EndoWrist SP 430152Recalls & reported problems
Intuitive Surgical, Inc. · MARYLAND BIPOLAR FORCEPS · Class 2
All models of EndoWrist SP (20)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00886874123461), received between Sep 9, 2025 and Jun 4, 2026 — classified as 2 injuries, 9 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Unintended System Motion2
- Arcing1
- Detachment of Device or Device Component1
- Material Fragmentation1
- Material Twisted/Bent1
- Thermal Decomposition of Device1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Insufficient Information3
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 4, 2026Thermal Decomposition of Device
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED PROSTATECTOMY - RADICAL EXTRAPERITONEAL W/ LYMPHADENECTOMY SURGICAL PROCEDURE, THE MARYLAND BIPOLAR FORCEPS INSTRUMENT WAS BROKEN. NO KNOWN IMPACT OR PATIENT CONSEQUENCE WAS REPORTED.
Patient problems: Insufficient Information
MAUDE report 25401287 ↗MalfunctionMay 19, 2026Break
IT WAS REPORTED THAT DURING CENTRAL PROCESSING, THE MARYLAND BIPOLAR FORCEPS INSTRUMENT HAD A CHIP AT THE TIP. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR PATIENT INJURY. INTUITIVE FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE EXACT TIME THE DAMAGE OCCURRED IS UNKNOWN, AS IT WAS ONLY NOTICED AFTER THE ITEM WAS SENT TO THE CENTRAL STERILE DEPARTMENT. THERE HAVE BEEN NO REPORTS OF FRAGMENTS FALLING, AND NO FRAGMENTS HAVE BEEN FOUND OR REPORTED. ADDITIONALLY, THERE HAVE BEEN NO COMPLICATIONS REPORTED, AND THE PATIENT HAS NOT RETURNED TO THE HOSPITAL. FROM MY RECOLLECTION, THE DAMAGE WAS A VERY SMALL CRACK, BUT I DO NOT REMEMBER IF ANYTHING WAS MISSING. THERE ARE NO PHOTOGRAPHS OR IMAGES AVAILABLE TO PROVIDE FURTHER DETAIL.
Patient problems: Insufficient Information
MAUDE report 25224336 ↗MalfunctionMay 13, 2026Break
IT WAS REPORTED THAT DURING CENTRAL PROCESSING, THE SP MARYLAND BIPOLAR FORCEPS INSTRUMENT WAS BROKEN IN THE JOINT, AND THE PLASTIC PART OF THE JOINT FELL OFF DURING THE RETURN PROCESS. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25157267 ↗MalfunctionApr 7, 2026Unintended System Motion
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24811353 ↗MalfunctionApr 3, 2026Material Twisted/Bent
IT WAS REPORTED THAT DURING DA VINCI-ASSISTED PEDIATRIC PYELOPLASTY SURGICAL PROCEDURE, SP MARYLAND BIPOLAR FORCEPS INSTRUMENT HAD MISALIGNED JAWS. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24773091 ↗MalfunctionMar 26, 2026Break
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED RADICAL PROSTATECTOMY WITH LYMPHADENECTOMY SURGICAL PROCEDURE, THE MARYLAND BIPOLAR FORCEPS INSTRUMENT'S TIP BROKE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24707016 ↗InjuryMar 9, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE TIP FELL OFF THE MARYLAND BIPOLAR FORCEPS INSTRUMENT DURING DISSECTION. THE PROCEDURE WAS COMPLETED. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE INSTRUMENT WAS INSPECTED PRIOR TO USE AND SHOWED NO INITIAL DAMAGE. DURING A DISSECTING TASK, A FRAGMENT FELL INSIDE THE PATIENT DUE TO A COLLISION WITH ANOTHER INSTRUMENT. THE CAUSE OF THE BREAKAGE WAS UNKNOWN, BUT THE FRAGMENT WAS RETRIEVED DURING THE SAME PROCEDURE, AND ALL FRAGMENTS WERE VISUALLY CONFIRMED AS RETRIEVED. DURING FINAL REMOVAL, THE WRIST WAS STRAIGHTENED AND THERE WAS NO RESISTANCE FELT; THERE WAS DAMAGE TO THE INSTRUMENT AND CANNULA. THERE WAS NO PATIENT INJURY, AND NO ADDITIONAL SURGICAL PROCEDURES OR POST-OPERATIVE TESTS WERE REQUIRED. THE PROCEDURE WAS COMPLETED ROBOTICALLY, AND THE INSTRUMENT IS AVAILABLE FOR RETURN TO INTUITIVE SURGICAL, INC. (ISI) FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24544217 ↗MalfunctionMar 6, 2026Arcing
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED RADICAL TRANSVESICAL PROSTATECTOMY SURGICAL PROCEDURE, THE SURGEON OBSERVED ARCING WHEN CAUTERY WAS APPLIED CLOSE TO THE CAMERA, CAUSING THE CAMERA IMAGE TO INTERMITTENTLY CUT IN AND OUT. THE CUSTOMER RECEIVED PHONE ASSISTANCE FROM INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE). THE TSE COULD NOT FIND ANY RELATED ERRORS IN THE LOGS. THE TSE RECOMMENDED MARKING THE CAMERA AND ENSURING THE SHEATH WAS NOT DAMAGED. IF THE ISSUE PERSISTED, THE TSE SUGGESTED RETURNING THE CAMERA. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24538914 ↗InjuryFeb 7, 2026Material Fragmentation
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24286115 ↗MalfunctionOct 23, 2025Break; Unintended Movement; Break; Unintended Movement
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE 6MM MARYLAND BIPOLAR FORCEPS JAWS WOULD NOT CLOSE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23364714 ↗MalfunctionSep 9, 2025Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED OVARIAN CYSTECTOMY SURGICAL PROCEDURE, THE MARYLAND BIPOLAR FORCEPS INSTRUMENT MOVED IN THE OPPOSITE DIRECTION OF THE SURGEON'S COMMAND. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THERE WERE NO REPORTS OF PATIENT INJURY. THE PROCEDURE WAS COMPLETED WITH A BACKUP SIMILAR INSTRUMENT.
Patient problems: Insufficient Information
MAUDE report 23008169 ↗
Other System, surgical, computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,225 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,156 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,581 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,302 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls235 deaths5,316 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,310 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,770 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,273 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page