Skip to content

Arthrex® AR-5100-07Recalls & reported problems

Arthrex, Inc. · GRAFTBOLT, TIBIAL 7MM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867035744), received between Feb 1, 2021 and Jun 26, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 26, 2025Break; Break

    ON 5TH JUNE 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA EMAIL THAT FOR AN AR-5100-07, GRAFTBOLT®, TIBIAL, 7 MM, ITS ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING RECONSTRUCTION OF POSTERIOR FORK LIGAMENT SURGERY ON (B)(6) 2025. THE CASE WAS COMPLETED SUCCESSFULLY BY CHANGING TO A NEW AR-5100-07. THERE WAS NO PATIENT HARM, AND NO SECONDARY PROCEDURE NEEDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22336819 ↗
  • MalfunctionJan 14, 2025Break; Break

    ON 24TH DECEMBER 2024, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-5100-07, GRAFTBOLT®, TIBIAL, 7 MM, THE ANCHOR BROKE DURING INSERTION. THIS WAS DETECTED DURING A RECONSTRUCTION OF POSTERIOR FORK LIGAMENT SURGERY ON (B)(6) 2024. AR-5107 WAS USED TO PREPARE THE BONE SOCKET. THE ANCHOR BROKE DURING INSERTION AND NO FRAGMENTS WERE LEFT IN THE PATIENT. PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED BY USING ANOTHER NEW AR-5100-07.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21149327 ↗
  • MalfunctionSep 18, 2024Break; Break

    ON 4TH SEPTEMBER 2024, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-5100-07, GRAFTBOLT®, TIBIAL, 7 MM, THE SHEATH BROKE DURING INSERTION. THIS WAS DETECTED DURING USE IN AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY ON (B)(6) 2024. THE CASE WAS COMPLETED SUCCESSFULLY USING ANOTHER NEW AR-5100-07. ALL FRAGMENTS WERE RETRIEVED FROM PATIENT. THERE WAS NO PATIENT HARM AND NO SECONDARY PROCEDURE NEEDED. AR-5108 WAS USED TO PREPARE THE BONE SOCKET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20244848 ↗
  • MalfunctionFeb 1, 2021Break

    IT WAS REPORTED THAT DURING A PROCEDURE, WHEN THE SURGEON WAS INSERTING THE TIBIAL GRAFTBOLT, AR-5100-07, THE SHEATH OF THE SCREW BROKE UP. THE SCREW WAS REMOVED FROM THE PATIENT AND THE CASE WAS COMPLETED WITH A SECOND TIBIAL GRAFTBOLT. NO ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11259003 ↗

Other FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page