Flexible Silicone Drill Driver 110010733 (UDI 00887868466014)Recalls & reported problems

Biomet Orthopedics, LLC · Class 1

All models of Flexible Silicone Drill Driver (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Biomet flexible silicone drill driver is used to drill holes during hip surgery. Most of the 28 reports describe the shaft cracking, snapping or twisting while drilling, and in a few cases a broken piece was found in the patient. There are no deaths and one injury. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 00887868466014), received between Jun 29, 2022 and Jun 11, 2026 — classified as 1 injury, 28 malfunctions.

The most frequently reported specific device problem is “Fracture” (27 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture27
  • Crack1
  • Difficult to Insert1
  • Material Split, Cut or Torn1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions28
  • Foreign Body In Patient1

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 11, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING A RIGHT TOTAL HIP PROCEDURE THE FLEXIBLE SILICONE DRILL DRIVER OUTER RUBBER SHEATH WAS NOTED TO BE BROKEN. THE ISSUE WAS NOTED INTRAOPERATIVELY. THERE WAS NO REPORT OF ADVERSE EVENT, PATIENT INJURY, OR PATIENT IMPACT ASSOCIATED WITH THE REPORTED DEVICE ISSUE. NO DELAY IN PROCEDURE, MEDICAL INTERVENTION, OR ADDITIONAL TREATMENT ATTRIBUTABLE TO THE EVENT WAS REPORTED. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25475045 ↗
  • MalfunctionApr 28, 2026Fracture

    IT WAS REPORTED THAT DURING ACETABULAR SCREW HOLE PREPARATION THE DRILL BROKE COMPLETELY OFF. INTRA-OPERATIVELY, A DIFFERENT DRILL WAS USED TO SUCCESSFULLY COMPLETE THE CASE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25016157 ↗
  • MalfunctionFeb 11, 2026Fracture

    IT WAS REPORTED THAT A FLEXIBLE DRIVER BROKE DURING THE PROCEDURE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24313353 ↗
  • MalfunctionFeb 4, 2026Fracture

    NO ADDITIONAL INFORMATION AVAILABLE FOR THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24257375 ↗
  • MalfunctionOct 28, 2025Fracture; Fracture

    IT WAS REPORTED THAT WHILE DRILLING HOLES, THE FLEXIBLE SILICONE DRILL DRIVER¿S SHAFT FRACTURED. THERE WAS NO PATIENT HARM OR IMPACT AS A RESULT. NO ADDITIONAL INFORMATION AVAILABLE FOR THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23406464 ↗
  • MalfunctionSep 3, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE FLEXIBLE DRILL DRIVER TWISTED AND DEFORMED WHILE DRILLING FOR A SCREW. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. IT WAS REPORTED NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22950928 ↗
  • MalfunctionAug 7, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE DRILL BIT SHAFT BROKE WHILE DRILLING HOLES IN THE CUP. THERE WAS NO PATIENT IMPACT, NO MEDICAL/SURGICAL INTERVENTIONS, NO SURGICAL COMPLICATIONS, NO SURGICAL DELAY, AND NO FOREIGN BODY RETAINED. THERE WERE ALSO NO CONTRIBUTING CONDITIONS RELATED TO THE REPORTED EVENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22738779 ↗
  • MalfunctionMay 6, 2025Fracture; Difficult to Insert; Fracture; Difficult to Insert

    IT WAS REPORTED THAT THE DRILL BIT DOESN¿T INSERT. THE DRILL BIT CONNECTION END OF THE FLEX DRIVER WAS FOUND TO HAVE A RETAINED FRACTURED SEGMENT DURING INVESTIGATION. THERE WAS NO REPORTED PATIENT INVOLVEMENT AND NO SURGICAL DELAY. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21971944 ↗
  • InjuryMar 20, 2025Fracture; Fracture

    IT WAS REPORTED THE SURGEON WAS DRILLING WITH THE FLEXIBLE DRIVER WHEN IT BROKE. SOME PIECES MAY HAVE FALLEN INTO THE PATIENT. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 21651313 ↗
  • MalfunctionNov 11, 2024Fracture; Fracture

    NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20652151 ↗
  • MalfunctionOct 24, 2024Fracture; Fracture

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20526315 ↗
  • MalfunctionOct 3, 2024Fracture; Fracture

    IT WAS REPORTED THAT BEFORE THE SURGERY IT WAS NOTICED THAT THE RUBBER ON THE FLEXIBLE DRILL SHAFT WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20359269 ↗
  • MalfunctionSep 17, 2024Fracture; Fracture

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20238096 ↗
  • MalfunctionAug 20, 2024Fracture; Fracture

    NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20004404 ↗
  • MalfunctionMay 10, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING THE SURGERY, THE SURGEON WAS TRYING TO DRILL A SCREW HOLE, BUT WITH THE RESISTANCE OF THE BONE, THE DRILL BIT WOULD NOT TURN. THE SURGEON OPENED A BACKUP TRAY TO USE ANOTHER SILICONE DRILL DRIVER. THIS TIME, WITH THE SAME DRILL BIT BUT A DIFFERENT DRILL DRIVER IT WORKED WELL. AFTER THE SURGERY, THE SILICONE DRILL DRIVER WAS EXAMINED, AND IT WAS NOTICED THAT THE INSTRUMENT IS BROKEN AT THE JUNCTION BETWEEN THE METAL AND THE SILICONE. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19285458 ↗
  • MalfunctionMay 6, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE RUBBER IS CRACKED ON THE DRILL DRIVER. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19247013 ↗
  • MalfunctionApr 30, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE FLEXIBLE DRIVER FROM THE G7 EFFICIENT TRAY BROKE WHILE DRILLING SCREW HOLE FOR THE CUP. THERE WAS NO PATIENT IMPACT OR DELAY IN THE CASE. SURGERY WAS COMPLETED WITH ANOTHER DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19213144 ↗
  • MalfunctionMar 4, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE FLEXIBLE SILICONE ON THE DRILL DRIVER IS STARTING TO CRACK. THE SILICONE WAS STARTING TO SPIN AND TWIST. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18829853 ↗
  • MalfunctionJan 9, 2024Fracture; Fracture

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18473819 ↗
  • MalfunctionDec 12, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE DRILL BROKE APART WHILE DRILLING. SOME PIECES FELL INTO THE PATIENT, BUT WILL WERE RECOVERED. THERE WAS NO HARM OR HEALTH CONSEQUENCES TO THE PATIENT. IT WAS REPORTED THAT NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18310766 ↗
  • MalfunctionDec 5, 2023Fracture; Fracture

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18265297 ↗
  • MalfunctionOct 26, 2023Fracture; Fracture

    THERE IS NO UPDATE TO THE PRIOR EVENT DESCRIPTION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18012631 ↗
  • MalfunctionJul 6, 2023Fracture; Fracture

    IT WAS REPORTED THE RUBBER IS COMING OFF THE FLEXIBLE SILICONE DRILL DRIVER. THERE WAS NO PATIENT INVOLVEMENT REPORTED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE, HOWEVER, NO MORE IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17271328 ↗
  • MalfunctionMar 7, 2023Fracture; Fracture

    IT WAS REPORTED THAT THE REAMER SHAFT BROKE WHILE DRILLING THROUGH HARD BONE. THE SURGERY WAS COMPLETED USING ANOTHER DEVICE. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16496323 ↗
  • MalfunctionFeb 27, 2023Crack; Fracture; Crack; Fracture

    IT WAS REPORTED THAT THE SILICON DRILL DRIVER RUBBER IS BROKEN. THERE WAS NO PATIENT INVOLVEMENT. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16444085 ↗

Other BIT, DRILL devices

All BIT, DRILL devices →