EndoWrist 470036Recalls & reported problems
Intuitive Surgical, Inc. · DEBAKEY FORCEPS · Class 2
All models of EndoWrist (62)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 00886874112236), received between Dec 6, 2020 and Mar 31, 2026 — classified as 19 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (7 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Dec 19, 2024Class IIOpen, Classified
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.
Z-0852-2025 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Split, Cut or Torn7
- Material Fragmentation6
- Material Frayed3
- Detachment of Device or Device Component2
- Break1
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions18
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 31, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24749108 ↗MalfunctionFeb 26, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24465376 ↗MalfunctionFeb 25, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24450361 ↗MalfunctionFeb 23, 2026Material Frayed
IT WAS REPORTED THAT THERE WAS A FRAYED CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24422822 ↗MalfunctionFeb 20, 2026Material Fragmentation
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24407415 ↗MalfunctionNov 5, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23481422 ↗MalfunctionSep 16, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23068786 ↗MalfunctionSep 15, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23058743 ↗MalfunctionSep 5, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22984796 ↗MalfunctionAug 29, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22929134 ↗MalfunctionApr 24, 2025Material Frayed; Material Frayed
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE 8MM DEBAKEY FORCEPS INSTRUMENT HAD A FRAYED CABLE. NO FRAGMENTS FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21904872 ↗MalfunctionApr 2, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21756982 ↗MalfunctionMar 18, 2025Material Frayed; Material Frayed
IT WAS REPORTED THAT PRIOR TO THE START OF A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CONDUCTOR WIRE (BLACK) WAS FOUND DAMAGED THE PROCEDURE WAS COMPLETING AS PLANNED WITH NO REPORTED INJURY. ISI MADE MULTIPLE FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21636355 ↗MalfunctionNov 28, 2024Material Fragmentation; Material Fragmentation
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20803725 ↗MalfunctionAug 15, 2024Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE DEBAKEY FORCEPS INSTRUMENT WRIST CABLE BROKE. A BACKUP INSTRUMENT OF THE SAME KIND WAS USED TO COMPLETE THE PROCEDURE WITH NO PATIENT HARM. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE SITE AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE INSTRUMENT WAS INSPECTED PRIOR TO USE. THERE WAS NO DAMAGE OR ANYTHING OUT OF THE ORDINARY OBSERVED. THE ISSUE OCCURRED DURING A MITRAL VALVE REPAIR. THE INSTRUMENT DID NOT COLLIDE WITH ANY OTHER INSTRUMENT OR TOOL DURING THE PROCEDURE. THERE WERE NO ISSUES RELATED TO OPENING OR CLOSING THE GRIPS, LEFT AND RIGHT (YAW) MOTION OF THE GRIPS, OR UP AND DOWN (PITCH) MOTION OF THE WRIST. ONE CABLE, WHICH CONTROLS THE OPENING AND CLOSING OF THE JAWS, WAS VISIBLY PROTRUDING FROM THE DISTAL END OF THE INSTRUMENT. NO FRAGMENT FALL INSIDE THE PATIENT'S ANATOMY. NO VIDEO OR PHOTOGRAPHIC IMAGES OF THE DEVICES ARE AVAILABLE FOR ISI TO REVIEW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19988916 ↗MalfunctionAug 10, 2024Material Fragmentation; Material Fragmentation
REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19957005 ↗MalfunctionMay 2, 2024Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED REVASCULARIZATION OF THE MYOCARDIUM SURGICAL PROCEDURE, THE DEBAKEY FORCEPS INSTRUMENT RELEASED A FILAMENT FROM THE STEEL CABLE AT ITS TIP JOINT, ON ITS SECOND USE, LEAVING 8 LIVES LEFT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE SITE AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE INSTRUMENT WAS INSPECTED PRIOR TO USE AND NO DAMAGE OR ANYTHING OUT OF THE ORDINARY WAS FOUND BEFORE THE PROCEDURE. THE SURGICAL TASK WAS COMPLETING THE SURGERY, MYOCARDIAL REVASCULARIZATION SURGERY. THE INSTRUMENT DID NOT COLLIDE WITH ANY OTHER INSTRUMENT OR TOOL. THERE WERE NO ISSUES OPENING/CLOSING THE GRIPS, LEFT/RIGHT (YAW) MOTION, OR UP/DOWN (PITCH) MOTION. ONE CABLE WAS VISIBLY PROTRUDING FROM THE DISTAL END, A PHOTO WAS PREVIOUSLY SENT. THE CUSTOMER COULD NOT ANSWER WHICH MOTION THE PROTRUDING CABLE CONTROLS. ONE MICRO FILAMENT BROKE BUT DID NOT RELEASE FROM THE TWEEZERS. A FRAGMENT DID NOT FALL INTO THE PATIENT. THE INSTRUMENT WAS AVAILABLE FOR RETURN AND WAS ALREADY SENT LAST WEEK. AN IMAGE WAS ATTACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19227332 ↗MalfunctionAug 3, 2022Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED TONGUE BASE RESECTION - BENIGN SURGICAL PROCEDURE, THE DEBAKEY FORCEPS INSTRUMENT HAD A BROKEN CABLE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) MADE MULTIPLE FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15158037 ↗MalfunctionDec 6, 2020Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI-ASSISTED RADICAL TONSILLECTOMY SURGICAL PROCEDURE, THE DEBAKEY FORCEPS JAWS WOULD NOT MOVE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10955651 ↗
FDA clearance and classification
The FDA regulates product code NAY (system,surgical,computer controlled instrument) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the EndoWrist 470036
Has the EndoWrist 470036 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0852-2025 (Class II), began on Dec 19, 2024: Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the EndoWrist 470036?
The FDA's MAUDE database lists 19 adverse event reports for this device, received between Dec 6, 2020 and Mar 31, 2026. In those reports, the most common specific device problems are “Material Split, Cut or Torn” (7), “Material Fragmentation” (6) and “Material Frayed” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the EndoWrist 470036?
None of the reports are classified as deaths, 0 as injuries and 19 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the EndoWrist 470036, and is it still sold?
The FDA's device database lists Intuitive Surgical, Inc. as the company responsible for it (brand EndoWrist, model 470036). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other System,surgical,computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,308 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,204 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,630 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,325 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls233 deaths5,369 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,342 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,802 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,281 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page