daVinci 5 380747Recalls & reported problems

Intuitive Surgical, Inc. · Patient Side Cart · Class 2

All models of daVinci 5 (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This da Vinci 5 surgical robot is a computer-controlled system used in surgery. Many of the 580 reports describe on-screen error messages that will not clear, and some mention arms overheating or moving unexpectedly. The reports include 12 deaths and 70 injuries, and reports rose from 113 in the previous 12 months to 457 in the latest 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
572
Deaths reported
14
Injuries reported
70
Recalls naming it
0

Summary

The FDA's MAUDE database contains 572 adverse event reports that list this exact device (DI 00886874119716), received between Apr 30, 2024 and Sep 30, 2026 — classified as 14 deaths, 70 injuries, 488 malfunctions.

The most frequently reported specific device problem is “Visual Prompts will not Clear” (269 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 3 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 3 reportsNov 2024: 4 reportsDec 2024: 2 reportsJan 2025: 6 reportsFeb 2025: 5 reportsMar 2025: 7 reportsApr 2025: 15 reportsMay 2025: 8 reportsJun 2025: 6 reportsJul 2025: 13 reportsAug 2025: 21 reportsSep 2025: 11 reportsOct 2025: 15 reportsNov 2025: 21 reportsDec 2025: 29 reportsJan 2026: 26 reportsFeb 2026: 44 reportsMar 2026: 202 reportsApr 2026: 55 reportsMay 2026: 29 reportsJun 2026: 0 reportsJul 2026: 4 reportsAug 2026: 15 reportsSep 2026: 24 reportsOct 2026: 0 reports202420252026202
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Visual Prompts will not Clear269
  • Output Problem88
  • Adverse Event Without Identified Device or Use Problem55
  • No Device Output27
  • Positioning Problem25
  • Unintended Movement23
  • Unintended System Motion19
  • Connection Problem14

Patient problems

  • No Clinical Signs, Symptoms or Conditions452
  • Insufficient Information50
  • Hemorrhage/Blood Loss/Bleeding28
  • Unspecified Tissue Injury25
  • Cardiac Arrest6
  • Hematoma3
  • Bowel Perforation2
  • Fever2

Recent adverse event reports

The 25 most recent of 572 reports, in the reporter's own words as published by the FDA.

  • DeathSep 30, 2026Insufficient Device Problem Information

    It was reported that following a da Vinci-assisted hiatal hernia procedure, the patient developed an incarcerated trocar-site hernia and subsequently expired. On postoperative day five, the patient presented to the emergency room, where the hernia was identified and an unspecified subsequent procedure was performed. The cause and date of death, as well as additional details regarding the events leading to the death, were not provided. Attempts to obtain additional information from the surgeon and site were unsuccessful.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 50948206 ↗
  • InjurySep 29, 2026Image Orientation Incorrect; Unintended Movement

    Patient problems: Unspecified Tissue Injury

    MAUDE report 50940901 ↗
  • MalfunctionSep 29, 2026Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939695 ↗
  • MalfunctionSep 29, 2026Visual Prompts will not Clear

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939643 ↗
  • MalfunctionSep 29, 2026Visual Prompts will not Clear

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939641 ↗
  • MalfunctionSep 28, 2026Visual Prompts will not Clear

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50925842 ↗
  • MalfunctionSep 28, 2026Visual Prompts will not Clear

    Patient problems: Insufficient Information

    MAUDE report 50925830 ↗
  • MalfunctionSep 26, 2026Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50906720 ↗
  • MalfunctionSep 25, 2026Visual Prompts will not Clear

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905861 ↗
  • InjurySep 25, 2026Visual Prompts will not Clear

    Patient problems: Unspecified Tissue Injury

    MAUDE report 50902775 ↗
  • MalfunctionSep 23, 2026No Device Output

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50875854 ↗
  • MalfunctionSep 22, 2026No Device Output

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50861939 ↗
  • MalfunctionSep 22, 2026Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50860983 ↗
  • MalfunctionSep 22, 2026Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50860749 ↗
  • MalfunctionSep 22, 2026Unintended Movement

    It was reported that during a da Vinci-assisted cholecystectomy surgical procedure, Intuitive Surgical, Inc. (ISI) Clinical Sales Representative (CSR) informed Technical Support Engineer (TSE) that when the surgeon was using head in menu to change from 30 up to 30 down, the horizon was inverted. The surgeon had to manually move the orientation to resolve the issue. The endoscope was installed on Universal Surgical Manipulator (USM) 2. The customer continued to use USM 2 for the procedure. The procedure was completing as planned with no reported injury. On the following day, the CSR called in to provide more information about the issue. The TSE was informed that the issue had not recurred during the subsequent case. The TSE walked the CSR through checking the presence pins on the Universal Surgical Manipulator (USM) and the disc integrity, with the CSR verifying that all felt solid and that no oddities were found. The TSE then had the CSR compare the carriage integrity of the USM to another USM, and the CSR confirmed that no irregularities were found. The TSE advised the CSR that if the issue were to occur again on the USM, port hopping could be utilized as an option.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50852640 ↗
  • MalfunctionSep 11, 2026Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50744281 ↗
  • MalfunctionSep 10, 2026Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50724245 ↗
  • MalfunctionSep 10, 2026No Device Output

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50723868 ↗
  • DeathSep 8, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Cardiac Arrest; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 50673579 ↗
  • MalfunctionSep 5, 2026Output Problem

    Patient problems: Insufficient Information

    MAUDE report 50626903 ↗
  • MalfunctionSep 4, 2026Visual Prompts will not Clear

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50618939 ↗
  • MalfunctionSep 4, 2026Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50618938 ↗
  • MalfunctionSep 4, 2026Visual Prompts will not Clear

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50618936 ↗
  • MalfunctionSep 1, 2026Connection Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50553081 ↗
  • MalfunctionAug 27, 2026Visual Prompts will not Clear

    It was reported that during a da Vinci-assisted surgical procedure, the customer reported that the arm temperature was higher than normal. The caller reported the system had been on since about 6 am that morning and draped for the last 30 to 45 minutes. The Technical Support Engineer (TSE) advised the caller to reposition the system arms to allow for adequate air flow. The TSE viewed the logs and noted an error 1119 in the logs pointing to an AC fan error. The customer continued with the case. The customer called back about the Universal Surgical Manipulator (USM) overheating. The caller said the site put a gown over the system arms after draping and waiting for the case to start. The TSE informed that third party covers like a gown can act like a blanket not allowing the USM to properly cool. They were going to disable USM 1 and do the case as a 3 arm system and power cycle between cases. The procedure was continued as planned with no reported injury.Intuitive Surgical, Inc. (ISI) followed up with the initial reporter and obtained the following additional information: The procedure was continued with three system arms instead of four. The customer was unable to resolve the error m

    Patient problems: Insufficient Information

    MAUDE report 50468051 ↗

Other System, surgical, computer controlled instrument devices