EndoWrist 470049Recalls & reported problems

Intuitive Surgical, Inc. · CADIERE FORCEPS · Class 2

All models of EndoWrist (62)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Intuitive EndoWrist Cadiere forceps is a robot-controlled surgical instrument used with da Vinci systems. The 238 reports mostly describe cables that break, fray or come loose, and in one case a wire was seen hanging out of the wrist during surgery, with the procedure finished using a backup instrument. The reports include 53 injuries and no deaths. There is 1 open recall, started in December 2024, over frayed or broken grip cables on reusable instruments.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
239
Deaths reported
0
Injuries reported
53
Recalls naming it
1

Summary

The FDA's MAUDE database contains 239 adverse event reports that list this exact device (DI 00886874112250), received between Apr 25, 2017 and Sep 25, 2026 — classified as 53 injuries, 186 malfunctions.

The most frequently reported specific device problem is “Break” (106 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 4 reportsApr 2025: 1 reportsMay 2025: 4 reportsJun 2025: 7 reportsJul 2025: 1 reportsAug 2025: 4 reportsSep 2025: 3 reportsOct 2025: 4 reportsNov 2025: 2 reportsDec 2025: 1 reportsJan 2026: 3 reportsFeb 2026: 13 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 19, 2024Class IIOpen, Classified

    Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

    Z-0853-2025 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break106
  • Detachment of Device or Device Component79
  • Material Split, Cut or Torn35
  • Material Fragmentation13
  • Material Frayed13
  • Difficult to Open or Close7
  • Detachment Of Device Component5
  • Defective Component4

Patient problems

  • No Clinical Signs, Symptoms or Conditions138
  • No Known Impact Or Consequence To Patient65
  • Insufficient Information14
  • Foreign Body In Patient11
  • Device Embedded In Tissue or Plaque6
  • Injury2
  • Bowel Perforation1
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 25 most recent of 239 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026Unintended Movement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905624 ↗
  • MalfunctionSep 23, 2026Material Split, Cut or Torn

    Patient problems: Insufficient Information

    MAUDE report 50862957 ↗
  • MalfunctionAug 11, 2026No Apparent Adverse Event

    It was reported that the customer experienced an issue with an Intuitive product. The customer reported event has no report of patient injury.

    Patient problems: Insufficient Information

    MAUDE report 50203260 ↗
  • MalfunctionApr 14, 2026Material Split, Cut or Torn

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24878420 ↗
  • MalfunctionApr 7, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24810268 ↗
  • MalfunctionFeb 26, 2026Material Split, Cut or Torn

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 24465861 ↗
  • MalfunctionFeb 26, 2026Material Frayed

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CADIERE FORCEPS INSTRUMENT HAD A WIRE HANGING OUT OF THE WRIST. THE PROCEDURE WAS COMPLETED AS PLANNED WITH NO REPORTED INJURY USING A BACKUP INSTRUMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24465048 ↗
  • MalfunctionFeb 25, 2026Material Frayed

    IT WAS REPORTED THAT THERE WAS A FRAYED CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24449887 ↗
  • MalfunctionFeb 24, 2026Material Split, Cut or Torn

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24437132 ↗
  • MalfunctionFeb 24, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24425455 ↗
  • MalfunctionFeb 23, 2026Material Frayed

    IT WAS REPORTED THAT THERE WAS A FRAYED CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24422732 ↗
  • MalfunctionFeb 23, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24422651 ↗
  • MalfunctionFeb 23, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24422636 ↗
  • MalfunctionFeb 20, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24407715 ↗
  • MalfunctionFeb 20, 2026Material Split, Cut or Torn

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 24407535 ↗
  • MalfunctionFeb 18, 2026Material Split, Cut or Torn

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 24374317 ↗
  • MalfunctionFeb 14, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24340370 ↗
  • MalfunctionFeb 5, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24273978 ↗
  • MalfunctionJan 22, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24154424 ↗
  • MalfunctionJan 9, 2026Material Frayed

    IT WAS REPORTED THAT THERE WAS A FRAYED CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24029705 ↗
  • MalfunctionJan 8, 2026Material Frayed

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24014797 ↗
  • MalfunctionDec 5, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO HTTPS://EMDR.FDA.GOV/EMDR/FORMINBOX/INDEX.JSF#REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23735074 ↗
  • MalfunctionNov 27, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23668042 ↗
  • MalfunctionNov 5, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23481407 ↗
  • MalfunctionOct 31, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 23442444 ↗

Other System,surgical,computer controlled instrument devices