EndoWrist 470184Recalls & reported problems
Intuitive Surgical, Inc. · PERMANENT CAUTERY SPATULA · Class 2
All models of EndoWrist (62)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Intuitive EndoWrist permanent cautery spatula is a robotic surgical instrument used with da Vinci systems. Its 705 reports, with 55 injuries and no deaths, often describe instrument parts that split, tear, break or detach, and some that fragment or arc. Recent reports include tips that broke during surgery and insulation chipped during reprocessing. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 711
- Deaths reported
- 0
- Injuries reported
- 56
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 711 adverse event reports that list this exact device (DI 00886874112328), received between Jun 29, 2017 and Sep 30, 2026 — classified as 56 injuries, 654 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (242 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Split, Cut or Torn242
- Detachment of Device or Device Component138
- Break127
- Material Fragmentation84
- Thermal Decomposition of Device42
- Arcing28
- Material Frayed21
- Energy Output Problem12
Patient problems
- No Clinical Signs, Symptoms or Conditions477
- Insufficient Information138
- No Known Impact Or Consequence To Patient56
- Foreign Body In Patient20
- Hemorrhage/Blood Loss/Bleeding6
- Device Embedded In Tissue or Plaque5
- Unspecified Tissue Injury5
- Burn(s)4
Recent adverse event reports
The 25 most recent of 711 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Energy Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50953179 ↗MalfunctionSep 29, 2026Thermal Decomposition of Device
It was reported that during a da Vinci-assisted surgical procedure, noticeable abnormal smoke occurred following the release of energy from the tip of the Permanent Cautery Spatula. The procedure was completed with no reported injury.
Patient problems: Insufficient Information
MAUDE report 50939631 ↗MalfunctionSep 29, 2026Detachment of Device or Device Component
It was reported that there was a broken cable with a da Vinci product. The customer reported event has no report of patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50928030 ↗MalfunctionSep 29, 2026Break
It was reported that prior to the start of a da Vinci-assisted surgical procedure, the Permanent Cautery Spatula instrument articulation/joint was broken. The procedure was completed with no reported injury.Intuitive followed up and obtained additional information via customer checklist. Issue was broken part at the articulation and joint of the instrument. There were five lives left. Procedure was on (b)(6)2026. Procedure was a robotic treatment of endometriosis and lysis of adhesions. A backup instrument was used and no patient harm. Nothing fell in the patient.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50928008 ↗MalfunctionSep 28, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50926104 ↗MalfunctionSep 28, 2026Material Split, Cut or Torn
Patient problems: Insufficient Information
MAUDE report 50924665 ↗InjurySep 28, 2026Arcing
Refer to H11 for Follow-Up information.
Patient problems: Burn(s); Hemorrhage/Blood Loss/Bleeding
MAUDE report 50913513 ↗MalfunctionSep 24, 2026Material Split, Cut or Torn
Patient problems: Insufficient Information
MAUDE report 50891232 ↗OtherSep 23, 2026Arcing
It was reported that during a da Vinci-assisted supraglottic laryngectomy, energy from the Permanent Cautery Spatula dispersed through the pulley cables, resulting in an injury to the interior of the patient's throat. The instrument had been inspected prior to use with no abnormalities observed. An arc was formed between one of the wires at the distal end of the instrument and the affected area of the throat at the time Monopolar Coag energy was activated. The tip of the Permanent Cautery Spatula did not come into contact with any other instrument or tool during the procedure. The procedure was completed with a 15-30-minute delay. The injury did not result in any complications, long-term sequelae, or the need for additional intervention, and the patient recovered well.
Patient problems: Unspecified Tissue Injury
MAUDE report 50867261 ↗MalfunctionSep 21, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50849363 ↗InjurySep 12, 2026Detachment of Device or Device Component
It was reported that during a da Vinci-assisted pulmonary lobectomy procedure, a pin from the distal tip area of the Permanent Cautery Spatula (PCS) instrument fell off inside the patient and was retrieved during the same procedure. The customer confirmed that the fragment (pin) was retrieved. No additional surgical procedures were required to retrieve the fragment. Additionally, a chest x-ray did not find any retained fragments. The procedure was completed with no delay.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50745064 ↗MalfunctionSep 3, 2026Material Split, Cut or Torn
Refer to H11 for Follow-Up information.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50596979 ↗MalfunctionSep 2, 2026Break; Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50576044 ↗MalfunctionAug 13, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50242938 ↗MalfunctionAug 11, 2026Material Split, Cut or Torn
Refer to H11 for Follow-Up information.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50216296 ↗MalfunctionAug 10, 2026Break
It was reported that during central processing, the Permanent Cautery Spatula instrument's insulated area near the tip was chipped. No known impact or patient consequence was reported.ISI followed up with the initial reporter and obtained the following additional information: There was a visible section missing at the insulated area. It wasn't used on a case as this was visually noticed beforehand. There was no arcing.
Patient problems: Insufficient Information
MAUDE report 50196616 ↗MalfunctionJun 1, 2026Thermal Decomposition of Device
IT WAS REPORTED THAT THE PERMANENT CAUTERY SPATULA INSTRUMENT TIP WAS DAMAGED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR PATIENT INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: NO REPORT OF MATERIAL BROKEN OFF OR DETACHED. THERE WAS THE LOCATION THAT LOOKED A SLIGHT CRACK OR SCRATCH ON THE TIP. THERE WERE NO NOTICEABLE DEFECTS WITH THE CABLE. NO PHOTOGRAPHIC IMAGES OF THE INSTRUMENT AVAILABLE FOR REVIEW.
Patient problems: Insufficient Information
MAUDE report 25352532 ↗MalfunctionMay 27, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: Insufficient Information
MAUDE report 25307378 ↗MalfunctionMay 26, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: Insufficient Information
MAUDE report 25291517 ↗MalfunctionMay 15, 2026Break
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE TIP OF A PERMANENT CAUTERY SPATULA INSTRUMENT HAS BEEN DAMAGED. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: Insufficient Information
MAUDE report 25179993 ↗MalfunctionApr 30, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25051176 ↗MalfunctionApr 17, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT AFTER A DA VINCI-ASSISTED RADICAL SURGICAL PROCEDURE, THE PERMANENT CAUTERY SPATULA INSTRUMENT WAS FOUND DEFECTIVE. THERE WAS NO REPORT OF PATIENT HARM DUE TO THIS EVENT. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION ON 13-APR-2026: SOME ARMS WERE MISSING FROM THE WHEELS THAT ATTACH TO THE MECHANISM CONNECTING TO THE ROBOT ARM. SOME ARMS HAD EXPOSED WIRING AND WERE NON-FUNCTIONAL. SOME ARMS HAD CABLING THAT HAD SNAPPED AT THE ELBOW OF THE DEVICE.
Patient problems: Insufficient Information
MAUDE report 24918179 ↗MalfunctionApr 17, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24917318 ↗MalfunctionApr 15, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: Insufficient Information
MAUDE report 24890696 ↗MalfunctionApr 15, 2026Thermal Decomposition of Device
IT WAS REPORTED THAT PRIOR TO THE START (POST-ANESTHESIA) OF A DA VINCI-ASSISTED TONGUE BASE RESECTION - MALIGNANT SURGICAL PROCEDURE, THE PERMANENT CAUTERY SPATULA INSULATOR AT WRIST WAS DEGRADED AND HAD BURNT MARKS ON IT. THE PROCEDURE WAS COMPLETED AS PLANNED WITH NO REPORTED INJURY USING A BACKUP INSTRUMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24890058 ↗
FDA clearance and classification
The FDA regulates product code NAY (system,surgical,computer controlled instrument) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the EndoWrist 470184
Has the EndoWrist 470184 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00886874112328), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the EndoWrist 470184?
The FDA's MAUDE database lists 711 adverse event reports for this device, received between Jun 29, 2017 and Sep 30, 2026. In those reports, the most common specific device problems are “Material Split, Cut or Torn” (242), “Detachment of Device or Device Component” (138) and “Break” (127); the most common patient problems are “Foreign Body In Patient” (20), “Hemorrhage/Blood Loss/Bleeding” (6) and “Device Embedded In Tissue or Plaque” (5). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the EndoWrist 470184?
None of the reports are classified as deaths, 56 as injuries and 654 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the EndoWrist 470184, and is it still sold?
The FDA's device database lists Intuitive Surgical, Inc. as the company responsible for it (brand EndoWrist, model 470184). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other System,surgical,computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,308 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,204 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,630 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,325 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls233 deaths5,369 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,342 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,802 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,281 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page