EndoWrist SP 431100Recalls & reported problems

Intuitive Surgical, Inc. · MONOPOLAR CURVED SCISSORS ER · Class 2

All models of EndoWrist SP (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
17
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 00886874124048), received between Jul 9, 2025 and Sep 29, 2026 — classified as 4 injuries, 13 malfunctions.

The most frequently reported specific device problem is “Break” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 4 reportsFeb 2026: 3 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break8
  • Detachment of Device or Device Component3
  • Material Split, Cut or Torn2
  • Unintended Movement2
  • Difficult to Open or Close1
  • Material Fragmentation1
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Insufficient Information5
  • Foreign Body In Patient1

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Unintended Movement

    It was reported that prior to starting a da Vinci-assisted Urology surgical procedure, a Monopolar Curved Scissors (MCS) instrument exhibited non intuitive motion. The customer continued with the procedure as planned, and no patient impact or consequence was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927526 ↗
  • MalfunctionMay 19, 2026Break

    IT WAS REPORTED THAT PRIOR TO A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE TIP OF THE MONOPOLAR CURVED SCISSORS INSTRUMENT COULD NOT BE ATTACHED. THERE WAS NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25224267 ↗
  • MalfunctionApr 29, 2026Break

    IT WAS REPORTED THAT THE SINGLE PORT (SP) EXTENDED RANGE MONOPOLAR CURVED SCISSORS HAD A BROKEN TIP. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) MADE A FOLLOW-UP ATTEMPT TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS COULD BE PROVIDED AND OBTAINED THE FOLLOWING: THE CUSTOMER REPORTED THAT THEY DID NOT HAVE ANY MORE INFORMATION, OTHER THAN WHAT WAS ORIGINALLY OFFERED.

    Patient problems: Insufficient Information

    MAUDE report 25037250 ↗
  • MalfunctionApr 22, 2026Break

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED MYOMECTOMY SURGICAL PROCEDURE, THE MONOPOLAR CURVED SCISSORS INSTRUMENT'S EXTENDED R HAD A BREAKAGE RESULTING IN THE SCISSORS TAB DETACHING FROM THE INSTRUMENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE PART OF THE MONOPOLAR CURVED SCISSORS (MCS) EXTENDED R BROKE WAS THE DISTAL END OF SP MONOPOLAR INSTRUMENT. THE INCIDENT DID NOT HAVE A FRAGMENT FALL INSIDE THE PATIENT'S ANATOMY. THE PROCEDURE THE INSTRUMENT WAS USED IN WAS OVARIAN CYSTECTOMY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24959494 ↗
  • InjuryMar 30, 2026Material Fragmentation

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24723125 ↗
  • InjuryFeb 25, 2026Detachment of Device or Device Component

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: Foreign Body In Patient

    MAUDE report 24442310 ↗
  • MalfunctionFeb 24, 2026Difficult to Open or Close; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED MYOMECTOMY SURGICAL PROCEDURE, WHILE EXCISING THE MYOMA PROTRUDING FROM THE FRONT OF THE UTERUS AND DISSECTING THE TISSUE, IT APPEARED ON THE SURGICAL SCREEN THAT THE JAW OF THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT WAS NOT CLOSING COMPLETELY. THE SURGEON ATTEMPTED TO OPEN AND CLOSE THE JAW SEVERAL TIMES AND, SUSPECTING IT WAS NOT FUNCTIONING PROPERLY, REQUESTED AN INSTRUMENT CHECK. UPON REMOVING THE SHEATH AND INSPECTING THE WIRE, IT WAS DISCOVERED THAT THE WIRE NEAR THE TIP WAS BROKEN. WHILE CHECKING THE INSTRUMENT'S OPERATION BY OPENING AND CLOSING THE RELEASE DIAL, IT WAS CONFIRMED THAT THE WIRE WAS BREAKING LITTLE BY LITTLE WHEN THE JAW WAS OPENED AND CLOSED. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24434734 ↗
  • MalfunctionFeb 9, 2026Break

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE TIP OF A SINGLE PORT MONOPOLAR CURVED SCISSORS INSTRUMENT HAD A BROKEN TIP. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) CONTACTED THE CUSTOMER BUT WAS NOT ABLE TO OBTAIN ANY ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24288877 ↗
  • MalfunctionJan 29, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ISSUE WITH THE SINGLE PORT (SP) MONOPOLAR CURVED SCISSORS (MCS). THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY. INTUITIVE SURGICAL, INC. (ISI) MADE MULTIPLE FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24217017 ↗
  • MalfunctionJan 13, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THERE WAS A BROKEN CABLE WITH A DA VINCI PRODUCT. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 24050514 ↗
  • InjuryJan 12, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED HYSTERECTOMY PROCEDURE, THE TIP OF THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT FELL OFF INSIDE THE PATIENT. THE TIP WAS RETRIEVED DURING THE SAME SURGICAL PROCEDURE WHICH WAS COMPLETED ROBOTICALLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24038076 ↗
  • InjuryJan 7, 2026Detachment of Device or Device Component

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23989800 ↗
  • MalfunctionDec 16, 2025Break; Break

    IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT'S TIP WAS CHIPPED. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23823893 ↗
  • MalfunctionDec 16, 2025Unintended Movement; Unintended Movement

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 23813396 ↗
  • MalfunctionNov 26, 2025Break; Break

    REFER TO H11 FOR FOLLOW-UP INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 23653852 ↗
  • MalfunctionNov 24, 2025Break; Break

    IT WAS REPORTED THAT PRIOR TO THE START OF A DA VINCI-ASSISTED NEPHRECTOMY SURGICAL PROCEDURE, THE MONOPOLAR CURVED SCISSORS INSTRUMENT'S TIP WAS BROKEN, AND THE SCISSOR COULD NOT BE PLACED. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE TIP WAS NOTICED BEFORE THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23639269 ↗
  • MalfunctionJul 9, 2025Break; Break

    IT WAS REPORTED THAT DURING CENTRAL PROCESSING, SINGLE-PORT (SP) MONOPOLAR CURVED SCISSORS (MCS) DISTAL TIP WAS FOUND BROKEN. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.

    Patient problems: Insufficient Information

    MAUDE report 22463706 ↗

Other System, surgical, computer controlled instrument devices

All System, surgical, computer controlled instrument devices →