Intuitive Surgical, Inc. · CAMERA INSTRUMENT SHEATH · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00886874114353), received between Feb 1, 2024 and Dec 11, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation2
  • Material Split, Cut or Torn2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 11, 2025Material Fragmentation; Material Fragmentation

    THE SHEATH ACCESSORY WAS RETURNED TO ISI FOR EVALUATION WITH NO REPORTED ISSUE. AN INVESTIGATION HAS BEEN COMPLETED. THE SHEATH WAS RETURNED STILL INSTALLED ON THE ENDOSCOPE. VISUAL INSPECTION OF THE SHEATH FOUND IT TO BE DAMAGED. MULTIPLE TEARS MEASURING APPROXIMATELY 0.77 MM X 4.91 MM WERE FOUND ALONG THE LENGTH OF THE SHEATH. THERE WAS NO MATERIAL MISSING AS A RESULT. THE ROOT CAUSE OF THIS FAILURE IS ATTRIBUTED TO USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23784269 ↗
  • MalfunctionMay 21, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT PRIOR TO THE START OF A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CUSTOMER FOUND THAT SINGLE-PORT (SP) CAMERA SHEATH HAD A HOLE. UPON REMOVING THE SHEATH, THE CONNECTION PART CRUMBLED AND DETACHED. THE CUSTOMER USED A BACKUP SHEATH TO CONTINUE WITH THE PLANNED PROCEDURE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE NURSE AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: PORTS WERE NOT PLACED WHEN THE ISSUE WAS IDENTIFIED. THERE WAS NO FRAGMENT FALLING INSIDE THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22058282 ↗
  • InjuryApr 28, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED THE RIGHT PARTIAL NEPHRECTOMY SURGICAL PROCEDURE, THE CAMERA SHEATH SHREDDED, AND A PIECE WAS HANGING. THE CUSTOMER WAS UNABLE TO ACCESS THE SHEATH INITIALLY DUE TO THE CLAP TIME. THE CUSTOMER PERFORMED AN X-RAY, AND IT WAS CONFIRMED THAT THERE WERE NO FRAGMENTS IN THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21917415 ↗
  • MalfunctionFeb 1, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    THE CAMERA SHEATH ACCESSORY WAS AN INCIDENTAL RETURN INCLUDED WITH RMA 1505840 WHICH ALLEGES THAT THE FIREFLY CAMERA WAS OUT OF LIVES. NO ALLEGATION WAS MADE AGAINST THIS PRODUCT. INTUITIVE SURGICAL INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND THE FOLLOWING ADDITIONAL INFORMATION WAS OBTAINED: THERE WERE NO REPORTS OF FRAGMENTS FALLING INTO THE PATIENT WHEN THE CAMERA SHEATH WAS LAST USED.

    Patient problems: Insufficient Information

    MAUDE report 18620454 ↗

Other System, surgical, computer controlled instrument devices

All System, surgical, computer controlled instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page