Skip to content

Arthrex® AR-2922PSRecalls & reported problems

Arthrex, Inc. · SUTURE ANCHOR, 2.4MM PUSHLOCK , PEEK · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867027992), received between Jan 9, 2023 and Dec 24, 2025 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Material Fragmentation1

Patient problems

  • Failure of Implant2
  • Foreign Body In Patient2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 24, 2025Break; Break

    IT WAS REPORTED THAT DURING A SHOULDER STABILIZATION SURGERY THE ANCHOR BROKE OFF DURING INSERTION. NO PART OF THE DEVICE REMAINED INSIDE THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23888179 ↗
  • MalfunctionJul 22, 2024Break; Break

    ON 12/20/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT (3) AR-2922PS PEEK PUSHLOCK ANCHORS BROKE AT THE SAME SPOT EACH TIME. THIS WAS DISCOVERED DURING A REVISION POSTERIOR LABRAL REPAIR ON (B)(6) 2022. ADDITIONAL INFORMATION RECEIVED ON 12/20/2022: ALL BROKEN FRAGMENTS FROM THE THREE PUSHLOCK'S WERE REMOVED. CASE WAS COMPLETED WITH ANOTHER AR-2922PS PEEK PUSHLOCK. DURING THIS REVISION SURGERY, ONLY THE PUSHLOCK WAS IMPLANTED, AND NOTHING WAS EXPLANTED. SALES REPRESENTATIVE CANNOT CONFIRM ANY INFORMANT REGARDING THE ORIGINAL SURGERY, AS IT WAS PERFORMED BY ANOTHER SURGEON. NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 19792370 ↗
  • MalfunctionFeb 16, 2023Break

    IT WAS REPORTED THAT DURING A LABRUM REPAIR SURGERY ON THE HIP THE PEEK TIP BROKE OFF BY ITSELF AND COULD NOT BE RELOADED. THE BROKEN PIECE WAS RETRIEVED FROM PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: Foreign Body In Patient

    MAUDE report 16385139 ↗
  • InjuryJan 9, 2023Break

    ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT (3) AR-2922PS PEEK PUSHLOCK ANCHORS BROKE AT THE SAME SPOT EACH TIME. THIS WAS DISCOVERED DURING A REVISION POSTERIOR LABRAL REPAIR ON (B)(6) 2022. ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2022: ALL BROKEN FRAGMENTS FROM THE THREE PUSHLOCK'S WERE REMOVED. CASE WAS COMPLETED WITH ANOTHER AR-2922PS PEEK PUSHLOCK. DURING THIS REVISION SURGERY, ONLY THE PUSHLOCK WAS IMPLANTED, AND NOTHING WAS EXPLANTED. SALES REPRESENTATIVE CANNOT CONFIRM ANY INFORMANT REGARDING THE ORIGINAL SURGERY, AS IT WAS PERFORMED BY ANOTHER SURGEON.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant

    MAUDE report 16120380 ↗
  • InjuryJan 9, 2023Break; Material Fragmentation

    ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT (3) AR-2922PS PEEK PUSHLOCK ANCHORS BROKE AT THE SAME SPOT EACH TIME. THIS WAS DISCOVERED DURING A REVISION POSTERIOR LABRAL REPAIR ON (B)(6) 2022. ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2022: ALL BROKEN FRAGMENTS FROM THE THREE PUSHLOCK'S WERE REMOVED. CASE WAS COMPLETED WITH ANOTHER AR-2922PS PEEK PUSHLOCK. DURING THIS REVISION SURGERY, ONLY THE PUSHLOCK WAS IMPLANTED, AND NOTHING WAS EXPLANTED. SALES REPRESENTATIVE CANNOT CONFIRM ANY INFORMANT REGARDING THE ORIGINAL SURGERY, AS IT WAS PERFORMED BY ANOTHER SURGEON.

    Patient problems: Failure of Implant; Foreign Body In Patient

    MAUDE report 16120376 ↗

Other Fastener, fixation, nondegradable, soft tissue devices

All Fastener, fixation, nondegradable, soft tissue devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page