EndoWrist SP 430300Recalls & reported problems

Intuitive Surgical, Inc. · CADIERE FORCEPS · Class 2

All models of EndoWrist SP (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00886874123843), received between Nov 5, 2025 and Feb 12, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Unintended Movement” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended Movement1
  • Unintended System Motion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 12, 2026Unintended System Motion

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED RADICAL EXTRAPERITONEAL PROSTATECTOMY WITHOUT LYMPHADENECTOMY SURGICAL PROCEDURE, NON-INTUITIVE MOTIONS OCCURRED ON AN INSTRUMENT (K10250918-0025) THAT WAS INSTALLED ON UNIVERSAL SURGICAL MANIPULATOR (USM) 2. A TECHNICAL SUPPORT ENGINEER (TSE) WAS CONTACTED BY THE CUSTOMER FOR TROUBLESHOOTING. THE TSE CHECKED THE SYSTEM LOGS AND FOUND SEVERAL ENGAGEMENT ERRORS RECORDED ON USM 2. THE TSE ASKED THE CUSTOMER TO RESEAT THE STERILE ADAPTER AND INSTRUMENT ON USM 2. THE CALLER STATED THAT THE ISSUE PERSISTED. THE TSE THEN ASKED THE CUSTOMER TO INSTALL A DIFFERENT INSTRUMENT ON USM 2. THE CALLER REPORTED THAT WITH A DIFFERENT INSTRUMENT THE ISSUE WAS RESOLVED. THE TSE ASKED THE CALLER TO RE-INSTALL THE AFFECTED INSTRUMENT ON USM 2. THE CALLER STATED THAT THE ISSUE REAPPEARED, CONFIRMING THAT THE PROBLEM WAS RELATED TO THE CADIERE FORCEPS INSTRUMENT. THE PROCEDURE WAS COMPLETING AS PLANNED WITH NO REPORTED INJURY.INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE SURGEON'S HEAD WAS INSIDE THE VIEWER AND THE ISSUE OCCURRED WHILE THE SURGEON WAS TRYING TO MANIPULATE INSTRUM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24328478 ↗
  • MalfunctionNov 5, 2025Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CADIERE FORCEPS INSTRUMENT WOULD NOT FOLLOW COMMANDS PROPERLY WITH NO OTHER INFORMATION GIVEN, AND THE CUSTOMER WAS UNAWARE IF ANY TROUBLESHOOTING WAS DONE LIKE RE-ATTACHING THE DRAPE STERILE ADAPTOR. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. FOLLOW-UP WAS ATTEMPTED TO GATHER MORE INFORMATION, BUT THE CUSTOMER WAS NOT ABLE TO PROVIDE ANY ADDITIONAL DETAILS RELATED TO THE EVENT/INSTRUMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23482288 ↗

Other System, surgical, computer controlled instrument devices