EndoWrist;daVinci Si 420093Recalls & reported problems
Intuitive Surgical, Inc. · PROGRASP FORCEPS · Class 2
All models of EndoWrist;daVinci Si (35)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This is a ProGrasp forceps, a wristed robotic instrument used with the da Vinci Si surgical system. The 63 reports mostly describe the instrument breaking, parts coming loose, and jaws that are hard to open or close. Six reports involve injuries and 57 describe malfunctions; no deaths were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 63
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 63 adverse event reports that list this exact device (DI 00886874111345), received between Mar 24, 2017 and Mar 16, 2026 — classified as 6 injuries, 57 malfunctions.
The most frequently reported specific device problem is “Break” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break15
- Detachment of Device or Device Component10
- Unintended System Motion9
- Difficult to Open or Close8
- Material Split, Cut or Torn8
- Detachment Of Device Component7
- Device Dislodged or Dislocated6
- Material Fragmentation2
Patient problems
- No Clinical Signs, Symptoms or Conditions34
- No Known Impact Or Consequence To Patient18
- Insufficient Information4
- Device Embedded In Tissue or Plaque3
- Foreign Body In Patient1
- Hemorrhage/Blood Loss/Bleeding1
- Tissue Damage1
- Unspecified Tissue Injury1
Recent adverse event reports
The 25 most recent of 63 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 16, 2026Material Split, Cut or Torn
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: Insufficient Information
MAUDE report 24612452 ↗MalfunctionFeb 26, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT PRIOR TO THE START OF A DA VINCI-ASSISTED PROSTATECTOMY SURGICAL PROCEDURE, THE PROGRASP FORCEPS HAD BROKEN CABLES. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24452044 ↗MalfunctionFeb 24, 2026Material Frayed
IT WAS REPORTED THAT THERE WAS A FRAYED CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24438247 ↗MalfunctionOct 23, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: Insufficient Information
MAUDE report 23374495 ↗MalfunctionJun 18, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22287787 ↗MalfunctionMar 23, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21674305 ↗MalfunctionMar 20, 2025Break; Break
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21658250 ↗MalfunctionMar 13, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21600084 ↗MalfunctionMar 13, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21600080 ↗MalfunctionMar 4, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21512074 ↗MalfunctionFeb 25, 2025Material Frayed; Material Frayed
IT WAS REPORTED THAT THERE WAS A FRAYED CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21459205 ↗MalfunctionJun 6, 2024Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED RADICAL CURE OF PROSTATE CANCER SURGICAL PROCEDURE, THE 8MM PROGRASP FORCEPS HAD THE JOINT DISCONNECTED FROM THE PLIERS. ISI FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE PROCEDURE NAME WAS RADICAL CURE OF PROSTATE CANCER. THE SURGEON'S NAME WAS (B)(6). THERE WERE NO FRAGMENTS THAT FELL IN THE PATIENT WHILE THE INSTRUMENT'S USE WITHIN THE PATIENT. THE INSTRUMENT WAS CHECKED BEFORE USE. NO ABNORMALITIES WERE FOUND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19474147 ↗MalfunctionDec 1, 2023Difficult to Open or Close; Difficult to Open or Close
REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18244176 ↗MalfunctionNov 3, 2023Unintended System Motion; Unintended System Motion
REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18067706 ↗MalfunctionOct 26, 2023Unintended Movement; Unintended Movement
REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.
Patient problems: Insufficient Information
MAUDE report 18013420 ↗InjuryAug 25, 2023Defective Component; Defective Component
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PROSTATECTOMY PROCEDURE, ONE SIDE OF THE PROGRASP FORCEPS INSTRUMENT JAW WAS NOT MOVING SMOOTHLY. THE INSTRUMENT WAS REMOVED AND IT WAS OBSERVED THAT A WIRE HAD BECOME DISLODGED WITHIN AND ¿THE DISC¿ DID NOT WORK. IT WAS REPORTED THAT TISSUE WAS LODGED IN THE SMALL CREVICES OF THE INSTRUMENT. IMMEDIATELY AFTER REMOVING THE INSTRUMENT, MINOR TISSUE DAMAGE WAS IDENTIFIED WHEN THE SURGICAL STAFF CHECKED THE PATH OF REMOVAL OF THE PROGRASP FORCEPS INSTRUMENT FROM THE CANNULA TO THE TARGET TISSUE TO ENSURE THERE WERE NO ISSUES IN THE ABDOMINAL CAVITY. IT IS UNKNOWN AS TO WHAT TYPE OF MINOR TISSUE DAMAGE/INJURY WAS NOTED ON THE INSIDE OF THE PERITONEUM. THERE WAS NO BLEEDING OBSERVED AND NO FRAGMENTS FELL INSIDE THE PATIENT¿S BODY. WHEN CHECKING THE INSTRUMENT AFTER REMOVAL, THEY ¿RESTORED¿ THE WIRE AND THE PROGRASP FORCEPS STILL DID NOT MOVE. AFTER CONSULTING WITH THE GENERAL SURGERY COLORECTAL TEAM, THE MINOR TISSUE DAMAGE WAS REPAIRED WITH TWO STITCHES. THE PROCEDURE WAS COMPLETED ROBOTICALLY USING A BACK-UP PROGRASP FORCEPS INSTRUMENT. THE SURGEON DID NOT ADMINISTER ANY ADDITIONAL TREATMENT TO THE TISSUES. THE PATIENT DID NOT EXPERIENCE
Patient problems: Unspecified Tissue Injury
MAUDE report 17627871 ↗MalfunctionAug 12, 2023Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE PROGRASP FORCEPS INSTRUMENT DID NOT MOVE THE WAY THE SURGEON WANTED IT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) MADE MULTIPLE FOLLOW-UPS ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17532045 ↗MalfunctionJul 26, 2023Unintended System Motion; Unintended System Motion
REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17398848 ↗MalfunctionJun 5, 2023Visual Prompts will not Clear; Visual Prompts will not Clear
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PARTIAL NEPHRECTOMY SURGICAL PROCEDURE, AFTER THE KIDNEY WAS UNCLAMPED AND THE BEGINNING OF THE RECONSTRUCTION WAS OCCURRING; REPEATED RECOVERABLE 48100 FAULTS OCCURRED. THE FAULT REMAINED EVEN AFTER TROUBLESHOOTING WITH DVSTAT AND REBOOTING THE SYSTEM AND COMPLETING A HARD CYCLE POWER. SINCE THE NEEDLE DRIVER INSTRUMENT WAS HOLDING A NEEDLE THE INSTRUMENT RELEASE KIT (IRK) WAS USED TO SAFELY REMOVE IT WHICH IS WHY THE INSTRUMENT IS BEING RETURNED. THE SURGEON MADE THE DECISION TO CONVERT THE PROCEDURE TO TRADITIONAL LAPAROSCOPIC TECHNIQUES WITH NO REPORTED INJURY. THE PROGRASP FORCEPS INSTRUMENTS ARE MULTIPLE-USE ENDOSCOPIC INSTRUMENTS WITH A GRASPING TIP TO BE USED IN CONJUNCTION WITH THE DA VINCI SYSTEM. THE INSTRUMENT IS DESIGNED TO GRAB, MANIPULATE, RETRACT, AND DISSECT TISSUE DURING A DA VINCI ASSISTED SURGICAL PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17067484 ↗MalfunctionJun 5, 2023Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE PROGRASP FORCEPS INSTRUMENT MADE MOVEMENTS IN THE OPPOSITE DIRECTIONS FROM WHAT WAS INDICATED/INTENDED BY THE SURGEON. THE SITE RESTARTED THE SYSTEM AND REPOSITIONED THE SYSTEM ARM WITH NO CHANGE. THE SITE THEN REPLACED THE INSTRUMENT TO RESOLVE THE PROBLEM AND CONTINUE WITH THE PROCEDURE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17067320 ↗MalfunctionJun 5, 2023Unintended System Motion; Unintended System Motion
IT WAS REPORTED, THAT DURING A DA VINCI-ASSISTED RADICAL SALVAGE PROSTATECTOMY SURGICAL PROCEDURE. THE PROGRASP FORCEPS INSTRUMENT SHOWED AN INVERSION OF ROTATION, YAW AND TILT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY USING A BACKUP PROGRASP FORCEPS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17067318 ↗MalfunctionJun 5, 2023Unintended System Motion; Difficult to Open or Close; Unintended System Motion; Difficult to Open or Close
IT WAS REPORTED DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE INSTRUMENT JAWS COULD NOT BE OPENED OR CLOSED PROPERLY. THE SITE USED A BACK-UP INSTRUMENT TO CONTINUE THE PROCEDURE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17067317 ↗MalfunctionJun 5, 2023Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PROSTATECTOMY SURGICAL PROCEDURE, THE INSTRUMENT WOULD MOVE IN THE OPPOSITE DIRECTIONS AS THAT OF THE MANIPULATORS. A BACKUP INSTRUMENT WAS USED. THE PROCEDURE WAS COMPLETED WITH A BACKUP INSTRUMENT AND THERE NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17067316 ↗MalfunctionJun 5, 2023Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE JAW OF THE PROGRASP FORCEPS WAS NOT FULLY OPENED AND WAS NOT MOVING SMOOTHLY. THE CUSTOMER USED A BACKUP INSTRUMENT TO CONTINUE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17067311 ↗MalfunctionMay 28, 2023Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE INSTRUMENT WAS FOUND TO HAVE A RESTRICTED JOINT MOVEMENT. A BACKUP INSTRUMENT WAS USED. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17016017 ↗
Other System,surgical,computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,225 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,156 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,581 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,302 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls235 deaths5,316 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,310 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,770 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,273 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗