Skip to content

LEVEL ONE 25-452-07-91Recalls & reported problems

KLS-Martin L.P. · TWIST DRILL, J NOTCH, 7 MM STOP · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This KLS-Martin twist drill is a surgical drilling bit. Reports mostly describe the drill breaking during surgery with a piece left in the patient. There are 25 reports that name its identifier, with no deaths and 25 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
25
Deaths reported
0
Injuries reported
25
Recalls naming it
0

Summary

The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 00888118040985), received between May 7, 2018 and Jan 28, 2026 — classified as 25 injuries.

The most frequently reported specific device problem is “Break” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break19
  • Fracture6
  • Use of Device Problem1

Patient problems

  • Device Embedded In Tissue or Plaque22
  • Impaired Healing1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 28, 2026Fracture

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL FRACTURED DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 24203527 ↗
  • InjuryJan 21, 2026Fracture

    IT WAS REPORTED DURING INITIAL SURGERY THAT TWIST DRILLS FRACTURED DURING USE. A PORTION OF THEM REMAIN IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 24136427 ↗
  • InjuryDec 9, 2025Fracture; Fracture

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL FRACTURED DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 23760863 ↗
  • InjurySep 9, 2025Fracture; Fracture

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL FRACTURED DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 23010168 ↗
  • InjuryJul 30, 2025Fracture; Fracture

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL FRACTURED DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 22668283 ↗
  • InjuryMar 5, 2025Fracture; Fracture

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL FRACTURED DURING USE. A PORTION OF IT CONFIRMED ON X-RAY REMAINS FIRMLY IMPLANTED WITHOUT PLANS FOR REMOVAL.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 21529164 ↗
  • InjuryJan 14, 2025Break; Break

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 21148935 ↗
  • InjuryDec 27, 2024Break; Break

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 21029624 ↗
  • InjuryJul 18, 2024Break; Break

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 19772057 ↗
  • InjuryJan 17, 2024Break; Break

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 18530409 ↗
  • InjuryJun 30, 2023Break; Break

    IT WAS REPORTED THAT A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 17240057 ↗
  • InjuryDec 29, 2022Break; Break

    IT WAS REPORTED A TWIST DRILL BROKE DURING USE. A PORTION REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 16075306 ↗
  • InjurySep 23, 2022Break; Break

    IT WAS REPORTED A TWIST DRILL BROKE DURING USE. A PORTION REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 15473414 ↗
  • InjuryAug 27, 2021Break; Break

    IT WAS REPORTED THAT A TWIST DRILL BROKE, AND A PORTION OF IT REMAINS IMPLANTED. THE REPORTED RPMS EXCEEDED THE MAXIMUM RECOMMENDED SPEED FOUND IN THE IFU.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 12377772 ↗
  • InjuryAug 4, 2021Break; Break

    IT WAS REPORTED THAT A TWIST DRILL BROKE, AND A PORTION OF IT REMAINS IMPLANTED. THE REPORTED RPMS EXCEEDED THE MAXIMUM RECOMMENDED SPEED FOUND IN THE IFU.

    Patient problems: Impaired Healing

    MAUDE report 12274316 ↗
  • InjuryNov 12, 2020Break; Break

    IT WAS REPORTED A TWIST DRILL BROKE DURING USE AND IT IS SUSPECTED THAT A PORTION REMAINED IMPLANTED. THE REPORTED RPMS EXCEEDED THE MAXIMUM VALUE FOUND IN THE IFU. PER IFU 91-600-02-00 FOR THE TWIST DRILL "OPERATE THE TWIST DRILLS WITH NO MORE THAN THE MAXIMUM RECOMMENDED HANDPIECE SPEED SETTING OF 2,500 RPM."

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 10829432 ↗
  • InjuryOct 15, 2020Break; Break

    IT WAS REPORTED A TWIST DRILL BROKE DURING USE. A PORTION REMAINS IMPLANTED. THE REPORTED RPMS EXCEEDED THE MAXIMUM VALUE FOUND IN THE IFU. PER IFU 91-600-02-00 FOR THE TWIST DRILL "OPERATE THE TWIST DRILLS WITH NO MORE THAN THE MAXIMUM RECOMMENDED HANDPIECE SPEED SETTING OF 2,500 RPM."

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 10683424 ↗
  • InjuryAug 11, 2020Break; Break

    IT WAS REPORTED A TWIST DRILL BROKE AND A PORTION REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 10393725 ↗
  • InjuryJun 24, 2019Break; Break

    IT WAS REPORTED THAT A TWIST DRILL BROKE. A PORTION REMAINED IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8728525 ↗
  • InjuryJun 19, 2019Break; Break

    IT WAS REPORTED THAT A TWIST DRILL BROKE. A PORTION REMAINED IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8712038 ↗
  • InjuryMar 8, 2019Break; Break

    IT WAS REPORTED A TWIST DRILL BROKE AND A PORTION REMAINED IN THE BONE.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8404202 ↗
  • InjuryFeb 1, 2019Break; Use of Device Problem; Break; Use of Device Problem

    IT WAS REPORTED A TWIST DRILL BROKE DURING USE. ALL PIECES WERE REMOVED. PER IFU 90-274-92-40 FOR THE TWIST DRILL "TWIST DRILLS ARE NOT DESIGNED TO BE USED AT SPEEDS IN EXCESS OF 5000 RPM."

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8298204 ↗
  • InjuryDec 7, 2018Break; Break

    IT WAS REPORTED THAT A TWIST DRILL DIAMETER INDICATOR RING BROKE OFF DURING USE AND IT WAS NOT LOCATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8143596 ↗
  • InjuryJun 15, 2018Break; Break

    IT WAS REPORTED THE TWIST DRILL BROKE. DOCTOR IS UNSURE IF IT WAS IMPLANTED IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7607358 ↗
  • InjuryMay 7, 2018Break; Break

    IT WAS REPORTED THAT A TWIST DRILL BROKE AND A PORTION REMAINS IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7490240 ↗

Other BIT, DRILL devices

All BIT, DRILL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page