EndoWrist 470194Recalls & reported problems
Intuitive Surgical, Inc. · MEGA NEEDLE DRIVER · Class 2
All models of EndoWrist (62)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Intuitive EndoWrist Mega Needle Driver is the subject of 1,047 adverse event reports, which mention no deaths and 16 injuries. Most reports describe cables or materials that split, tore, fragmented or frayed. A few mention material left in the patient. 1 open recall, started in December 2024, concerns frayed or broken grip cables on reusable instruments with jaws.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,043
- Deaths reported
- 0
- Injuries reported
- 16
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1,043 adverse event reports that list this exact device (DI 00886874112342), received between Jun 1, 2017 and Sep 30, 2026 — classified as 16 injuries, 1,027 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (620 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Dec 19, 2024Open, Classified
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.
Z-0858-2025 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Split, Cut or Torn620
- Material Fragmentation197
- Material Frayed97
- Break44
- Detachment of Device or Device Component39
- No Apparent Adverse Event25
- Unintended Movement11
- Difficult to Open or Close8
Patient problems
- No Clinical Signs, Symptoms or Conditions724
- Insufficient Information303
- No Known Impact Or Consequence To Patient12
- Foreign Body In Patient7
- Bowel Perforation1
- Device Embedded In Tissue or Plaque1
- Unspecified Tissue Injury1
Recent adverse event reports
The 25 most recent of 1,043 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954073 ↗MalfunctionSep 30, 2026Material Split, Cut or Torn
It was reported that there was a broken cable with a da Vinci product. The customer reported event has no report of patient injury.Review of the provided image is consistent with broken cable. Root cause of the failure mode cannot be confirmed without the returned device.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942380 ↗MalfunctionSep 25, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50905317 ↗MalfunctionSep 25, 2026Material Split, Cut or Torn
It was reported that the Mega Needle Driver instrument had a broke tip. The procedure was completed with no reported injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50893577 ↗MalfunctionSep 23, 2026Break
Patient problems: Insufficient Information
MAUDE report 50875949 ↗MalfunctionSep 23, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50874586 ↗MalfunctionSep 22, 2026Material Split, Cut or Torn
It was reported that there was a broken cable with a da Vinci product. The customer reported event has no report of patient injury.
Patient problems: Insufficient Information
MAUDE report 50856395 ↗MalfunctionAug 24, 2026Material Split, Cut or Torn
Patient problems: Insufficient Information
MAUDE report 50420624 ↗MalfunctionAug 20, 2026Material Split, Cut or Torn
Patient problems: Insufficient Information
MAUDE report 50359548 ↗MalfunctionAug 14, 2026Material Frayed; Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50254120 ↗MalfunctionAug 11, 2026No Apparent Adverse Event
It was reported that the customer experienced an issue with an Intuitive product. The customer reported event has no report of patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50216717 ↗MalfunctionAug 7, 2026Material Frayed
It was reported that there was a frayed cable with a da Vinci product. The customer reported event has no report of patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50184397 ↗MalfunctionMay 28, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25324997 ↗MalfunctionMay 20, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: Insufficient Information
MAUDE report 25235316 ↗MalfunctionMay 18, 2026Material Frayed
IT WAS REPORTED THAT THERE WAS A FRAYED CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25196983 ↗MalfunctionMay 15, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25179975 ↗MalfunctionMay 14, 2026No Apparent Adverse Event
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: Insufficient Information
MAUDE report 25178169 ↗MalfunctionMay 12, 2026Break
IT WAS REPORTED THAT DURING CENTRAL PROCESSING, THE JAWS OF A MEGA NEEDLE DRIVER INSTRUMENT WERE MISALIGNED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: NO FRAGMENTS FELL INTO PATIENT'S ANATOMY. THE DAMAGE TO THE INSTRUMENT OCCURRED OUTSIDE OF A SURGICAL PROCEDURE.
Patient problems: Insufficient Information
MAUDE report 25155498 ↗MalfunctionMay 11, 2026Break
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY. ISI FOLLOWED UP WITH THE INITIAL REPORTER TO OBTAIN ADDITIONAL INFORMATION, HOWEVER, NO FURTHER DETAILS ARE AVAILABLE REGARDING THE REPORTED EVENT.
Patient problems: Insufficient Information
MAUDE report 25143853 ↗MalfunctionMay 9, 2026Break
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY. ISI MADE MULTIPLE FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25131709 ↗MalfunctionMay 7, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: Insufficient Information
MAUDE report 25118773 ↗MalfunctionMay 6, 2026Unintended Movement
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED BENIGN HYSTERECTOMY SURGICAL PROCEDURE, THE MEGA NEEDLE DRIVER INSTRUMENT WAS GOING THE OPPOSITE WAY. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: NO PATIENT INJURY OR VISIBLE DAMAGE TO THE INSTRUMENT, INCLUDING JAWS, TIPS, OR CABLES, WAS REPORTED. THE ISSUE WAS RESOLVED BY REPLACING THE INSTRUMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25107432 ↗MalfunctionMay 6, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A FRAYED CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25107307 ↗MalfunctionMay 4, 2026Material Integrity Problem
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25074310 ↗MalfunctionApr 29, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25029375 ↗
Other System,surgical,computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,225 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,156 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,581 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,302 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls235 deaths5,316 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,310 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,770 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,273 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page