Skip to content

LEVEL ONE 25-469-11-07Recalls & reported problems

KLS-Martin L.P. · TWIST DRILL, J NOTCH, 30 MM STOP, 25 MM FLUTE · Class 1

All models of LEVEL ONE (36)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This KLS-Martin twist drill is a surgical bone-cutting bit used during orthopedic procedures. The 33 reports mostly describe the drill tip breaking off during use, sometimes leaving a fragment in the patient, along with a few handling complaints. All 33 reports involve injuries, with no deaths reported. There are no recalls naming this device, and no reports came in during the past 12 months compared with 2 in the prior12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
33
Deaths reported
0
Injuries reported
33
Recalls naming it
0

Summary

The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 00888118041272), received between Aug 16, 2017 and Jun 26, 2025 — classified as 33 injuries.

The most frequently reported specific device problem is “Break” (30 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break30
  • Use of Device Problem4
  • Fracture3
  • Device Handling Problem1

Patient problems

  • Device Embedded In Tissue or Plaque19
  • No Consequences Or Impact To Patient12
  • Laceration(s)1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 26, 2025Fracture; Fracture

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL FRACTURED DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 22337778 ↗
  • InjuryMay 2, 2025Fracture; Fracture

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL FRACTURED DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 21954332 ↗
  • InjuryOct 25, 2024Break; Break

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 20532062 ↗
  • InjuryAug 6, 2024Break; Break

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 19916968 ↗
  • InjuryMar 20, 2024Break; Break

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 18946008 ↗
  • InjuryAug 25, 2023Break; Break

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 17624362 ↗
  • InjuryJul 21, 2023Break; Break

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 17374721 ↗
  • InjuryJul 7, 2023Break; Break

    IT WAS REPORTED THAT A TWIST DRILL BROKE, AND A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 17283054 ↗
  • InjuryJun 30, 2023Break; Break

    IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 17242087 ↗
  • InjuryMar 18, 2022Break; Break

    IT WAS REPORTED THAT A TWIST DRILL BROKE, AND A PORTION OF IT REMAINS IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 13804745 ↗
  • InjurySep 3, 2020Break; Break

    IT WAS REPORTED A TWIST DRILL BROKE DURING USE AND A PORTION REMAINS IMPLANTED. PER IFU 91-600-02-00 FOR THE TWIST DRILL "OPERATE THE TWIST DRILLS WITH NO MORE THAN THE MAXIMUM RECOMMENDED HANDPIECE SPEED SETTING OF 2,500 RPM."

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 10490906 ↗
  • InjuryJan 21, 2020Break; Break

    IT WAS REPORTED THAT A 115MM TWIST DRILL BROKE DURING USE AND A BURR GUARD OR DRILL GUIDE WAS NOT USED. A PORTION OF THE TWIST DRILL REMAINED IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 9610265 ↗
  • InjuryAug 13, 2019Break; Break

    IT WAS REPORTED THAT A TWIST DRILL BROKE AND A PORTION REMAINED IMPLANTED. PER IFU 90-274-92-40 FOR THE TWIST DRILL "TWIST DRILLS ARE NOT DESIGNED TO BE USED AT SPEEDS IN EXCESS OF 5000 RPM."

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8888079 ↗
  • InjuryApr 9, 2019Break; Device Handling Problem; Break; Device Handling Problem

    IT WAS REPORTED TWIST DRILLS BROKE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8496397 ↗
  • InjuryMar 18, 2019Break; Break

    IT WAS REPORTED A TWIST DRILL BROKE. ALL PIECES WERE REMOVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8428560 ↗
  • InjuryFeb 25, 2019Break; Break

    IT WAS REPORTED THE TWIST DRILL BROKE. ALL PIECES WERE REMOVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8367373 ↗
  • InjuryFeb 1, 2019Break; Break

    IT WAS REPORTED THE TWIST DRILL BROKE. ALL PIECES WERE REMOVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8298211 ↗
  • InjuryFeb 1, 2019Break; Use of Device Problem; Break; Use of Device Problem

    IT WAS REPORTED A TWIST DRILL BROKE DURING USE. ALL PIECES WERE REMOVED. PER IFU 90-274-92-40 FOR THE TWIST DRILL "TWIST DRILLS ARE NOT DESIGNED TO BE USED AT SPEEDS IN EXCESS OF 5000 RPM."

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8298203 ↗
  • InjuryDec 3, 2018Break; Use of Device Problem; Break; Use of Device Problem

    IT WAS REPORTED THAT A TWIST DRILL BROKE DURING USE. A PORTION REMAINED IMPLANTED. PER IFU (B)(4) FOR THE TWIST DRILL "TWIST DRILLS ARE NOT DESIGNED TO BE USED AT SPEEDS IN EXCESS OF 5000 RPM."

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8126150 ↗
  • InjuryNov 16, 2018Break; Break

    IT WAS REPORTED A TWIST DRILL BROKE DURING USE. A PORTION REMAINED IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8077751 ↗
  • InjuryNov 5, 2018Break; Break

    IT WAS REPORTED THAT A TWIST DRILL BROKE DURING USE. A PORTION REMAINED IMPLANTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8038596 ↗
  • InjuryNov 5, 2018Break; Break

    IT WAS REPORTED THAT A BROKEN TWIST DRILL WAS REMOVED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8038571 ↗
  • InjuryNov 5, 2018Break; Break

    IT WAS REPORTED THAT A TWIST DRILL BROKE DURING USE. ALL PIECES WERE REMOVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8038561 ↗
  • InjuryOct 15, 2018Break; Break

    IT WAS REPORTED A TWIST DRILL BROKE DURING USE. A PORTION REMAINED IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7966119 ↗
  • InjuryOct 5, 2018Break; Use of Device Problem; Break; Use of Device Problem

    IT WAS REPORTED THAT A TWIST DRILL BROKE DURING USE. THE TWIST DRILL WAS REMOVED FROM THE PATIENT. THE TWIST DRILL WAS BEING USED AT 20,000 RMPS. IT IS STATED IN IFU 90-274-92-40 "FOR TWIST DRILLS WITH AN OVERALL LENGTH OF GREATER THAN 50MM, AN APPROPRIATE BURR GUARD OR DRILL GUIDE SHOULD BE USED AT ALL TIMES."

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7939734 ↗

Other BIT, DRILL devices