None 470002Recalls & reported problems
Intuitive Surgical, Inc. · CANNULA, STANDARD, REUSABLE · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00886874112137), received between Apr 18, 2024 and Dec 4, 2025 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Detachment of Device or Device Component1
- Material Fragmentation1
- Material Separation1
- No Apparent Adverse Event1
Patient problems
- Foreign Body In Patient3
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 4, 2025No Apparent Adverse Event; No Apparent Adverse Event
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: Insufficient Information
MAUDE report 23722026 ↗InjuryJul 10, 2025Material Separation; Material Separation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED MYOMECTOMY SURGICAL PROCEDURE, THE TIP WAS BENT, AND A PIECE OF THE BLACK PART FELL INSIDE THE PATIENT, BUT THE DOCTOR RETRIEVED IT. THE PROCEDURE WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22469190 ↗InjuryFeb 25, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PARAESOPHAGEAL HIATAL HERNIA REPAIR PROCEDURE, THE COATING/METAL SHAVINGS ON SHAFT OF THE VESSEL SEALER EXTEND (VSE) INSTRUMENT SCRAPED OFF AND FELL INSIDE THE PATIENT'S ABDOMEN. THE INTUITIVE SURGICAL, INC. (ISI) CUSTOMER SALES REPRESENTATIVE (CSR) THAT WAS PRESENT DURING THE PROCEDURE STATED THAT THE SURGEON WAS ABLE TO REMOVE MOST OF THE METAL SHAVINGS. IT IS UNKNOWN IF ALL THE SHAVINGS WERE RETRIEVED. A POST OPERATIVE X-RAY WAS PERFORMED BUT THE RESULTS OF THE X-RAY ARE UNKNOWN. THE SURGEON REPORTED THAT THE PATIENT HAS NOT RETURNED WITH ANY POST-OPERATIVE COMPLICATIONS.
Patient problems: Foreign Body In Patient
MAUDE report 21461411 ↗InjuryAug 20, 2024Break; Break
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED ABDOMINOPERINEAL RESECTION (APR) SURGICAL PROCEDURE, THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT BROKE AT THE TIP AND COULD NOT BE REMOVED BY THE TROCAR SO THE TROCAR, ALONG WITH THE INSTRUMENT HAD TO BE REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: Foreign Body In Patient
MAUDE report 20003800 ↗MalfunctionJun 8, 2024Break; Break
REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19494532 ↗InjuryApr 18, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS INITIALLY REPORTED THAT DURING DA VINCI ASSISTED SURGICAL PROCEDURE, A MEGA NEEDLE DRIVER INSTRUMENT BROKE INSIDE THE PATIENT AND RESULTED IN FRAGMENT(S) FALLING INSIDE THE PATIENT. IT WAS LATER REPORTED THAT THE SURGICAL STAFF NOTICED PIECES BROKEN OFF FROM THE CANNULA. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE INSTRUMENT WAS INSPECTED PRIOR TO USE WITH NO NOTED DAMAGE. SUTURING WAS BEING PERFORMED WHEN THE INSTRUMENT BROKE. THE REPORTED STATED THAT TWO PIECES BROKE OFF AND ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. THE SURGEON DID NOT NOTICE ANY ISSUES WITH THE FUNCTIONALITY OF THE INSTRUMENT DURING THE SURGICAL PROCEDURE. THE INSTRUMENT REPORTEDLY DID NOT COLLIDE WITH ANY OTHER INSTRUMENTS OR OTHER HARD MATERIALS DURING THE SURGICAL PROCEDURE; HOWEVER, THE REPORTER INDICATED THAT THE FRAGMENT FELL INSIDE THE PATIENT DURING AN INSTRUMENT TIP/ACCESSORY COLLISION. UPON FINAL REMOVAL OF THE INSTRUMENT, THE WRIST WAS STRAIGHTENED, AND THE SURGICAL STAFF DID NOT FEEL ANY RESISTANCE OF THE INSTRUMENT THROUGH THE CANNULA. THE SURGICAL STAFF DID NOT NOTICE ANY OTHER DAMAGE TO THE INSTRUMENT AFTER
Patient problems: Foreign Body In Patient
MAUDE report 19130256 ↗
Other System,surgical,computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,225 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,156 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,581 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,302 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls235 deaths5,316 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,310 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,770 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,273 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page