Skip to content

EndoWrist 470007Recalls & reported problems

Intuitive Surgical, Inc. · ROUND TIP SCISSORS · Class 2

All models of EndoWrist (62)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00886874112168), received between Nov 3, 2020 and Apr 9, 2026 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (9 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 4 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 19, 2024Class IIOpen, Classified

    Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

    Z-0850-2025 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Split, Cut or Torn9
  • Material Fragmentation3
  • Material Frayed2
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • Insufficient Information2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 9, 2026Material Frayed

    IT WAS REPORTED THAT THERE WAS A FRAYED CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24836825 ↗
  • MalfunctionMar 2, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24477803 ↗
  • MalfunctionMar 2, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24477802 ↗
  • MalfunctionNov 20, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 23611047 ↗
  • MalfunctionOct 23, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23374440 ↗
  • MalfunctionSep 2, 2025Material Frayed; Material Frayed

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22947249 ↗
  • MalfunctionMay 29, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22110397 ↗
  • MalfunctionMay 28, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22101418 ↗
  • MalfunctionMay 13, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22014660 ↗
  • MalfunctionMay 9, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21998328 ↗
  • MalfunctionApr 30, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT AFTER A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE 8MM ROUND TIP SCISSORS INSTRUMENT EXHIBITED A SMALL FRACTURE OF THE CABLE. NO FRAGMENTS FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21938793 ↗
  • MalfunctionFeb 23, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE 8MM ROUND TIP SCISSORS INSTRUMENT'S CABLE BROKE. IT WAS UNKNOWN IF A FRAGMENT FELL INTO THE PATIENT. THE PROCEDURE AND PATIENT OUTCOME ARE UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 21441986 ↗
  • MalfunctionJun 6, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE ROUND TIP SCISSORS INSTRUMENT WAS DAMAGED. THE WIRE OF THE INSTRUMENT WAS COMING OUT. THE PROCEDURE WAS COMPLETED ROBOTICALLY WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19481289 ↗
  • MalfunctionJul 18, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT PRIOR TO STARTING A PROCEDURE THE ROUND TIP SCISSORS INSTRUMENT HAD BROKEN CABLES. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL INC. (ISI) FOLLOWED UP WITH THE CUSTOMER AND CONFIRMED THAT THE ISSUE WAS NOTICED DURING INSPECTION, PRIOR TO THE CASE, AND THERE WAS NO REPORT OF FRAGMENTS FALLING FROM THE DEVICE WHILE USED WITHIN A PATIENT SO THERE WAS NO FRAGMENT FROM THE BROKEN CABLES THAT FELL INSIDE A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17340535 ↗
  • MalfunctionNov 3, 2020Break; Break

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED MYOMECTOMY SURGICAL PROCEDURE, THE 8MM ROUND TIP SCISSORS INSTRUMENT CABLE BROKE DURING SURGERY. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) MADE MULTIPLE FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10776457 ↗

Other System,surgical,computer controlled instrument devices

All System,surgical,computer controlled instrument devices →