EndoWrist 470007Recalls & reported problems
Intuitive Surgical, Inc. · ROUND TIP SCISSORS · Class 2
All models of EndoWrist (62)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00886874112168), received between Nov 3, 2020 and Apr 9, 2026 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (9 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 19, 2024Class IIOpen, Classified
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.
Z-0850-2025 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Split, Cut or Torn9
- Material Fragmentation3
- Material Frayed2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- Insufficient Information2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 9, 2026Material Frayed
IT WAS REPORTED THAT THERE WAS A FRAYED CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24836825 ↗MalfunctionMar 2, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24477803 ↗MalfunctionMar 2, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24477802 ↗MalfunctionNov 20, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: Insufficient Information
MAUDE report 23611047 ↗MalfunctionOct 23, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23374440 ↗MalfunctionSep 2, 2025Material Frayed; Material Frayed
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22947249 ↗MalfunctionMay 29, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22110397 ↗MalfunctionMay 28, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22101418 ↗MalfunctionMay 13, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22014660 ↗MalfunctionMay 9, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21998328 ↗MalfunctionApr 30, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT AFTER A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE 8MM ROUND TIP SCISSORS INSTRUMENT EXHIBITED A SMALL FRACTURE OF THE CABLE. NO FRAGMENTS FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21938793 ↗MalfunctionFeb 23, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE 8MM ROUND TIP SCISSORS INSTRUMENT'S CABLE BROKE. IT WAS UNKNOWN IF A FRAGMENT FELL INTO THE PATIENT. THE PROCEDURE AND PATIENT OUTCOME ARE UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 21441986 ↗MalfunctionJun 6, 2024Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE ROUND TIP SCISSORS INSTRUMENT WAS DAMAGED. THE WIRE OF THE INSTRUMENT WAS COMING OUT. THE PROCEDURE WAS COMPLETED ROBOTICALLY WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19481289 ↗MalfunctionJul 18, 2023Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT PRIOR TO STARTING A PROCEDURE THE ROUND TIP SCISSORS INSTRUMENT HAD BROKEN CABLES. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL INC. (ISI) FOLLOWED UP WITH THE CUSTOMER AND CONFIRMED THAT THE ISSUE WAS NOTICED DURING INSPECTION, PRIOR TO THE CASE, AND THERE WAS NO REPORT OF FRAGMENTS FALLING FROM THE DEVICE WHILE USED WITHIN A PATIENT SO THERE WAS NO FRAGMENT FROM THE BROKEN CABLES THAT FELL INSIDE A PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17340535 ↗MalfunctionNov 3, 2020Break; Break
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED MYOMECTOMY SURGICAL PROCEDURE, THE 8MM ROUND TIP SCISSORS INSTRUMENT CABLE BROKE DURING SURGERY. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) MADE MULTIPLE FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10776457 ↗
Other System,surgical,computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,225 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,156 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,581 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,302 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls235 deaths5,316 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,310 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,770 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,273 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page