Force Feedback 476093Recalls & reported problems
Intuitive Surgical, Inc. · Force Feedback ProGrasp Forceps · Class 2
All models of Force Feedback (11)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00886874125595), received between Jul 2, 2025 and May 18, 2026 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Detachment of Device or Device Component1
- Material Separation1
- Tactile Prompts/Feedback1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Insufficient Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 18, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CUSTOMER EXPERIENCED DIFFICULTY REMOVING THE FORCE FEEDBACK PROGRASP FORCEPS (FFPF) INSTRUMENT FROM ONE OF THE ARMS BECAUSE A PIN WAS DISLODGED AND PREVENTED THE INSTRUMENT FROM RETRACTING THROUGH THE CANNULA. THE CUSTOMER MANUALLY PUSHED THE PIN BACK INTO PLACE AND WAS THEN ABLE TO REMOVE THE INSTRUMENT FROM THE CANNULA. THE PROCEDURE WAS COMPLETING AS PLANNED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25210413 ↗MalfunctionMay 6, 2026Tactile Prompts/Feedback
IT WAS REPORTED THAT DURING A DA VINCI ASSISTED SURGICAL PROCEDURE, POTENTIAL ERRORS OCCURRED REPEATEDLY. AFTER 3 RETRIES, A DIFFERENT FORCE FEEDBACK ARM WAS RECOGNIZED AND NORMAL OPERATION WAS CONFIRMED. THE CORRESPONDING ARM IS NOT BEING RECOGNIZED; THIS IS PRESUMED TO BE AN ERROR ISSUE. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25105105 ↗InjuryFeb 11, 2026Material Separation
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE PIN FELL OFF THE TIP OF THE FORCE FEEDBACK INSTRUMENT AND INTO THE PATIENT. THE FRAGMENT WAS RETRIEVED DURING THE SAME PROCEDURE. THE PIN WAS PLACED IN A JAR ACCOMPANYING THE INSTRUMENT. THE PROCEDURE WAS COMPLETED. INTUITIVE SURGICAL, INC. (ISI) MADE MULTIPLE FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24313574 ↗MalfunctionFeb 3, 2026Break
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE THE FORCE FEEDBACK PROGRASP FORCEPS INSTRUMENT SHEATH WAS BROKEN. THE CUSTOMER REPORTED EVENT HAS NO REPORT OF PATIENT INJURY. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER TO OBTAIN ADDITIONAL INFORMATION, HOWEVER, NO FURTHER DETAILS ARE AVAILABLE REGARDING THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24253901 ↗MalfunctionDec 11, 2025Break; Break
IT WAS REPORTED THAT DURING CENTRAL PROCESSING, THE FORCE FEEDBACK INSTRUMENT SHAFT WAS BROKEN. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR REPORTED INJURY.
Patient problems: Insufficient Information
MAUDE report 23784092 ↗MalfunctionJul 2, 2025Break; Break
REFER TO H11 FOR FOLLOW-UP INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22379912 ↗
Other System, surgical, computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,225 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,156 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,581 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,302 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls235 deaths5,316 reports
- daVinci Xi 381121Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall2 deaths4,310 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,770 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,273 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page