Skip to content

LEVEL ONE 50-515-05-07Recalls & reported problems

KLS-Martin L.P. · TWIST DRILL, J NOTCH, 5 MM STOP · Class 1

All models of LEVEL ONE (36)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888118077264), received between Sep 24, 2018 and Sep 24, 2018 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1
  • Use of Device Problem1

Patient problems

  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 24, 2018Break; Use of Device Problem; Break; Use of Device Problem

    IT WAS REPORTED THAT A TWIST DRILL BROKE DURING USE AND A PORTION REMAINED IMPLANTED. TWIST DRILL WAS BEING USED AT 40,000 RPM. IT IS STATED IN IFU 90-274-92-40 "TWIST DRILLS ARE NOT DESIGNED TO BE USED AT SPEEDS IN EXCESS OF 5000 RPM."

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7901647 ↗

Other BIT, DRILL devices

All BIT, DRILL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page