EndoWrist 471400Recalls & reported problems

Intuitive Surgical, Inc. · BIPOLAR GRASPER, LONG · Class 2

All models of EndoWrist (62)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Intuitive EndoWrist bipolar grasper is the subject of 1,027 adverse event reports, which mention no deaths and 34 injuries. Most reports describe cables or materials that split, tore, fragmented or frayed. 1 open recall, started in December 2024, concerns frayed or broken grip cables on reusable instruments with jaws.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,023
Deaths reported
0
Injuries reported
34
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1,023 adverse event reports that list this exact device (DI 00886874121528), received between May 16, 2021 and Sep 30, 2026 — classified as 34 injuries, 989 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (632 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 3 reportsJan 2024: 4 reportsFeb 2024: 1 reportsMar 2024: 6 reportsApr 2024: 10 reportsMay 2024: 4 reportsJun 2024: 5 reportsJul 2024: 9 reportsAug 2024: 5 reportsSep 2024: 8 reportsOct 2024: 10 reportsNov 2024: 5 reportsDec 2024: 7 reportsJan 2025: 6 reportsFeb 2025: 4 reportsMar 2025: 120 reportsApr 2025: 47 reportsMay 2025: 87 reportsJun 2025: 66 reportsJul 2025: 52 reportsAug 2025: 41 reportsSep 2025: 26 reportsOct 2025: 32 reportsNov 2025: 46 reportsDec 2025: 35 reportsJan 2026: 23 reportsFeb 2026: 203 reportsMar 2026: 32 reportsApr 2026: 18 reportsMay 2026: 13 reportsJun 2026: 2 reportsJul 2026: 1 reportsAug 2026: 3 reportsSep 2026: 11 reportsOct 2026: 0 reports202420252026203
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 19, 2024Open, Classified

    Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

    Z-0882-2025 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Split, Cut or Torn632
  • Material Frayed119
  • Material Fragmentation68
  • Break44
  • Thermal Decomposition of Device39
  • Detachment of Device or Device Component31
  • No Apparent Adverse Event26
  • Unintended Movement13

Patient problems

  • No Clinical Signs, Symptoms or Conditions761
  • Insufficient Information246
  • Foreign Body In Patient12
  • Hemorrhage/Blood Loss/Bleeding5
  • Unspecified Tissue Injury4

Recent adverse event reports

The 25 most recent of 1,023 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Material Split, Cut or Torn

    Patient problems: Insufficient Information

    MAUDE report 50954081 ↗
  • MalfunctionSep 30, 2026Entrapment of Device; Difficult to Open or Close

    It was reported that during a da Vinci-assisted Radical Prostatectomy without Lymphadenectomy surgical procedure, the Surgeon lost complete vision on the console while the ProGrasp Forceps was grasping tissue. A system restart was done, but the system could not complete its self-checks since two instruments, a ProGrasp Forceps instrument and a Long Bipolar Grasper instrument, were grasping tissue. The system was put into an emergency state, and the instruments had to manually be opened with the emergency tool kit. The tissue was removed and the instruments were removed. The system was resumed for use, and the self-check was completed. The procedure was completed with no reported injury.Intuitive Surgical, Inc. (ISI) followed up with the Distributor Clinical Sales Representative (CSR) and obtained the following additional information: The customer confirmed the Long Bipolar Grasper was the other instrument involved and reported that both instruments were on tissue and had to be manually opened and removed using the Instrument Release Kit (IRK). Once the jaws were successfully opened, an assistant used a laparoscopic grasper to detach and remove the tissue caught between the jaws of

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948888 ↗
  • MalfunctionSep 28, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50925878 ↗
  • MalfunctionSep 28, 2026Unintended Movement

    It was reported that the Long Bipolar Grasper instrument was not articulating properly. The customer reported event has no report of patient injury.

    Patient problems: Insufficient Information

    MAUDE report 50920112 ↗
  • MalfunctionSep 25, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905859 ↗
  • MalfunctionSep 25, 2026Material Frayed

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905835 ↗
  • MalfunctionSep 24, 2026Material Split, Cut or Torn

    Patient problems: Insufficient Information

    MAUDE report 50891437 ↗
  • MalfunctionSep 24, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50891414 ↗
  • MalfunctionSep 22, 2026No Device Output

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50861262 ↗
  • MalfunctionSep 9, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50701082 ↗
  • MalfunctionSep 4, 2026Thermal Decomposition of Device

    It was reported that during a da Vinci-assisted surgical procedure, the Long Bipolar Grasper instrument stopped coagulating and started to smell burnt. The ceramic component at the tip of the instrument appears to have overheated. The procedure was completed with no reported injury.ISI followed up with the initial reporter and obtained the following additional information: No arcing was observed. The non-energy instrument did not come into contact with or collide with another activated energy instrument during the procedure.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50618888 ↗
  • MalfunctionAug 13, 2026Break; Thermal Decomposition of Device

    It was reported that during central processing, the Long Bipolar Grasper instrument white plastic at tip has a chunk missing on one side, looks like its dirty but a chunk was gone. The procedure was completed with no reported injury.Intuitive Surgical, Inc. (ISI) followed up with the initial reporter to obtain additional information; however, no further details are available regarding the reported event

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50231314 ↗
  • MalfunctionAug 12, 2026Material Fragmentation; Material Split, Cut or Torn

    Patient problems: Insufficient Information

    MAUDE report 50218064 ↗
  • InjuryAug 10, 2026Material Fragmentation

    It was reported that during a da Vinci-assisted pulmonary lobectomy procedure, the Long Bipolar Grasper instrument tip broke. A fragment fell into the patient and was retrieved during the same procedure. The customer confirmed the fragment was located in a suction canister. It was removed and matched with the Long Bipolar Grasper instrument along the break line. The surgeon confirmed full fragment retrieval. No post-operative tests were performed. No additional procedure was required. The procedure was completed. There were no complications and the patient has not returned to the hospital due to experiencing any post-surgical complications related to the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50193631 ↗
  • MalfunctionJul 7, 2026Thermal Decomposition of Device

    IT WAS REPORTED THAT DURING CENTRAL PROCESSING, THE LONG BIPOLAR GRASPER TIPS ARE BLACKENED. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25741691 ↗
  • MalfunctionJun 30, 2026Thermal Decomposition of Device

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE 8MM LONG BIPOLAR GRASPER INSTRUMENT HAD INTRA-ABDOMINAL SPARKS HAD OCCURRED. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25672865 ↗
  • MalfunctionJun 30, 2026Thermal Decomposition of Device

    IT WAS REPORTED THAT DURING CENTRAL PROCESSING, THE DISTAL TIP OF A LONG BIPOLAR GRASPER INSTRUMENT APPEARED MELTED/BURNT IN AREA OF MANIPULATOR WIRES. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR PATIENT INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: IT WAS UNKNOWN IF ARCING WAS OBSERVED. THERE WAS NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25662478 ↗
  • MalfunctionMay 27, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 25309023 ↗
  • MalfunctionMay 26, 2026Energy Output Problem

    IT WAS REPORTED THAT DURING CENTRAL PROCESSING, THE LONG BIPOLAR GRASPER CAUTERY WAS NOT WORKING ON THE INSTRUMENT. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25290162 ↗
  • MalfunctionMay 22, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 25262526 ↗
  • MalfunctionMay 21, 2026Material Frayed

    IT WAS REPORTED THAT THERE WAS A FRAYED CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25249124 ↗
  • MalfunctionMay 19, 2026

    WOULDN'T RELEASE FROM ARM/STERILE ADAPTOR. PROCEDURE COMPLETED AND DEVICE PRIED OFF OF STERILE ADAPTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25215623 ↗
  • MalfunctionMay 18, 2026Unintended Movement

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH AN INTUITIVE PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25210449 ↗
  • MalfunctionMay 18, 2026Material Frayed

    IT WAS REPORTED THAT THERE WAS A FRAYED CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 25196986 ↗
  • MalfunctionMay 14, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25178321 ↗

Other System, surgical, computer controlled instrument devices