Hill-Rom Services Pte. Ltd. · VOLARA, HC NA A · Class 2

All models of Volara System (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00887761981690), received between Mar 20, 2026 and Apr 8, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Crack” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 3, 2026Open, Classified

    Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.

    Z-3257-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Crack1
  • Insufficient Device Problem Information1
  • Insufficient Flow or Under Infusion1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Pneumonia1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 8, 2026Use of Device Problem; Insufficient Flow or Under Infusion; Insufficient Device Problem Information

    IT WAS REPORTED THAT AROUND FEBRUARY THE PATIENT WAS DIAGNOSED WITH CAVITARY PNEUMONIA PRESUMABLY DUE TO VOLARA DEVICE NOT FUNCTIONING PROPERLY. IT WAS REPORTED THAT VOLARA DEVICE HAS ABOUT TWO MONTHS NOT FUNCTIONING PROPERLY, SOUNDING QUIETER, LESS STRONG AND DIFFERENT THAN IT USED TO. DAYS AFTER THE EVENT WAS REPORTED, THE VOLARA AND CIRCUIT WAS REPLACED BY THE TRAINER WHO NOTED THE PRESENCE OF WATER IN THE CIRCUIT AND THE NEBULIZER TUBING DURING THE RETRAINING SESSION. PATIENT'S PULMONOLOGIST INDICATED THE PATIENT HAD PUS IN A CAVITY IN HIS LUNG AND WAS PUT ON ANTIBIOTIC. THE PATIENT WAS GIVEN A NEW DEVICE AND THERAPY WITH THE VOLARA HAS BEEN CONTINUED AS PRESCRIBED. THE EXACT CAUSE OF THE EVENT IS UNKNOWN AT THE MOMENT AND NO FURTHER INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Pneumonia

    MAUDE report 24822656 ↗
  • MalfunctionMar 20, 2026Crack

    ON (B)(6) 2026, A PATIENT CONTACTED TECHNICAL SERVICE TO REPORT THAT VOLARA SYSTEM SERVICED, HC,NA (PRODUCT CODE M08594A, SERIAL NUMBER (B)(6)), HAD SMART FILTER CRACKED. THIS OCCURRED AT HOME DURING PATIENT USE. THERE WAS NO PATIENT INJURY OR DEATH REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24651026 ↗