LEVEL ONE 25-452-00-91Recalls & reported problems
KLS-Martin L.P. · TWIST DRILL, J NOTCH, 17 MM STOP · Class 1
All models of LEVEL ONE (36)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This KLS-Martin twist drill is a surgical bit used to drill a starter hole with a J notch. The 30 reports mostly describe the drill tip snapping off during surgery, sometimes leaving a broken piece behind in the bone or tissue. No deaths were reported; all 30 reports describe injuries. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 29
- Deaths reported
- 0
- Injuries reported
- 29
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 00888118040954), received between May 16, 2018 and Jun 5, 2026 — classified as 29 injuries.
The most frequently reported specific device problem is “Break” (21 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break21
- Fracture8
- Use of Device Problem1
Patient problems
- Device Embedded In Tissue or Plaque23
- No Consequences Or Impact To Patient5
- Impaired Healing1
Recent adverse event reports
The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.
InjuryJun 5, 2026Fracture
IT WAS REPORTED THAT DURING INITIAL SURGERY A TWIST DRILL FRACTURED DURING USE. A PORTION OF IT REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 25412669 ↗InjuryApr 30, 2026Fracture
IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL FRACTURED DURING USE. A PORTION OF IT REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 25046693 ↗InjuryFeb 10, 2026Fracture
IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL FRACTURED DURING USE. A PORTION OF IT REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 24306763 ↗InjuryDec 9, 2025Fracture; Fracture
IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL FRACTURED DURING USE. A PORTION OF IT REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 23760864 ↗InjuryNov 5, 2025Fracture; Fracture
IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL FRACTURED DURING USE. A PORTION OF IT REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 23477570 ↗InjuryJul 30, 2025Fracture; Fracture
IT WAS REPORTED DURING INITIAL SURGERY THAT TWIST DRILLS FRACTURED DURING USE. A PORTION OF THEM REMAIN IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 22668280 ↗InjuryApr 24, 2025Fracture; Fracture
IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 21904081 ↗InjuryMar 5, 2025Fracture; Fracture
IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL FRACTURED DUE TO IT WAS NOT PLACED PERPENDICULAR TO THE BONE. A PORTION OF IT REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 21529165 ↗InjuryNov 15, 2024Break; Break
IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 20698810 ↗InjuryAug 22, 2024Break; Break
IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 20034211 ↗InjuryJun 5, 2024Break; Break
IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 19467629 ↗InjuryDec 13, 2023Break; Break
IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 18317865 ↗InjuryOct 26, 2023Break; Break
IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 18015921 ↗InjuryFeb 14, 2023Break; Break
IT WAS REPORTED THAT A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 16375261 ↗InjuryDec 14, 2022Break; Break
IT WAS REPORTED A TWIST DRILL BROKE DURING USE. A PORTION REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 15978815 ↗InjuryMar 4, 2022Break; Break
IT WAS REPORTED A TWIST DRILL BROKE. A PORTION REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 13672060 ↗InjuryJul 21, 2021Break; Break
IT WAS REPORTED THAT A TWIST DRILL BROKE, AND A PORTION OF IT REMAINS IMPLANTED. THE REPORTED RPMS EXCEEDED THE MAXIMUM RECOMMENDED SPEED FOUND IN THE IFU.
Patient problems: Impaired Healing
MAUDE report 12205941 ↗InjuryJul 2, 2021Break; Break
IT WAS REPORTED THAT A TWIST DRILL BROKE, AND A PORTION OF IT REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 12112000 ↗InjuryMay 21, 2021Break; Break
IT WAS REPORTED A TWIST DRILL BROKE AND A PORTION REMAINED IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 11866983 ↗InjuryMay 13, 2021Break; Break
IT WAS REPORTED A TWIST DRILL BROKE AND A PORTION REMAINS IMPLANTED. PER IFU 91-600-02-00 FOR THE TWIST DRILL "TWIST DRILLS ARE NOT DESIGNED TO BE USED AT SPEEDS IN EXCESS OF 2500 RPM."
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 11819367 ↗InjuryMar 26, 2020Break; Break
IT WAS REPORTED A TWIST DRILL BROKE AND A PORTION REMAINS IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9885078 ↗InjuryFeb 26, 2020Break; Break
IT WAS REPORTED A TWIST DRILL BROKE AND A PORTION REMAINED IMPLANTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9757021 ↗InjuryFeb 1, 2019Break; Break
IT WAS REPORTED THE TWIST DRILL BROKE. ALL PIECES WERE REMOVED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8298210 ↗InjuryJan 11, 2019Break; Break
IT WAS REPORTED A TWIST DRILL BROKE. ALL PIECES WERE REMOVED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8240545 ↗InjuryAug 17, 2018Break; Break
IT WAS REPORTED THAT A TWIST DRILL BROKE DURING USE. THE TWIST DRILL WAS REMOVED FROM THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7794006 ↗
FDA clearance and classification
The FDA regulates product code HTW (bit, drill) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the LEVEL ONE 25-452-00-91
Has the LEVEL ONE 25-452-00-91 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00888118040954), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the LEVEL ONE 25-452-00-91?
The FDA's MAUDE database lists 29 adverse event reports for this device, received between May 16, 2018 and Jun 5, 2026. In those reports, the most common specific device problems are “Break” (21), “Fracture” (8) and “Use of Device Problem” (1); the most common patient problems are “Device Embedded In Tissue or Plaque” (23) and “Impaired Healing” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the LEVEL ONE 25-452-00-91?
None of the reports are classified as deaths, 29 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the LEVEL ONE 25-452-00-91 FDA approved?
Devices of this type (product code HTW) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by KLS-Martin L.P.
Who makes the LEVEL ONE 25-452-00-91, and is it still sold?
The FDA's device database lists KLS-Martin L.P. as the company responsible for it (brand LEVEL ONE, model 25-452-00-91). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BIT, DRILL devices
- QUICKSET 2274-25-500Depuy Orthopaedics, Inc. · BIT, DRILL1,978 reports
- QUICKSET 2274-35-500Depuy Orthopaedics, Inc. · BIT, DRILL769 reports
- QUICKSET 2274-30-500Depuy Orthopaedics, Inc. · BIT, DRILL761 reports
- NA 03.037.022Synthes GmbH · BIT, DRILL335 reports
- NA 323.062Synthes GmbH · BIT, DRILL308 reports
- QUICKSET 2274-56-000Depuy (Ireland) · BIT, DRILL275 reports
- QUICKSET 2274-45-500Depuy Orthopaedics, Inc. · BIT, DRILL210 reports
- LCS 2366-83-000Depuy (Ireland) · BIT, DRILL204 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page