EndoWrist 471171Recalls & reported problems
Intuitive Surgical, Inc. · MICRO BIPOLAR FORCEPS · Class 2
All models of EndoWrist (62)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00886874121474), received between Sep 5, 2023 and Feb 25, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (6 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Dec 19, 2024Open, Classified
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.
Z-0875-2025 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Split, Cut or Torn6
- Thermal Decomposition of Device3
- Break1
- Energy Output Problem1
- Material Fragmentation1
- Material Frayed1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information5
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 25, 2026Material Frayed
IT WAS REPORTED THAT THERE WAS A FRAYED CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: Insufficient Information
MAUDE report 24449752 ↗MalfunctionDec 12, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23796288 ↗MalfunctionDec 3, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: Insufficient Information
MAUDE report 23709649 ↗MalfunctionOct 30, 2025Break; Break
IT WAS REPORTED THAT THE MARYLAND BIPOLAR FORCEPS INSTRUMENT OPEN/CLOSE PART OF THE TIP IS MISSING. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR REPORTED INJURY. ISI FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE INSTRUMENT DAMAGE WAS FIRST IDENTIFIED DURING POST-CLEANING INSPECTION AFTER STERILIZATION. THE SURGEON DID NOT COMMENT ON THE CAUSE SINCE NO FRAGMENT FELL AND NOTED NO FUNCTIONAL ISSUES DURING SURGERY. THERE WAS NO PATIENT INJURY. THE BREAKAGE LIKELY OCCURRED WHILE THE INSTRUMENT WAS INSIDE THE PATIENT'S ANATOMY, HOWEVER NO FRAGMENT WAS CONFIRMED INSIDE PATIENT'S ANATOMY AND NO INJURY WAS REPORTED.
Patient problems: Insufficient Information
MAUDE report 23421483 ↗MalfunctionMay 7, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE 8MM MICRO BIPOLAR FORCEPS INSTRUMENT HAD A BROKEN WIRE. NO FRAGMENTS FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21980902 ↗MalfunctionApr 24, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED CARDIAC SURGICAL PROCEDURE, AT THE END OF THE SURGICAL PROCEDURE, THE STEEL CABLE AT THE TIP OF THE MICRO BIPOLAR FORCEPS JOINT CAME LOOSE COMPLETELY. AS IT WAS AT THE END OF THE SURGERY, IT WAS NOT NECESSARY TO USE ANOTHER CLAMP. INSTRUMENT HAVE 10 LIVES REMAINING. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE SITE AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE CUSTOMER CLARIFIED THAT THE DAMAGE TO THE MATERIAL WAS IDENTIFIED OUTSIDE THE PATIENT, DURING THE CHANGE OF CLAMPS, I.E. OUTSIDE THE SURGICAL FIELD AND NOT DURING DIRECT USE ON THE PATIENT. THERE WERE NO REPORTS OF FRAGMENTS FALLING FROM THE DEVICE DURING USE ON THE PATIENT, NOR ANY INJURY OR COMPLICATION RELATED TO THE RETENTION OF MATERIAL. THERE WAS ALSO NO NEED FOR THE PATIENT TO RETURN TO HOSPITAL FOR THIS REASON.
Patient problems: Insufficient Information
MAUDE report 21904624 ↗MalfunctionApr 2, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21757141 ↗MalfunctionJun 27, 2024Thermal Decomposition of Device; Material Twisted/Bent; Material Split, Cut or Torn; Thermal Decomposition of Device; Material Twisted/Bent; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE TEETH OF THE MICRO BIPOLAR FORCEPS INSTRUMENT DID NOT MATCH. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19630458 ↗MalfunctionDec 22, 2023Thermal Decomposition of Device; Energy Output Problem; Material Split, Cut or Torn; Thermal Decomposition of Device; Energy Output Problem; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE MICRO BIPOLAR FORCEPS INSTRUMENT WAS NOT DELIVERING ENERGY CONSISTENTLY AND NOT CAUTERIZING. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE CUSTOMER NOTED NO THERMAL DAMAGE WAS OBSERVED. NO ARCING WAS OBSERVED DURING THE PROCEDURE. NO SPARKING, SMOKE OCCURRED DURING PROCEDURE. NO PATIENT INJURY OR ADVERSE EVENT DURING OR AFTER SURGICAL PROCEDURE. IT IS UNSURE IF THERE ARE ANY PHOTOGRAPHIC IMAGES OF THE DEVICE(S) OR A VIDEO RECORDING OF THE PROCEDURE AVAILABLE FOR ISI REVIEW. THE INSTRUMENT WAS INSPECTED PRIOR TO USE AND NO DAMAGE OR ANYTHING OUT OF THE ORDINARY. THERE WAS NO THERMAL DAMAGE OBSERVED. TO THEIR KNOWLEDGE, THERE WAS NO INSTRUMENT COLLISION WITH AN ENERGY INSTRUMENT DURING THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18381280 ↗MalfunctionSep 5, 2023Thermal Decomposition of Device; Thermal Decomposition of Device
IT WAS REPORTED THAT DURING CENTRAL PROCESSING, THE MICRO BIPOLAR FORCEPS INSTRUMENT TIP (YELLOWISH PART) LOOKED MELTED OFF. THE INSTRUMENT WAS NEVER USED FOR A CASE JUST IN CASE THE MELTED PART HAS ROUGH EDGES THAT MAY HURT THE PATIENT. THERE WAS NO REPORT OF PATIENT INVOLVEMENT. INTUITIVE SURGICAL, INC. (ISI) MADE MULTIPLE FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17686521 ↗
Other System, surgical, computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,225 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,156 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,581 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,302 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls235 deaths5,316 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,310 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,770 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,273 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page