Force Feedback 476049Recalls & reported problems
Intuitive Surgical, Inc. · Force Feedback Cadiere Forceps · Class 2
All models of Force Feedback (11)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00886874125564), received between Oct 15, 2025 and May 4, 2026 — classified as 3 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Material Fragmentation3
- Crack1
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Bacterial Infection1
- Insufficient Information1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 4, 2026Break
IT WAS REPORTED THAT DURING CENTRAL PROCESSING, THE CUSTOMER NOTED THE FORCE FEEDBACK CADIERE FORCEPS JAW WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25074609 ↗InjuryApr 17, 2026Material Fragmentation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED GASTRIC BYPASS PROCEDURE, A SMALL CURVED METALLIC FRAGMENT FELL FROM THE FORCE FEEDBACK CADIERE FORCEPS INSTRUMENT AND WAS RETRIEVED FROM THE PATIENT'S ABDOMEN DURING THE SAME PROCEDURE. THE CUSTOMER CONFIRMED THAT ALL VISIBLE FRAGMENTS WERE RETRIEVED; ONE PIECE WAS IDENTIFIED AND REMOVED USING A LAPAROSCOPIC GRASPER. THE PIECE WAS SMALL, AND NO POST-OPERATIVE IMAGING SUCH AS X-RAY OR ULTRASOUND WAS PERFORMED. IT WAS NOT KNOWN WHETHER THE PATIENT HAD RETURNED TO THE HOSPITAL, BUT NO COMPLICATIONS OR ADDITIONAL SURGICAL PROCEDURES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24917957 ↗MalfunctionFeb 6, 2026No Apparent Adverse Event
IT WAS REPORTED THAT DURING CENTRAL PROCESSING, THE FORCE FEEDBACK CADIERE FORCEPS INSTRUMENT WAS DAMAGED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24284452 ↗MalfunctionJan 30, 2026Break; Crack
IT WAS REPORTED THAT THE FORCE FEEDBACK CADIERE FORCEPS INSTRUMENT DROPPED ON THE FLOOR AND CRACKED AT THE DISTAL END. NO KNOWN IMPACT OR PATIENT CONSEQUENCE WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24226257 ↗MalfunctionJan 30, 2026Break
IT WAS REPORTED THAT DURING CENTRAL PROCESSING, THE 8MM FORCE FEEDBACK CADIERE FORCEPS INSTRUMENT TIP WAS DAMAGED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24225321 ↗MalfunctionDec 10, 2025Break; Break
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: Insufficient Information
MAUDE report 23765214 ↗InjuryOct 15, 2025Material Fragmentation; Material Fragmentation
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: Bacterial Infection; No Clinical Signs, Symptoms or Conditions
MAUDE report 23300950 ↗InjuryOct 15, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED RIGHT HEMICOLECTOMY PROCEDURE, THE SHAFT OF THE FORCE FEEDBACK CADIERE FORCEPS BROKE INSIDE THE PATIENT. THE CUSTOMER STATED THE PLASTIC ON THE SHAFT OF THE INSTRUMENT BROKE OFF INSIDE THE PATIENT. THE CUSTOMER RETRIEVED THE PIECES DURING THE SAME PROCEDURE WHICH WAS COMPLETED ROBOTICALLY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED THE FOLLOWING ADDITIONAL INFORMATION: THE INSTRUMENT WAS INSPECTED PRIOR TO USE, AND NO DAMAGE WAS NOTED. THE SURGICAL TASK BEING PERFORMED AT THE TIME OF THE FRAGMENT FELL INSIDE THE PATIENT WAS RETRACTION OF TISSUE. IT IS UNKNOWN HOW LONG THE INSTRUMENT WAS IN USE WHEN THE ISSUE WAS IDENTIFIED. THE INSTRUMENT DID NOT COLLIDE WITH ANY OTHER INSTRUMENTS OR OTHER HARD MATERIAL DURING THE SURGICAL PROCEDURE. THE INSTRUMENT WAS NEVER REMOVED FROM THE PATIENT UNTIL IT WAS NOTICED TO BE BROKEN. THE SURGICAL STAFF DID NOT FEEL ANY RESISTANCE UPON REMOVAL OF THE INSTRUMENT THROUGH THE CANNULA. AFTER REMOVING THE INSTRUMENT, THE STAFF NOTICED ANOTHER PIECE MISSING. THE FIRST PIECE WAS REMOVED ROBOTICALLY; TWO ADDITIONAL FRAGMENTS WERE ATTACHED TO THE SPECIMEN. THE FRAGMENTS WE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23297541 ↗
Other System, surgical, computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,225 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,156 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,581 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,302 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls235 deaths5,316 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,310 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,770 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,273 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page