Skip to content

LEVEL ONE 25-473-70-07Recalls & reported problems

KLS-Martin L.P. · TWIST DRILL, CYLINDRICAL, MORRISON, 25 MM STOP, FOR 50-501-XX-07 · Class 1

All models of LEVEL ONE (36)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888118041548), received between Feb 1, 2018 and Feb 12, 2019 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Break2
  • Device Handling Problem2
  • Temperature Problem2
  • Human-Device Interface Problem1

Patient problems

  • Burn(s)2
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 12, 2019Temperature Problem; Device Handling Problem; Temperature Problem; Device Handling Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED HEAT BURNS ON THEIR CHEEK. THE TWIST DRILL WAS BEING USED AT 30,000 RPM WITH A TROCAR. IT IS STATED IN IFU 90-274-92-40 AND 91-600-00-41 UNDER PRECAUTIONS "WHEN USING TWIST DRILLS, COPIOUS AMOUNTS OF IRRIGATION ARE MANDATORY TO ALLEVIATE EXCESSIVE HEAT GENERATION, WHICH CAN LEAD TO BONE NECROSIS, AND IN EXTREME CASES, SOFT TISSUE BURNS THROUGH CONTACT WITH TROCAR AND CANNULA." AND "TWIST DRILLS ARE NOT DESIGNED TO BE USED AT SPEEDS IN EXCESS OF 5000 RPM."

    Patient problems: Burn(s)

    MAUDE report 8332125 ↗
  • InjuryJun 8, 2018Break; Break

    IT WAS REPORTED THAT TWO TWIST DRILLS BROKE. ALL PIECES WERE REMOVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7584209 ↗
  • InjuryFeb 15, 2018Temperature Problem; Device Handling Problem; Temperature Problem; Device Handling Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED HEAT BURNS BILATERALLY ON THEIR CHEEKS. THE TWIST DRILL WAS BEING USED AT 35,000 RPM WITH A TROCAR. IT IS STATED IN IFU 90-274-92-40 AND 91-600-00-41 UNDER PRECAUTIONS "WHEN USING TWIST DRILLS, COPIOUS AMOUNTS OF IRRIGATION ARE MANDATORY TO ALLEVIATE EXCESSIVE HEAT GENERATION, WHICH CAN LEAD TO BONE NECROSIS, AND IN EXTREME CASES, SOFT TISSUE BURNS THROUGH CONTACT WITH TROCAR AND CANNULA." AND "TWIST DRILLS ARE NOT DESIGNED TO BE USED AT SPEEDS IN EXCESS OF 5000 RPM." THE BURNT TISSUE WAS CUT OUT LIKE A SCAR REVISION.

    Patient problems: Burn(s)

    MAUDE report 7272819 ↗
  • MalfunctionFeb 1, 2018Break; Human-Device Interface Problem; Break; Human-Device Interface Problem

    THE END OF THE TWIST DRILL BROKE OFF AND STUCK INTO THE SLEEVE OF THE SCRUB TECH'S GOWN. THE SKIN WAS NOT PUNCTURDED AND MEDICAL CARE WAS NOT NEEDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7234824 ↗

FDA clearance and classification

The FDA regulates product code HTW (bit, drill) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the LEVEL ONE 25-473-70-07

Has the LEVEL ONE 25-473-70-07 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00888118041548), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the LEVEL ONE 25-473-70-07?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Feb 1, 2018 and Feb 12, 2019. In those reports, the most common specific device problems are “Break” (2), “Device Handling Problem” (2) and “Temperature Problem” (2); the most common patient problem is “Burn(s)” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the LEVEL ONE 25-473-70-07?

None of the reports are classified as deaths, 3 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the LEVEL ONE 25-473-70-07 FDA approved?

Devices of this type (product code HTW) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by KLS-Martin L.P.

Who makes the LEVEL ONE 25-473-70-07, and is it still sold?

The FDA's device database lists KLS-Martin L.P. as the company responsible for it (brand LEVEL ONE, model 25-473-70-07). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other BIT, DRILL devices

All BIT, DRILL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page