Skip to content

LEVEL ONE 25-454-07-91Recalls & reported problems

KLS-Martin L.P. · TWIST DRILL, J NOTCH, 7 MM STOP · Class 1

All models of LEVEL ONE (36)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00888118041067), received between Mar 12, 2018 and Jun 13, 2025 — classified as 7 injuries.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Fracture1
  • Material Fragmentation1

Patient problems

  • Device Embedded In Tissue or Plaque5
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 13, 2025Fracture; Fracture

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 22205090 ↗
  • InjuryFeb 13, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 21375086 ↗
  • InjuryOct 15, 2020Break; Break

    IT WAS REPORTED A TWIST DRILL BROKE DURING USE AND IT IS SUSPECTED THAT A PORTION REMAINED IMPLANTED. THE REPORTED RPMS EXCEEDED THE MAXIMUM VALUE FOUND IN THE IFU. PER IFU 91-600-02-00 FOR THE TWIST DRILL "OPERATE THE TWIST DRILLS WITH NO MORE THAN THE MAXIMUM RECOMMENDED HANDPIECE SPEED SETTING OF 2,500 RPM."

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 10683426 ↗
  • InjuryMay 11, 2020Break; Break

    IT WAS REPORTED THAT A TWIST DRILL BROKE AND A PORTION REMAINED IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 10042900 ↗
  • InjuryNov 5, 2018Break; Break

    IT WAS REPORTED THAT A TWIST DRILL BROKE DURING USE. ALL PIECES WERE REMOVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8038591 ↗
  • InjurySep 7, 2018Break; Break

    IT WAS REPORTED THAT A TWIST DRILL BROKE DURING USE. THE TWIST DRILL WAS REMOVED FROM THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7856078 ↗
  • InjuryMar 12, 2018Break; Break

    IT WAS REPORTED THAT A TWIST DRILL BROKE AND THE TIP REMAINS IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7330122 ↗

Other BIT, DRILL devices

All BIT, DRILL devices →

Sources and updates