Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00886874114353), received between Feb 1, 2024 and Dec 11, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Material Fragmentation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Fragmentation2
- Material Split, Cut or Torn2
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 11, 2025Material Fragmentation; Material Fragmentation
THE SHEATH ACCESSORY WAS RETURNED TO ISI FOR EVALUATION WITH NO REPORTED ISSUE. AN INVESTIGATION HAS BEEN COMPLETED. THE SHEATH WAS RETURNED STILL INSTALLED ON THE ENDOSCOPE. VISUAL INSPECTION OF THE SHEATH FOUND IT TO BE DAMAGED. MULTIPLE TEARS MEASURING APPROXIMATELY 0.77 MM X 4.91 MM WERE FOUND ALONG THE LENGTH OF THE SHEATH. THERE WAS NO MATERIAL MISSING AS A RESULT. THE ROOT CAUSE OF THIS FAILURE IS ATTRIBUTED TO USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23784269 ↗MalfunctionMay 21, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT PRIOR TO THE START OF A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CUSTOMER FOUND THAT SINGLE-PORT (SP) CAMERA SHEATH HAD A HOLE. UPON REMOVING THE SHEATH, THE CONNECTION PART CRUMBLED AND DETACHED. THE CUSTOMER USED A BACKUP SHEATH TO CONTINUE WITH THE PLANNED PROCEDURE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE NURSE AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: PORTS WERE NOT PLACED WHEN THE ISSUE WAS IDENTIFIED. THERE WAS NO FRAGMENT FALLING INSIDE THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22058282 ↗InjuryApr 28, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED THE RIGHT PARTIAL NEPHRECTOMY SURGICAL PROCEDURE, THE CAMERA SHEATH SHREDDED, AND A PIECE WAS HANGING. THE CUSTOMER WAS UNABLE TO ACCESS THE SHEATH INITIALLY DUE TO THE CLAP TIME. THE CUSTOMER PERFORMED AN X-RAY, AND IT WAS CONFIRMED THAT THERE WERE NO FRAGMENTS IN THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21917415 ↗MalfunctionFeb 1, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
THE CAMERA SHEATH ACCESSORY WAS AN INCIDENTAL RETURN INCLUDED WITH RMA 1505840 WHICH ALLEGES THAT THE FIREFLY CAMERA WAS OUT OF LIVES. NO ALLEGATION WAS MADE AGAINST THIS PRODUCT. INTUITIVE SURGICAL INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND THE FOLLOWING ADDITIONAL INFORMATION WAS OBTAINED: THERE WERE NO REPORTS OF FRAGMENTS FALLING INTO THE PATIENT WHEN THE CAMERA SHEATH WAS LAST USED.
Patient problems: Insufficient Information
MAUDE report 18620454 ↗
Other System, surgical, computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,225 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,156 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,581 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,302 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls235 deaths5,316 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,310 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,770 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,273 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page