BY INTRAOPERATIVE X-RAYS, BUT WHEN THE CUSTOMER CHECKED THE INTRAOPERATIVE VIDEO, AND CONFIRMED THE IMAGE OF THE TIP COVER FALLING OFF. THE PATIENT'S CONDITION IS UNKNOWN AT THIS TIME. THE PROCEDURE WAS COMPLETED. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: BASED ON THE SURGICAL CASE INVOLVING THE MCS TIP COVER ACCESSORY, THE HOSPITAL PROVIDED A CLEAR EXPLANATION TO THE PATIENT REGARDING THE ACCESSORY¿S STATUS. WITH THE PATIENT¿S INFORMED CONSENT, NO ADDITIONAL SURGICAL PROCEDURE WAS PERFORMED TO RETRIEVE THE MCS TIP COVER ACCESSORY; INSTEAD, THE PATIENT IS CURRENTLY UNDER OBSERVATION. THERE IS NO PLAN TO REMOVE OR RETURN THE TIP COVER BECAUSE THE FRAGMENT WAS NOT RETRIEVED AS THE SURGERY WAS NOT PERFORMED. WHETHER THE MCS INSTRUMENT COLLIDED WITH OTHER INSTRUMENTS, IF THE TIP COVER ACCESSORY W

    Patient problems: Foreign Body In Patient

    MAUDE report 25081902 ↗
  • MalfunctionApr 28, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE SCISSORS TIP COVER ACCESSORY WAS BROKEN DURING INTRA-OP PROCEDURE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25011213 ↗
  • MalfunctionApr 28, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE SCISSORS TIP COVER ACCESSORY WAS BROKEN DURING INTRA-OP PROCEDURE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25011210 ↗
  • MalfunctionApr 28, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, A TEAR WAS SEEN ON THE SCISSORS TIP COVER ACCESSORY DURING INTRA-OP PROCEDURE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25011184 ↗
  • MalfunctionApr 28, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE TIP COVER ACCESSORY (DA VINCI XI) TORN INTRAOPERATIVELY. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25011178 ↗
  • MalfunctionApr 24, 2026Material Split, Cut or Torn

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24989518 ↗
  • MalfunctionApr 23, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE MONOPOLAR CURVED SCISSORS (MCS) TIP COVER ACCESSORY HAD TEARS IN THE GRAY SECTION. THE PROCEDURE WAS COMPLETING AS PLANNED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24973968 ↗

FDA clearance and classification

The FDA regulates product code NAY (system,surgical,computer controlled instrument) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the None 400180

Has the None 400180 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00886874111048), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the None 400180?

The FDA's MAUDE database lists 440 adverse event reports for this device, received between Dec 13, 2017 and Sep 23, 2026. In those reports, the most common specific device problems are “Material Split, Cut or Torn” (196), “Detachment of Device or Device Component” (87) and “Material Separation” (64); the most common patient problems are “Foreign Body In Patient” (35), “Device Embedded In Tissue or Plaque” (14) and “Burn(s)” (10). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the None 400180?

None of the reports are classified as deaths, 144 as injuries and 295 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the None 400180, and is it still sold?

The FDA's device database lists Intuitive Surgical, Inc. as the company responsible for it (brand None, model 400180). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other System,surgical,computer controlled instrument devices

All System,surgical,computer controlled instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page