EndoWrist;daVinci Si 420049Recalls & reported problems

Intuitive Surgical, Inc. · CADIERE FORCEPS · Class 2

All models of EndoWrist;daVinci Si (35)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Intuitive Surgical da Vinci EndoWrist Cadieres forceps instrument has 40 adverse event reports naming its identifier, mostly describing broken cables and tips, sometimes with detached pieces or torn material. The reports include no deaths, 6 injuries, and 34 malfunctions. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 00886874111338), received between Jan 30, 2017 and Mar 16, 2026 — classified as 6 injuries, 34 malfunctions.

The most frequently reported specific device problem is “Break” (18 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break18
  • Detachment of Device or Device Component11
  • Material Split, Cut or Torn6
  • Material Fragmentation4
  • Unintended System Motion3
  • Detachment Of Device Component2
  • Material Frayed1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions21
  • No Known Impact Or Consequence To Patient14
  • Insufficient Information2
  • Foreign Body In Patient1
  • No Patient Involvement1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 16, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24612459 ↗
  • MalfunctionJan 30, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24228438 ↗
  • MalfunctionJan 30, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24228390 ↗
  • MalfunctionSep 9, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23010923 ↗
  • MalfunctionSep 5, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22984676 ↗
  • MalfunctionJun 4, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22145657 ↗
  • MalfunctionMar 10, 2025Break; Break

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED MEDIASTINAL MASS RESECTION SURGICAL PROCEDURE, THE CADIERE FORCEPS HAD A BROKEN TIP. THE INSTRUMENT STILL HAD A LIFE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. NO FRAGMENT FELL INTO THE PATIENT. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE SITE AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE INCIDENT OCCURRED DURING SURGERY, BUT NO FRAGMENTS OF THE MATERIAL WERE LOST AND NO HARM WAS DONE TO THE PATIENT. THERE HAVE BEEN NO REPORTS OF FRAGMENTS FALLING FROM THE DEVICE DURING USE. NO ACTION WAS TAKEN AS NO FRAGMENTS FELL DURING USE. THE SURGERY WAS COMPLETED SUCCESSFULLY AND THERE WERE NO COMPLICATIONS IN THE POST-OPERATIVE PERIOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21567688 ↗
  • MalfunctionFeb 7, 2025Material Fragmentation; Material Fragmentation

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21330327 ↗
  • MalfunctionOct 3, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED BILIODIGESTIVE BYPASS SURGICAL PROCEDURE, THE PULLEY CABLES ON THE CADIERE FORCEPS WERE EXPOSED AND BROKEN. THE SURGEON HAD USED THE CADIERE FORCEPS AS A SEPARATOR WITH NO ISSUE IN THE LAST FIVE HOURS PRIOR TO THE INSTRUMENT BREAKAGE. THE SURGEON WAS NOT GRABBING ANYTHING WITHIN THE JAWS WHEN THE INCIDENT OCCURRED. THE CUSTOMER THEN STRAIGHTENED, OPENED, AND REMOVED THE INSTRUMENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL INC. (ISI) FOLLOWED UP WITH THE CUSTOMER TO CONFIRM THAT THE CADIERE FORCEPS WAS INSPECTED PRIOR TO USE. THE CADIERE FORCEPS DID NOT COLLIDE WITH ANY OTHER INSTRUMENTS. NO FRAGMENTS FELL INSIDE THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17865014 ↗
  • MalfunctionJun 15, 2023Unintended System Motion; No Apparent Adverse Event; Unintended System Motion; No Apparent Adverse Event

    REFER TO H10/H11 FOR FOLLOW-UP INFORMATION. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE CADIERE FORCEPS INSTRUMENT DID NOT MOVE BY ITSELF, ONLY THE SURGEON COULD NO LONGER MOVE IT IN A CERTAIN POSITION DUE TO THE FAILURE, IT REMAINED STATIONARY WITHOUT GENERATING ANY RISK TO THE PATIENT. THE ISSUE OCCURRED WITH THE SURGEON¿S HEAD INSIDE THE SURGEON CONSOLE¿S HIGH RESOLUTION STEREO VIEWER WHILE THE SURGEON WAS ATTEMPTING TO MANIPULATE INSTRUMENTS FROM THE SURGEON CONSOLE. THE INSTRUMENT REMAINED STATIC. INSTRUMENT MOVEMENT SCALED APPROPRIATELY, MOVED IN THE INTENDED DIRECTION, WITH APPROPRIATE TIMING. THERE WAS NO SHAKINESS/FRICTION OBSERVED. THE CUSTOMER WAS UNABLE TO TELL IF THERE WAS INSTRUMENT OR ARM INTERFERENCE. THERE WERE NO COLLISIONS. THE ISSUE WAS PERSISTENT. THE SURGEON WAS ATTEMPTING TO TAKE USUAL MOVEMENTS WITH THE INSTRUMENT. THE INSTRUMENT WAS REPLACED TO RESOLVE THE ISSUE. THE SAFETY WRENCH WAS USED TO REMOVE THE DEFECTIVE INSTRUMENT. THERE WAS NO INJURY TO THE PATIENT. THE INSTRUMENT WAS INSPECTED PRIOR TO USE AND NOTHING OUT OF THE ORDINARY WAS NOTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17136539 ↗
  • MalfunctionMay 21, 2023Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT DURING CENTRAL PROCESSING, THE CADIERE FORCEPS HAD A GRASPER PADDLE WHICH BROKE OFF DURING THE FIRST STERILIZATION. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16971248 ↗
  • MalfunctionMay 21, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED LOBECTOMY SURGICAL PROCEDURE, THE CADIERE FORCEPS INSTRUMENT TIP AND THE INSERT WERE MISALIGNED AND IT WAS NOT POSSIBLE TO TAKE THE INSTRUMENT OUT FROM THE CANNULA. THE INSTRUMENT WAS RETRIEVED FROM THE PATIENT ALONG WITH THE CANULA AND THE ARM. THEN THE ONLY WAY TO TAKE OUT THE INSTRUMENT FROM THE CANULA WAS CUTTING THE DISTAL END OF THE INSTRUMENT. THIS WAS DONE AWAY FROM THE PATIENT. A 3RD PARTY INSTRUMENT WAS USED AND THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. THE CUSTOMER PROVIDED A PHOTOGRAPH SHOWING THE INSTRUMENT TIP DISLODGED FROM THE INSTRUMENT TUBE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16971188 ↗
  • InjuryMay 13, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED CHOLECYSTECTOMY SURGICAL PROCEDURE, THE CUSTOMER NOTED THE CADIERE FORCEPS INSTRUMENT JAW BROKE OFF WHEN HITTING ANOTHER INSTRUMENT AND FELL INTO THE PATIENT. THE SURGEON REMOVED THE FRAGMENT FROM THE PATIENT. IT WAS NOTED THAT THE FRAGMENT WAS RETRIEVED USING A BACKUP INSTRUMENT. THE INTUITIVE REPRESENTATIVE MADE SURE THE FRAGMENT REMOVED MATCHED UP TO THE INSTRUMENT TO CONFIRM NOTHING WAS LEFT BEHIND. THE INSTRUMENT WRIST WAS STRAIGHTENED UPON REMOVAL. THE FRAGMENT WAS AVAILABLE FOR RETURN. NO VIDEO/IMAGE WAS AVAILABLE FOR REVIEW. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE CLINICAL TERRITORY ASSOCIATE (CTA) WHO CREATED THIS COMPLAINT AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE CTA INFORMED THE CADIERE FORCEPS INSTRUMENT WAS IN USE FOR ABOUT IN ON HOUR PRIOR TO THE REPORTED FRAGMENT FALLING INTO THE PATIENT. IT WAS NOTED THE INSTRUMENT WAS INSPECTED PRIOR TO USE WITH NO DAMAGE OBSERVED. THE CTA INFORMED THE CADIERE FORCEPS INSTRUMENT WAS REMOVED AT LEAST ONCE PRIOR TO THE FRAGMENT FALLING. WHEN THE REPORTED FRAGMENT FELL INTO THE PATIENT, THE CTA INFORMED THE CA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16928186 ↗
  • MalfunctionMay 6, 2023Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED ENUCLEATION OF ESOPHAGEAL LEIOMYOMA SURGICAL PROCEDURE, THE CADIERE FORCEPS INSTRUMENT WAS FOUND TO HAVE PHYSICAL DAMAGE. BACKUP INSTRUMENT OF SAME KIND WAS USED TO COMPLETE THE PROCEDURE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16883588 ↗
  • MalfunctionOct 13, 2022Unintended System Motion; Unintended System Motion

    REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15601883 ↗
  • MalfunctionOct 1, 2022Unintended System Motion; Unintended System Motion

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CADIERE FORCEPS INSTRUMENT MALFUNCTIONED. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE CADIERE FORCEPS INSTRUMENT WAS REPORTEDLY MOVING NON-INTUITIVELY DURING THE PROCEDURE. THE INSTRUMENT WAS INSPECTED PRIOR TO USE WITH NO DAMAGE. THERE WAS NO PATIENT INJURY. THE PROCEDURE WAS COMPLETED ROBOTICALLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15526103 ↗
  • MalfunctionSep 7, 2022Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PULMONARY LOBECTOMY SURGICAL PROCEDURE, A STEEL CABLE CONTROLLING THE CADIERE FORCEPS INSTRUMENT BROKE. A BACKUP INSTRUMENT OF THE SAME TYPE WAS USED AND THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND PATIENT DEMOGRAPHIC INFORMATION WAS REQUESTED, BUT NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15375435 ↗
  • MalfunctionAug 14, 2022Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PULMONARY LOBECTOMY PROCEDURE, THE CADIERE FORCEPS INSTRUMENT'S CABLE BROKE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. THE CADIERE FORCEPS IS A MULTIPLE-USE ENDOSCOPIC INSTRUMENT WITH A GRASPING TIP TO BE USED IN CONJUNCTION WITH THE DA VINCI SYSTEM. THE INSTRUMENT IS DESIGNED TO GRAB, MANIPULATE, RETRACT, AND DISSECT TISSUE DURING A DA VINCI ASSISTED SURGICAL PROCEDURE. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER FOR ADDITIONAL INFORMATION; HOWEVER, NO RESPONSE HAS BEEN RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15222719 ↗
  • InjuryNov 24, 2021Material Frayed; Material Frayed

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PULMONARY SEGMENTECTOMY PROCEDURE OF THE LEFT UPPER LUNG, FRAYING OF A WIRE ON A CADIERE FORCEPS INSTRUMENT WAS IDENTIFIED AFTER ARTERIOVENOUS TREATMENT. THE WIRE/CABLE DAMAGE WAS IDENTIFIED IN THE FINAL STAGE OF THE PROCEDURE. HOWEVER, IT IS UNCLEAR IF THE SURGEON CONTINUED TO USE THE INSTRUMENT AFTER THE INSTRUMENT DAMAGE WAS FOUND. DURING A LEAK TEST, LEAKS WERE CONFIRMED FROM THE LOWER LOBES INSTEAD OF THE UPPER LOBES. THE LEAK WAS FROM THE LOWER LOBE, WHICH THE SURGEON DID NOT SPECIFICALLY TOUCH ACCORDING TO THE INITIAL REPORTER. SINCE IT WAS ON THE INSERTION PATH OF THE CADIERE FORCEPS, IT WAS SPECULATED THAT THE FRAYED WIRE MAY HAVE DAMAGED THE LOWER LOBE AND CAUSED THE LEAK. THE SURGEON TRIED TO CLOSE THE LEAK; HOWEVER, THE SURGEON CONVERTED THE PROCEDURE TO VATS TO FIX THE LEAK DUE TO A LACK OF A CLEAR SURGICAL VIEW. AN UNSPECIFIED DOCTOR FROM THE SITE COMMENTED: THE LEAK WAS COMING FROM THE LOWER LOBE, WHICH HAD NOT BEEN SPECIFICALLY TOUCHED. THE AIR LEAK WAS NOTED TO BE COMING FROM A SMALL HOLE THAT WAS DISCOVERED IN THE LOWER LOBE WHICH WAS ALONG THE INSERTION PATH OF THE CADIERE FORCEPS INSTRUMENT. THE SURGEON SUSPECTED T

    Patient problems: Unspecified Tissue Injury

    MAUDE report 12872677 ↗
  • InjuryAug 27, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12381060 ↗
  • InjuryJul 20, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12200572 ↗
  • InjuryMay 27, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS INITIALLY REPORTED THAT THE CADIERE FORCEPS INSTRUMENT WAS USED TO GUIDE AN ULTRASOUND PROBE DURING THE PROCEDURE, AND ONE OF THE INSTRUMENT TIPS BROKE OFF AND FELL INTO THE PATIENT. THE INSTRUMENT TIP WAS REPORTED TO HAVE BEEN RETAINED IN THE PATIENT. FOLLOW-UP: ON 05/07/2021, INTUITIVE SURGICAL INC. (ISI) CONTACTED THE CUSTOMER AND ADDITIONAL INFORMATION WAS OBTAINED ABOUT THE COMPLAINT: THE CADIERE FORCEPS INSTRUMENT BROKE WHEN IT WAS GRASPING AN ULTRASOUND PROBE TO ASSIST IN GUIDING IT TO OBTAIN A VIEW OF THE TUMOR. THE SURGEON SEARCHED FOR THE FRAGMENT IN THE BOWEL, BUT COULD NOT FIND IT. THE SURGEON SAID THAT RETAINING THIS FRAGMENT WAS EQUAL TO A PATIENT RETAINING A CLIP ON THEIR ARTERY AND THERE WAS NO HARM. AN X-RAY WAS PERFORMED AFTER THE PROCEDURE WAS COMPLETED DURING THE SAME DAY; THE FRAGMENT COULD NOT BE FOUND IN THE X-RAY. IT WAS UNKNOWN IF THE SURGEON HAD AN OPINION ON HOW THE INSTRUMENT BROKE OR FOR HOW LONG THE INSTRUMENT WAS IN USE FOR BEFORE IT BROKE. IT WAS REPORTED THERE WAS NO PATIENT INJURY DURING THIS PROCEDURE. IT WAS REPORTED THAT THE PATIENT WAS DISCHARGED AND HAS NOT RETURNED TO THE HOSPITAL. THE PROCEDURE WAS A SMALL INTESTINE SEGMENTAL RESECTIO

    Patient problems: Foreign Body In Patient

    MAUDE report 11891918 ↗
  • InjuryFeb 15, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11330712 ↗
  • MalfunctionJan 30, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PULMONARY LOBECTOMY SURGICAL PROCEDURE, THE CADIERE FORCEPS INSTRUMENT STOPPED RESPONDING TO COMMANDS. THE STEEL CABLES HAD BEEN CHECKED. A BACKUP INSTRUMENT OF THE SAME TYPE WAS USED AND THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11255620 ↗
  • MalfunctionNov 18, 2020Break

    WE HAD AN INSTRUMENT BREAK WHILE INSIDE THE PATIENT DURING A DA VINCI HERNIA REPAIR WITH THE DOCTOR. IT WAS THE CADIERE FORCEPS AND THEY JUST QUIT FUNCTIONING BUT THEY WERE INSIDE THE PATIENT WHEN THAT HAPPENED. NOTE: FORCEP CHECKED AFTER REMOVAL AND IT WAS NOTED TO HAVE ZERO (0) LIVES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10860009 ↗

Other System,surgical,computer controlled instrument devices

All System,surgical,computer controlled instrument devices →