None 470015Recalls & reported problems
Intuitive Surgical, Inc. · ARM DRAPE · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00886874112199), received between Jun 29, 2022 and Jun 29, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Unintended System Motion” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Sep 21, 2021Terminated
Incorrect expiration date on its package labeling (shipper box and inner carton label).
Z-0275-2022 · Root cause: Labeling Change Control · FDA record ↗
Most reported problems
Device problems
- Unintended System Motion1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 29, 2022Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED LOW ANTERIOR RESECTION PROCEDURE, THE MEDIUM-LARGE CLIP APPLIER INSTRUMENT EXHIBITED NON-INTUITIVE MOVEMENT. THE INSTRUMENT JAW ALLEGEDLY "MOVED IN THE WRONG DIRECTION." TROUBLESHOOTING WAS PERFORMED BY REMOVING AND REINSTALLING THE SAME INSTRUMENT BACK INTO THE ARM AND THIS RESOLVED THE ISSUE. FOLLOWING THIS, THE INSTRUMENT MOVEMENT WAS CONFIRMED AND OPERATED WITHOUT ERROR. THE USER COMPLETED THE PROCEDURE WITH NO FURTHER ISSUE REPORTED. THE USER SUSPECTS A FAULTY DRAPE. NO FRAGMENT FELL INTO THE PATIENT. NO KNOWN IMPACT OR PATIENT CONSEQUENCE WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14863899 ↗
Other System,surgical,computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,225 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,156 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,581 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,302 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls235 deaths5,316 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,310 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,770 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,273 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.