Skip to content

None 470002Recalls & reported problems

Intuitive Surgical, Inc. · CANNULA, STANDARD, REUSABLE · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00886874112137), received between Apr 18, 2024 and Dec 4, 2025 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Detachment of Device or Device Component1
  • Material Fragmentation1
  • Material Separation1
  • No Apparent Adverse Event1

Patient problems

  • Foreign Body In Patient3
  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 4, 2025No Apparent Adverse Event; No Apparent Adverse Event

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 23722026 ↗
  • InjuryJul 10, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED MYOMECTOMY SURGICAL PROCEDURE, THE TIP WAS BENT, AND A PIECE OF THE BLACK PART FELL INSIDE THE PATIENT, BUT THE DOCTOR RETRIEVED IT. THE PROCEDURE WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22469190 ↗
  • InjuryFeb 25, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PARAESOPHAGEAL HIATAL HERNIA REPAIR PROCEDURE, THE COATING/METAL SHAVINGS ON SHAFT OF THE VESSEL SEALER EXTEND (VSE) INSTRUMENT SCRAPED OFF AND FELL INSIDE THE PATIENT'S ABDOMEN. THE INTUITIVE SURGICAL, INC. (ISI) CUSTOMER SALES REPRESENTATIVE (CSR) THAT WAS PRESENT DURING THE PROCEDURE STATED THAT THE SURGEON WAS ABLE TO REMOVE MOST OF THE METAL SHAVINGS. IT IS UNKNOWN IF ALL THE SHAVINGS WERE RETRIEVED. A POST OPERATIVE X-RAY WAS PERFORMED BUT THE RESULTS OF THE X-RAY ARE UNKNOWN. THE SURGEON REPORTED THAT THE PATIENT HAS NOT RETURNED WITH ANY POST-OPERATIVE COMPLICATIONS.

    Patient problems: Foreign Body In Patient

    MAUDE report 21461411 ↗
  • InjuryAug 20, 2024Break; Break

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED ABDOMINOPERINEAL RESECTION (APR) SURGICAL PROCEDURE, THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT BROKE AT THE TIP AND COULD NOT BE REMOVED BY THE TROCAR SO THE TROCAR, ALONG WITH THE INSTRUMENT HAD TO BE REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: Foreign Body In Patient

    MAUDE report 20003800 ↗
  • MalfunctionJun 8, 2024Break; Break

    REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19494532 ↗
  • InjuryApr 18, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS INITIALLY REPORTED THAT DURING DA VINCI ASSISTED SURGICAL PROCEDURE, A MEGA NEEDLE DRIVER INSTRUMENT BROKE INSIDE THE PATIENT AND RESULTED IN FRAGMENT(S) FALLING INSIDE THE PATIENT. IT WAS LATER REPORTED THAT THE SURGICAL STAFF NOTICED PIECES BROKEN OFF FROM THE CANNULA. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE INSTRUMENT WAS INSPECTED PRIOR TO USE WITH NO NOTED DAMAGE. SUTURING WAS BEING PERFORMED WHEN THE INSTRUMENT BROKE. THE REPORTED STATED THAT TWO PIECES BROKE OFF AND ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. THE SURGEON DID NOT NOTICE ANY ISSUES WITH THE FUNCTIONALITY OF THE INSTRUMENT DURING THE SURGICAL PROCEDURE. THE INSTRUMENT REPORTEDLY DID NOT COLLIDE WITH ANY OTHER INSTRUMENTS OR OTHER HARD MATERIALS DURING THE SURGICAL PROCEDURE; HOWEVER, THE REPORTER INDICATED THAT THE FRAGMENT FELL INSIDE THE PATIENT DURING AN INSTRUMENT TIP/ACCESSORY COLLISION. UPON FINAL REMOVAL OF THE INSTRUMENT, THE WRIST WAS STRAIGHTENED, AND THE SURGICAL STAFF DID NOT FEEL ANY RESISTANCE OF THE INSTRUMENT THROUGH THE CANNULA. THE SURGICAL STAFF DID NOT NOTICE ANY OTHER DAMAGE TO THE INSTRUMENT AFTER

    Patient problems: Foreign Body In Patient

    MAUDE report 19130256 ↗

Other System,surgical,computer controlled instrument devices

All System,surgical,computer controlled instrument devices →