Arthrex® AR-1934BCF-2Recalls & reported problems

Arthrex, Inc. · SUTR ANCH,BIO-COMP S-TAK · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Arthrex Bio-Comp S-TAK suture anchor appears in 89 adverse event reports, with no deaths and 6 injuries. Reports mostly describe the anchor breaking, and anchors coming loose or moving out of place. Some patient problems listed include foreign bodies and infection. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
87
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 87 adverse event reports that list this exact device (DI 00888867023888), received between Nov 23, 2016 and Jul 31, 2026 — classified as 6 injuries, 81 malfunctions.

The most frequently reported specific device problem is “Break” (69 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 3 reportsFeb 2024: 9 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 3 reportsJul 2024: 2 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 4 reportsDec 2024: 1 reportsJan 2025: 4 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 10 reportsJul 2025: 1 reportsAug 2025: 3 reportsSep 2025: 3 reportsOct 2025: 2 reportsNov 2025: 1 reportsDec 2025: 9 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 6 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break69
  • Device Dislodged or Dislocated10
  • Adverse Event Without Identified Device or Use Problem2
  • Positioning Failure2
  • Detachment of Device or Device Component1
  • Failure to Advance1
  • Fitting Problem1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions67
  • Insufficient Information15
  • Foreign Body In Patient4
  • Hypersensitivity/Allergic reaction2
  • Unspecified Infection2
  • Abscess1
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Cyst(s)1

Recent adverse event reports

The 25 most recent of 87 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 31, 2026Break; Device Dislodged or Dislocated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50099525 ↗
  • MalfunctionJul 31, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50099302 ↗
  • MalfunctionMar 30, 2026Break

    ON 10TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING AN ANKLE JOINT INSTABILITY SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24715232 ↗
  • MalfunctionMar 30, 2026Break

    ON 10TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING AN ANKLE JOINT INSTABILITY SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24714120 ↗
  • MalfunctionMar 30, 2026Break

    ON 10TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING AN ANKLE JOINT INSTABILITY SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24713871 ↗
  • MalfunctionMar 30, 2026Positioning Failure

    ON 11TH MARCH 2206, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWI RE® AND #2 TIGERWIRE®, AFTER THE ANCHOR WAS IMPLANTED INTO THE LABRUM, THE SURGEON TIED THE LABRUM UP AND THEN EXPLORED THE ACETABULUM. HE FOUND THAT SOME OF THE ANCHOR WAS EXPOSED INSIDE THE HIP, SO HAD TO EXTRACT THE ANCHOR AND INSERT ANOTHER ONE. THIS WAS DETECTED DURING A HIP ARTHROSCOPIC LABRAL REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW ANCHOR. BONE QUALITY OF PATIENT WAS NORMAL.THERE WAS NO PATIENT HARM AND NO SECONDARY SURGERY NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24712814 ↗
  • MalfunctionMar 26, 2026Break

    ON 6TH MARCH 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A ROTATOR CUFF REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698064 ↗
  • MalfunctionMar 25, 2026Break

    ON (B)(6) 2026, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, THE ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A HIP JOINT LABRUM REPAIR SURGERY ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AS THE DEVICE WAS CHANGED FOR A NEW ONE. AR-1949 AND AR-1250LT WERE USED TO PREPARE THE BONE SOCKET AND THE BONE QUALITY OF THE PATIENT WAS HARD. NO FRAGMENTS WERE LEFT IN THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 24686203 ↗
  • MalfunctionFeb 25, 2026Break

    ON 9TH FEB 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A PATELLAR DISLOCATION REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442035 ↗
  • MalfunctionFeb 13, 2026Break

    ON 27TH JAN 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A GLENOID LABRUM INJURY REPAIR SURGERY ON (B)(6) 2026. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24331568 ↗
  • MalfunctionFeb 10, 2026Positioning Failure

    ON (B)(6) 2026, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWI RE® AND #2 TIGERWIRE®, ITS ANCHOR WAS NOT STABLE AFTER INSERTION. THIS WAS DETECTED DURING AN ANKLE JOINT REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24302976 ↗
  • MalfunctionDec 22, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON 1ST DEC 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, ITS ANCHOR WAS PULLED OUT WHEN RETRIEVING THE POLE. THIS WAS DETECTED DURING A REPAIR OF THE ANTERIOR CALCANEOFIBULAR LIGAMENT SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23862700 ↗
  • MalfunctionDec 22, 2025Break; Break

    ON 1ST DEC 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A HIP JOINT LABRUM REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23862626 ↗
  • MalfunctionDec 22, 2025Break; Break

    ON 1ST DEC 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 2 UNITS OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, BOTH ITS ANCHORS BROKE DURING INSERTION. THIS WAS DETECTED DURING A HIP JOINT LABRUM REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING THE THIRD NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23862555 ↗
  • MalfunctionDec 22, 2025Break; Break

    ON 1ST DEC 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 2 UNITS OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, BOTH ITS ANCHORS BROKE DURING INSERTION. THIS WAS DETECTED DURING A HIP JOINT LABRUM REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING THE THIRD NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23862507 ↗
  • MalfunctionDec 22, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON (B)(6) 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, ITS ANCHOR WAS PULLED OUT WHEN RETRIEVING THE POLE. THIS WAS DETECTED DURING A LABRAL SUTURING SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23862376 ↗
  • MalfunctionDec 22, 2025Break; Break

    ON (B)(6) 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, THE END OF THE ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A LABRAL SUTURING SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23862318 ↗
  • MalfunctionDec 18, 2025Break; Break

    ON 28TH NOV 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A CALCANEAL OSTEOTOMY SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23837138 ↗
  • MalfunctionDec 8, 2025Break; Break

    ON 11/17/2025, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA (B)(4) THAT AN AR-1934BCF-2 BIOCOMPOSITE¿ SUTURETAK ANCHOR WAS DAMAGED DURING INSERTION. THE BASE OF THE SUTURES BROKE, MAKING IT UNUSABLE. THIS WAS DISCOVERED DURING A BICEPS LESION PROCEDURE WITH NO PATIENT HARM. THE CASE WAS COMPLETED WITH ANOTHER DEVICE. ADDITIONAL INFORMATION PROVIDED ON 11/26/2025: IT WAS FURTHER REPORTED THAT THE HOLE FOR THE ANCHOR WAS MADE USING A SUTURETAK ANCHOR DRILL AND THE DOCTOR USED THE MATERIAL CORRECTLY, THE PATIENT'S BONE WAS VERY HARD, WHICH CAUSED THE ANCHOR TO BREAK, NOT THE DRILL. THERE WERE NO LOOSE FRAGMENTS REPORTED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23742056 ↗
  • MalfunctionDec 4, 2025Break; Break

    ON 17TH NOV 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A REPAIR OF THE DELTOID LIGAMENT SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23717447 ↗
  • MalfunctionNov 5, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON (B)(6) 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, THE IMPLANT WAS PULLED OUT AFTER STITCHING. THIS WAS DETECTED DURING AN ANKLE SCOPE; LAT LIG STABILIZATION PROCEDURE ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING OTHER IMPLANT TO COMPLETE THE STITCHING.

    Patient problems: Insufficient Information

    MAUDE report 23478041 ↗
  • MalfunctionOct 20, 2025Break; Break

    ON 14TH OCT 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A FRACTURE OF THE RIGHT TIBIAL PLATEAU SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23335269 ↗
  • MalfunctionOct 20, 2025Break; Break

    ON 14TH OCT 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, THE ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A FRACTURE OF THE RIGHT TIBIAL PLATEAU SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23335199 ↗
  • MalfunctionSep 19, 2025Break; Break

    ON 2ND SEP 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, THE END OF THE ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A SHOULDER INSTABILITY SURGERY REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23095793 ↗
  • MalfunctionSep 19, 2025Break; Break

    ON 2ND SEP 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1934BCF-2, SUTURE ANCHOR, BIOCOMPOSITE¿ SUTURETAK® 3 X 14.5 MM, VE5 WITH #2 FIBERWIRE® AND #2 TIGERWIRE®, THE END OF THE ANCHOR WAS BROKEN. THIS WAS DETECTED DURING A SHOULDER INSTABILITY SURGERY REPAIR SURGERY ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ANOTHER NEW AR-1934BCF-2. THERE WERE NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED. ADDITIONAL INFORMATION RECEIVED ON 10TH SEP 2025: THE ANCHOR WAS BROKEN AFTER INSERTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23095690 ↗

FDA clearance and classification

The FDA regulates product code MBI (fastener, fixation, nondegradable, soft tissue) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Arthrex® AR-1934BCF-2

Has the Arthrex® AR-1934BCF-2 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00888867023888), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Arthrex® AR-1934BCF-2?

The FDA's MAUDE database lists 87 adverse event reports for this device, received between Nov 23, 2016 and Jul 31, 2026. In those reports, the most common specific device problems are “Break” (69), “Device Dislodged or Dislocated” (10) and “Positioning Failure” (2); the most common patient problems are “Foreign Body In Patient” (4), “Hypersensitivity/Allergic reaction” (2) and “Unspecified Infection” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Arthrex® AR-1934BCF-2?

None of the reports are classified as deaths, 6 as injuries and 81 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Arthrex® AR-1934BCF-2 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.

Who makes the Arthrex® AR-1934BCF-2, and is it still sold?

The FDA's device database lists Arthrex, Inc. as the company responsible for it (brand Arthrex®, model AR-1934BCF-2). Its record shows it as in commercial distribution.