LEVEL ONE 25-472-75-07Recalls & reported problems
KLS-Martin L.P. · TWIST DRILL, J NOTCH, MORRISON, 25 MM STOP, FOR 25-471-05-07 · Class 1
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888118041432), received between Sep 28, 2018 and Jan 4, 2019 — classified as 2 injuries.
The most frequently reported specific device problem is “Temperature Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Temperature Problem2
- Use of Device Problem1
Patient problems
- Burn(s)2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 4, 2019Temperature Problem; Temperature Problem
IT WAS REPORTED A PATIENT WAS BURNED WHILE USING A TWIST DRILL AND TROCAR.
Patient problems: Burn(s)
MAUDE report 8217430 ↗InjurySep 28, 2018Use of Device Problem; Temperature Problem; Use of Device Problem; Temperature Problem
IT WAS REPORTED A PATIENT WAS BURNED WHILE USING A TWIST DRILL AND TROCAR. THE TWIST DRILL WAS BEING USED AT 20,000 RMPS. IT IS STATED IN IFU 90-274-92-40 "TWIST DRILLS ARE NOT DESIGNED TO BE USED AT SPEEDS IN EXCESS OF 5000 RPM."
Patient problems: Burn(s)
MAUDE report 7919303 ↗
Other BIT, DRILL devices
- QUICKSET 2274-25-500Depuy Orthopaedics, Inc. · BIT, DRILL1,985 reports
- QUICKSET 2274-35-500Depuy Orthopaedics, Inc. · BIT, DRILL770 reports
- QUICKSET 2274-30-500Depuy Orthopaedics, Inc. · BIT, DRILL761 reports
- NA 03.037.022Synthes GmbH · BIT, DRILL339 reports
- NA 323.062Synthes GmbH · BIT, DRILL309 reports
- QUICKSET 2274-56-000Depuy (Ireland) · BIT, DRILL275 reports
- QUICKSET 2274-45-500Depuy Orthopaedics, Inc. · BIT, DRILL211 reports
- LCS 2366-83-000Depuy (Ireland) · BIT, DRILL207 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.