daVinci 5 380746Recalls & reported problems
Intuitive Surgical, Inc. · Vision Side Cart · Class 2
All models of daVinci 5 (3)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This da Vinci 5 vision side cart is part of the Intuitive robotic surgery system. Its 460 reports mostly describe on-screen prompts that will not clear. The reports include 2 deaths and 39 injuries, and rose to 391 in the last 12 months from 56 in the 12 months before. There is 1 open recall, started in April 2026, because a handheld camera can switch on by accident when staff try to turn off its light.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 457
- Deaths reported
- 2
- Injuries reported
- 40
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 457 adverse event reports that list this exact device (DI 00886874119662), received between May 2, 2024 and Sep 28, 2026 — classified as 2 deaths, 40 injuries, 415 malfunctions.
The most frequently reported specific device problem is “Visual Prompts will not Clear” (248 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Apr 24, 2026Class IIOpen, Classified
Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the handheld camera to the robotic endoscope.
Z-2310-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Visual Prompts will not Clear248
- Output Problem104
- Failure to Power Up24
- Connection Problem11
- Energy Output Problem9
- Pressure Problem9
- Unexpected Shutdown9
- Adverse Event Without Identified Device or Use Problem6
Patient problems
- No Clinical Signs, Symptoms or Conditions379
- Insufficient Information41
- Unspecified Tissue Injury25
- Hemorrhage/Blood Loss/Bleeding8
- Burn(s)4
- Abnormal Blood Gases2
- Cardiac Arrest2
- Abdominal Pain1
Recent adverse event reports
The 25 most recent of 457 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50926058 ↗MalfunctionSep 28, 2026Visual Prompts will not Clear
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50925269 ↗MalfunctionSep 28, 2026Visual Prompts will not Clear
It was reported that during a da Vinci assisted surgical procedure, the customer contacted the Technical Support Engineer (TSE) and reported that errors 170 and 31089 were occurring. A new blue fiber cable had been opened, a power cycle had been performed, and the blue fiber cable had been reconnected prior to contacting support. At the time of the call, ports had been placed but the system had not yet been docked. Upon power up, neither blue cable port on the back of the Vision Side Cart (VSC) displayed blue LEDs. Due to a patient being on the table, the decision was made to bring in a different vision cart rather than continue swapping cables. The replacement vision cart was connected and the system powered on without errors for approximately three minutes before faulting again with the same 31089 errors. The backup blue fiber cable was then attempted, but the errors returned. A replacement Patient Side Cart (PSC) was subsequently obtained and connected alongside the replacement vision cart, while the original Surgeon Side Console (SSC)remained connected. The system was considered down. The procedure was converted to another DV system.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50910824 ↗MalfunctionSep 26, 2026Visual Prompts will not Clear
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50906717 ↗MalfunctionSep 9, 2026Output Problem
Patient problems: Insufficient Information
MAUDE report 50701625 ↗MalfunctionAug 22, 2026Display or Visual Feedback Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50389409 ↗MalfunctionAug 13, 2026Visual Prompts will not Clear
Refer to H11 for Follow-Up information.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50242646 ↗MalfunctionAug 12, 2026Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50224871 ↗InjuryAug 7, 2026Display or Visual Feedback Problem
It was reported that during a da Vinci-assisted abdonimoperineal resection procedure, the operating room staff called Intuitive Surgical Inc. (ISI) Technical Support Engineers (TSE) to state the left eye image was split on both Surgeon Side Console (SSC) 1 and 2. The customer replaced the endoscope, but the issue remained. At the time of the call the customer was unable to power cycle the system due to active bleeding from the renal artery. It was reported that the bleeding from the renal artery started prior to the reported event; however, the split image on the left eye did cause it to take longer to suture the artery and stop the bleeding. The estimated blood loss was not provided; however, it was reported the patient received 3 units of blood and was admitted intubated to the Intensive Care Unit (ICU) post-operatively. After the bleeding was controlled a second call was made to ISI TSE and it was stated the staff was able to perform a hard power cycle to the system; however, the endoscope would not initiate. A second hard power cycle was performed, and an image was produced for a few moments before it was lost. TSE reviewed live system logs and saw an error from tower fiber por
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50175050 ↗MalfunctionJul 30, 2026Appropriate Device Problem Term/Code Not Available
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50092839 ↗MalfunctionJul 30, 2026Appropriate Device Problem Term/Code Not Available
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50092838 ↗MalfunctionJul 29, 2026Pressure Problem
Refer to H11 for Follow-Up information.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50075740 ↗MalfunctionMay 27, 2026No Display/Image
IT WAS REPORTED THAT DURING A DA VINCI ASSISTED SURGICAL PROCEDURE, STAFF EXPERIENCED A VISION ISSUE IN THE RIGHT EYE OF THE SURGEON CONSOLE. TECHNICAL SUPPORT TROUBLESHOOTING FIRST INVOLVED CONFIRMING THE SITE DID NOT HAVE A BACKUP SCOPE AVAILABLE AND THAT THE SITE DID NOT WANT TO POWER CYCLE THE SYSTEM. TECHNICAL SUPPORT THEN HAD THE SITE REQUEST A BACKUP SCOPE FOR SWAP TESTING. AFTER ANOTHER SCOPE WAS TRIED, THE ISSUE REMAINED UNCHANGED, AND THE SITE PLANNED TO INSTALL THE BACKUP SCOPE ONCE IT ARRIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25309029 ↗MalfunctionMay 12, 2026Visual Prompts will not Clear
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI ASSISTED SURGICAL PROCEDURE, THERE WAS A MESSAGE INDICATING THE CONSOLE WAS NOT CONNECTED. THE STAFF HAD ALREADY PERFORMED A HARD POWER CYCLE ONCE AND THE ERROR PERSISTED. TECHNICAL SUPPORT THEN INSTRUCTED A SECOND HARD POWER CYCLE, BUT THE ISSUE REMAINED AND ERROR 31089 WAS OBSERVED IN THE LOGS, INDICATING A PROBLEM WITH FIBER PORT 3 ON THE TOWER. TECHNICAL SUPPORT DIRECTED THE STAFF TO MOVE THE BLUE FIBER CABLE FROM PORT 3 TO ANOTHER PORT AND REBOOT THE SYSTEM, BUT THE ERROR STILL PERSISTED. THE STAFF PLANNED TO INFORM THE SURGEON AND HAD NOT YET DETERMINED HOW THEY WOULD PROCEED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25155396 ↗MalfunctionMay 7, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A DA VINCI ASSISTED SURGICAL PROCEDURE, THE OPERATING ROOM STAFF ENCOUNTERED A "ROBOT NOT CONNECTED TO TOWER" MESSAGE AT THE START OF A PROCEDURE. LOGS WERE UNAVAILABLE AT THE TIME OF THE CALL. THE TECHNICAL SUPPORT ENGINEER (TSE) INSTRUCTED THE CUSTOMER TO POWER DOWN THE SYSTEM. THE CUSTOMER WAS UNABLE TO IDENTIFY THE SPECIFIC BLUE FIBER PORT ON THE TOWER CONNECTED TO THE ROBOT. THE TSE THEN GUIDED THE CUSTOMER TO RESEAT ALL BLUE CABLE CONNECTIONS, BUT THIS DID NOT RESOLVE THE ISSUE. THE TSE HAD THE CUSTOMER POWER UP THE PATIENT CART AND VISION TOWER IN STANDALONE MODE, AND BOTH COMPONENTS POWERED UP WITHOUT ERRORS. THE TSE INQUIRED IF A BACKUP BLUE FIBER CABLE WAS AVAILABLE, WHICH THE CUSTOMER CONFIRMED. THE TSE DIRECTED THE CUSTOMER TO REPLACE THE ROBOT'S BLUE FIBER CABLE AND CONNECT IT TO A DIFFERENT PORT ON THE TOWER. AFTER SWITCHING THE CONNECTION FROM TOWER BLUE FIBER PORT 3 TO PORT 2, THE SYSTEM BOOTED UP SUCCESSFULLY, ALLOWING THE STAFF TO PROCEED WITH THE CASE. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25118750 ↗MalfunctionApr 29, 2026Visual Prompts will not Clear
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, AN ADVANCED PROCESSOR ERROR 66200 OCCURRED. THE CIRCULATING NURSE STATED THAT THE SURGEON UNDOCKED THE SYSTEM AND CONTINUED THE PROCEDURE WITHOUT THE ROBOT BEFORE CONTACTING TECHNICAL SUPPORT. BEFORE THE CALL, THE NURSE HAD ALREADY POWER CYCLED THE SYSTEM TWICE, BUT THE ERROR PERSISTED. AFTER REVIEWING THIS INFORMATION, TECHNICAL SUPPORT REQUESTED THAT THE ISSUE BE FURTHER TROUBLESHOT ONSITE, AND IT WAS NOTED THAT THE CUSTOMER DECLINED TO USE THE SYSTEM FOR SUBSEQUENT PROCEDURES. THE PROCEDURE WAS CONVERTED TO LAPAROSCOPIC APPROACH WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25029298 ↗MalfunctionApr 28, 2026Visual Prompts will not Clear
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI ASSISTED SURGICAL PROCEDURE, THE CUSTOMER EXPERIENCED NON-RECOVERABLE 40014 ERROR OCCURRING DURING SYSTEM STARTUP WHILE THE SYSTEM WAS CONFIGURED IN DUAL CONSOLE MODE. TECHNICAL SUPPORT HAD THE STAFF REMOVE ONE CONSOLE, AFTER WHICH THE SYSTEM STARTED SUCCESSFULLY AND COMPLETED SELF TEST. AT THE TIME OF TROUBLESHOOTING, THE SYSTEM REMAINED DISCONNECTED SO LOGS WERE NOT AVAILABLE, AND THE SYSTEM CONNECTION WAS REESTABLISHED SHORTLY AFTER THE CALL ENDED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25022930 ↗MalfunctionApr 23, 2026Visual Prompts will not Clear
IT WAS REPORTED THAT PRIOR TO START OF A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CUSTOMER CONTACTED THE TECHNICAL SUPPORT ENGINEER (TSE) AND REPORTED THEY EXPERIENCED MULTIPLE NON-RECOVERABLE FAULTS ON THE SYSTEM, SPECIFICALLY FAULTS 170, 40068, AND 307. THE TSE INSTRUCTED THE CUSTOMER TO PERFORM A HARD POWER CYCLE ON ALL SYSTEM COMPONENTS, BUT THE FAULTS CONTINUED TO RECUR AFTER THE REBOOT. THE SYSTEM WAS UNABLE TO PROVIDE LOGS FOR FURTHER ANALYSIS. THE PROCEDURE WAS ABORTED WITH NO PATIENT INVOLVEMENT. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE PROCEDURE WAS ABORTED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24961866 ↗MalfunctionApr 22, 2026Output Problem
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CUSTOMER CONTACTED THE TECHNICAL SUPPORT ENGINEER (TSE) TO REPORT A RECURRING TOWER OVERHEATING MESSAGE. THE TSE HAD THE CALLER VERIFY THAT THERE WERE NO OBSTRUCTIONS IN FRONT OF THE TOWER, WHICH THE CALLER CONFIRMED, AND ALSO CONFIRMED THERE WAS NO VISIBLE DUST IN THE FRONT OF THE TOWER. THE TSE THEN RECOMMENDED THAT THE CUSTOMER POWER DOWN THE SYSTEM BETWEEN CASES TO ALLOW THE SYSTEM TO COOL OFF, AND THE CALLER AGREED TO INFORM THE STAFF BEFORE THE CALL ENDED. THE PROCEDURE WAS COMPLETING AS PLANNED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24949115 ↗MalfunctionApr 21, 2026Visual Prompts will not Clear
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24946577 ↗MalfunctionApr 21, 2026Unexpected Shutdown
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24934406 ↗MalfunctionApr 20, 2026Visual Prompts will not Clear
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24932752 ↗MalfunctionApr 20, 2026Visual Prompts will not Clear
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24932741 ↗MalfunctionApr 20, 2026Visual Prompts will not Clear
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24932740 ↗InjuryApr 20, 2026Visual Prompts will not Clear
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED MALIGNANT HYSTERECTOMY PROCEDURE, THE CUSTOMER WAS ENCOUNTERING NON-RECOVERABLE SYSTEM ERRORS. AN INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE)) WAS UNABLE TO REVIEW THE SYSTEM LOGS SINCE THE SYSTEM WAS NOT CONNECTED TO THE INTERNET. BEFORE CALLING THE TSE FOR ASSISTANCE, THE CUSTOMER HAD ALREADY RESEATED ALL FIBER CABLE CONNECTIONS EXCEPT AT THE CONSOLE. THE TSE THEN GUIDED THE CUSTOMER TO POWER THE SYSTEM OFF AND RESEAT THE CONSOLE FIBER CABLE. AFTER THIS, THE SYSTEM POWERED BACK ON WITHOUT ERROR, AND THE CUSTOMER NOTED IT HAD PREVIOUSLY POWERED UP NORMALLY BEFORE FAILING AGAIN. THE TSE REMAINED ON THE LINE FOR SEVERAL MINUTES TO MONITOR WHETHER THE ERROR WOULD RECUR WHILE PORTS WERE BEING PLACED, DURING WHICH TIME THE SURGEON DECIDED TO CONVERT THE PROCEDURE TO OPEN SURGERY DUE TO FRUSTRATION WITH THE REPEATED ERROR. NO COMPLICATIONS TO THE PATIENT WERE REPORTED FOLLOWING THE CONVERSION TO OPEN SURGERY.
Patient problems: Unspecified Tissue Injury
MAUDE report 24932105 ↗
FDA clearance and classification
The FDA regulates product code NAY (system, surgical, computer controlled instrument) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the daVinci 5 380746
Has the daVinci 5 380746 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2310-2026 (Class II), began on Apr 24, 2026: Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the handheld camera to the robotic… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the daVinci 5 380746?
The FDA's MAUDE database lists 457 adverse event reports for this device, received between May 2, 2024 and Sep 28, 2026. In those reports, the most common specific device problems are “Visual Prompts will not Clear” (248), “Output Problem” (104) and “Failure to Power Up” (24); the most common patient problems are “Unspecified Tissue Injury” (25), “Hemorrhage/Blood Loss/Bleeding” (8) and “Burn(s)” (4). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the daVinci 5 380746?
2 reports are classified as deaths, 40 as injuries and 415 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the daVinci 5 380746, and is it still sold?
The FDA's device database lists Intuitive Surgical, Inc. as the company responsible for it (brand daVinci 5, model 380746). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other System, surgical, computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,308 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,204 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,630 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,325 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls233 deaths5,369 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,342 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,802 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,281 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page