None 480275Recalls & reported problems
Intuitive Surgical, Inc. · HARMONIC ACE · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Intuitive Harmonic Ace is a computer-controlled surgical instrument used with da Vinci robotic systems. Its 2,321 reports often describe parts breaking off or detaching, and some mention fragments ending up inside patients. They also mention 889 injuries and 1 death. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 2,325
- Deaths reported
- 1
- Injuries reported
- 899
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2,325 adverse event reports that list this exact device (DI 00886874112731), received between Mar 23, 2018 and Sep 30, 2026 — classified as 1 death, 899 injuries, 1,425 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1,114 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1,114
- Break752
- No Device Output201
- Material Fragmentation183
- Output Problem75
- Material Integrity Problem30
- Thermal Decomposition of Device27
- Material Separation23
Patient problems
- No Clinical Signs, Symptoms or Conditions1,750
- Foreign Body In Patient390
- Insufficient Information140
- Device Embedded In Tissue or Plaque69
- No Known Impact Or Consequence To Patient20
- Hemorrhage/Blood Loss/Bleeding10
- Unspecified Tissue Injury7
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
Recent adverse event reports
The 25 most recent of 2,325 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Material Fragmentation
Refer to H11 for Follow-Up information.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954553 ↗MalfunctionSep 30, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50953592 ↗InjurySep 30, 2026Material Fragmentation
It was reported that during a da Vinci-assisted nissen fundoplication procedure, a few minutes after using the coagulation function, the jaw of the Harmonic ACE instrument was observed to have broken. The customer confirmed that a piece broke off and fell inside the patient. The piece was identified and removed immediately during the same procedure which was completed with no delays.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50952792 ↗MalfunctionSep 30, 2026No Apparent Adverse Event
It was reported that during a da Vinci-assisted surgical procedure, a 8mm Harmonic Ace instrument had a blade activation failure. The procedure was completing as planned with no reported injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942385 ↗MalfunctionSep 28, 2026Break
It was reported that during a da Vinci-assisted surgical procedure, the cutting tip/blade of the Harmonic Ace instrument was fractured. No fragment fell into the patient. The user completed the procedure and no known impact or patient consequence was reported.ISI made multiple follow-up attempts to obtain additional information. However, no further details have been received as of the date of this report.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50910833 ↗MalfunctionSep 25, 2026No Apparent Adverse Event
Patient problems: Insufficient Information
MAUDE report 50905854 ↗MalfunctionSep 25, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50905024 ↗InjurySep 25, 2026Material Fragmentation
It was reported that during a da Vinci-assisted procedure, a screw on a Harmonic Ace instrument was observed to have loosened or broken and a fragment fell inside the patient. The fragment was retrieved during the same surgery.\\The customer stated that the broken fragment was immediately retrieved using a long forceps instrument along with another robotic arm. To confirm complete retrieval, the surgical site was thoroughly irrigated and carefully inspected using the da Vinci robotic endoscope, ensuring no fragments remained in the area. A postoperative ultrasound examination was conducted following the visual confirmation that no residual fragment was present.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50899446 ↗MalfunctionSep 24, 2026Break; Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50890605 ↗InjurySep 24, 2026Material Fragmentation
It was reported that during a da Vinci-assisted donor hepatectomy procedure, the Harmonic ACE instrument blade broke and damage was observed at the proximal end of the active blade. A fragment reportedly fell inside the patient. The instrument was new and had been confirmed to be intact prior to use. Following the break, the fragment was successfully retrieved from the patient and the procedure was completed.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50887079 ↗InjurySep 24, 2026Material Fragmentation
It was reported that during a da Vinci-assisted procedure, the blade of the Harmonic Ace instrument was broken, and a fragment fell inside the patient. The fragment was retrieved during the same procedure. The procedure was completed. No post-operative tests were performed. \\Intuitive Surgical, Inc. (ISI) made multiple follow-up attempts to obtain additional information. However, no further details have been received as of the date of this report.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50885141 ↗MalfunctionSep 23, 2026No Device Output
Patient problems: Insufficient Information
MAUDE report 50875983 ↗MalfunctionSep 23, 2026Break
Patient problems: Insufficient Information
MAUDE report 50875895 ↗MalfunctionSep 23, 2026Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50875872 ↗InjurySep 21, 2026Material Fragmentation
It was reported that during a da Vinci-assisted hepatectomy procedure, the ultrasonic blade of the Harmonic Ace instrument fractured while the surgeon was resecting liver tissue, resulting in a fragment falling inside the patient. The fragment was retrieved during the same procedure using an laparoscopic instrument, and the procedure was completed without delay. \\The customer confirmed that retrieval was visually verified by the surgeon, assistant, and nurse; however, the retrieved fragment was subsequently discarded during cleaning and disinfection of the instrument in the supply room.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50848368 ↗MalfunctionSep 11, 2026Detachment of Device or Device Component
Patient problems: Insufficient Information
MAUDE report 50743441 ↗InjurySep 8, 2026Material Fragmentation
It was reported that during a da Vinci-assisted segmental pancreatectomy procedure, the front end of the Harmonic ACE instrument was fractured, and a fragment fell into the patient. The customer confirmed that the fractured part (fragment) was retrieved during the same procedure and no fragment remained in the patient. The procedure was completed. Intuitive Surgical Inc., (ISI) made multiple follow-ups attempts to obtain additional information. However, no further details have been received as of the date of this report.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50676356 ↗InjurySep 4, 2026Material Fragmentation
Refer to H11 for Follow-Up information.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50623043 ↗MalfunctionSep 3, 2026Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50579771 ↗InjuryAug 28, 2026Material Fragmentation
It was reported that during a da Vinci-assisted surgical procedure, the customer encountered an error message stating ¿Reduce pressure on the blade¿ and the blade on a Harmonic Ace instrument subsequently broke. A fragment fell inside the patient but was retrieved during the same procedure. There were no reports of patient injury; however, the event resulted in a 15¿30 minute delay.\\Intuitive Surgical, Inc. (ISI) made multiple follow-up attempts to obtain additional information. However, no further details have been received as of the date of this report.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50508274 ↗MalfunctionAug 27, 2026Break
It was reported that during a da Vinci-assisted surgical procedure, the blade of the Harmonic Ace instrument was broken. No fragment fell into the patient. The user completed the procedure and no known impact or patient consequence was reported.\\Intuitive Surgical, Inc. (ISI) followed up with the initial reporter and obtained the following additional information: The broken blade was detached from the instrument.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50486331 ↗MalfunctionAug 26, 2026Thermal Decomposition of Device
It was reported that during a da Vinci-assisted surgical procedure, the Harmonic ACE teflon pad was melted. The user completed the procedure with no further issue reported. No fragment fell inside the patient. No known impact or patient consequence was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50463648 ↗MalfunctionAug 21, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50382281 ↗InjuryAug 21, 2026Material Fragmentation
It was reported that during a da Vinci¿assisted low anterior resection procedure, the Harmonic Ace blade tip fractured and a fragment fell inside the patient. The fragment was retrieved using laparoscopic instruments during the same procedure which was completed robotically.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50380989 ↗InjuryAug 20, 2026Material Fragmentation
It was reported that during a da Vinci-assisted low anterior resection procedure, the blade tip of the Harmonic Ace instrument fractured and a fragment fell inside the patient. All fragments were retrieved during the same procedure which was completed robotically.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50356478 ↗
FDA clearance and classification
The FDA regulates product code NAY (system,surgical,computer controlled instrument) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the None 480275
Has the None 480275 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00886874112731), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the None 480275?
The FDA's MAUDE database lists 2,325 adverse event reports for this device, received between Mar 23, 2018 and Sep 30, 2026. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (1,114), “Break” (752) and “No Device Output” (201); the most common patient problems are “Foreign Body In Patient” (390), “Device Embedded In Tissue or Plaque” (69) and “Hemorrhage/Blood Loss/Bleeding” (10). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the None 480275?
1 report is classified as a death, 899 as injuries and 1,425 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the None 480275, and is it still sold?
The FDA's device database lists Intuitive Surgical, Inc. as the company responsible for it (brand None, model 480275). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other System,surgical,computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,308 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,204 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,630 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,325 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls233 deaths5,369 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,342 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,802 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,281 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page