The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867041479), received between May 6, 2020 and Nov 23, 2020 — classified as 2 injuries.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break1
- Device Dislodged or Dislocated1
- Loosening of Implant Not Related to Bone-Ingrowth1
- Material Twisted/Bent1
Patient problems
- Fall1
- Pain1
- Unspecified Infection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 23, 2020Break; Material Twisted/Bent
IT WAS REPORTED THAT DURING A (B)(6) 2020 REVISION REVERSE TOTAL SHOULDER ARTHROPLASTY PROCEDURE THE FOLLOWING ISSUES OCCURRED WITH TWO ARTHREX INSTRUMENTS: INSTRUMENT 1: THE MEDIUM T-15 DRIVER (AR-9545-T15-02 #8001928) HAS BEEN COMPROMISED. WHEN TRYING TO EXTRACT THE CENTRAL SCREW FROM A UNIVERSAL GLENOID, THE TIP OF THE DRIVER BECAME TWISTED AND A SMALL PIECE BROKE OFF INTO THE HEAD OF THE CENTRAL SCREW. THERE WAS A 5 MINUTE DELAY IN SURGERY AS THE SCREW HEAD COULD NO LONGER BE ACCESSED. THE SURGEON USED OSTEOTOMES AROUND THE PERIPHERY OF THE BASEPLATE TO EVENTUALLY EXTRACT THE BASEPLATE AND CENTRAL SCREW AS ONE PIECE. THE INSTRUMENT BROKE INTO TWO PIECES. BOTH PIECES WERE RECOVERED, BUT THE SMALLER PIECE WAS LODGED INTO THE CENTRAL SCREW IMPLANT WHICH WAS NOT AVAILABLE FOR RETURN. THERE WAS NO UNPLANNED INCISION. A SECOND SURGERY IS NOT NECESSARY. DEVICE WILL BE RETURNING FOR EVALUATION. THE EVENT DID NOT OCCUR DURING THE INITIAL USE OF THE DEVICE. INSTRUMENT 2: THE TRINITY BASEPLATE IMPACTOR (AR-9165CDG #(10)051917) HAS BEEN COMPROMISED. THE SURGEON ENGAGED THE ALREADY IMPLANTED BASEPLATE WITH THE IMPACTOR SO THAT IMPACTION MIGHT DISRUPT THE BONY BOND WITH THE BASEPLATE. AFTER
Patient problems: Pain; Unspecified Infection
MAUDE report 10884233 ↗InjuryMay 6, 2020Device Dislodged or Dislocated; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED DURING A REVISION REVERSE SHOULDER TO A CLOSED REDUCTION PROCEDURE ON (B)(6) 2020, THE SURGEON WAS CHECKING IF THE PERIPHERAL SCREWS OF THE 24MM +2 MGS BASEPLATE WERE LOOSE. WHEN THE SURGEON TIGHTENED THE SCREWS, THE BLUE HANDLE OF THE AR-1999HH, RATCHETING HANDLE WAS CLICKING AND DID NOT HAVE A POSITIVE STOP. THE SURGEON DETERMINED THAT THE MGS BASEPLATE WAS LOOSE, SO THEY PROCEEDED TO EXPLANT THE SCREWS AND BASEPLATE WITHOUT ANY ISSUES. THE MGS BASEPLATE WAS IMPLANTED PRIOR TO THE REVISION SURGERY. THE IMPLANTS WERE CONVERTED TO TORNIER. AT THE TIME OF INITIAL REPORT TO ARTHREX NO FURTHER DETAILS WERE AVAILABLE. ADDITIONAL INFORMATION OBTAINED 4/13/2020 AND 4/15/2020: THE INITIAL REPORT PROVIDED WAS NOT A COMPLETE DESCRIPTION AND FURTHER, MORE DETAILED, INFORMATION HAS BEEN OBTAINED TO CLARIFY THE ENTIRE PATIENT RELATED EVENTS. THE FEMALE PATIENT UNDERWENT A PRIMARY REVERSE TOTAL SHOULDER PROCEDURE ON 10/8/19. IT WAS REPORTED THAT THE PATIENT HAD SUFFERED A FALL (DATE UNKNOWN) WHICH RESULTED IN A PERIPROSTHETIC HUMERAL FRACTURE. THE PATIENT UNDERWENT A FIRST REVISION ON 3/5/20 WHERE THE SURGEON EXPLANTED THE FOLLOWING DEVICES AND REPLACED THEM WITH ANOTHER MANUFAC
Patient problems: Fall
MAUDE report 10028413 ↗
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices
- FAST-FIX 72202468Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 death6,626 reports
- FAST-FIX 72201491Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE863 reports
- FAST-FIX 72202469Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE561 reports
- FAST-FIX 72205676Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE439 reports
- Arthrex® AR-4501Arthrex, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE403 reports
- FAST-FIX 72205324Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE383 reports
- LUPINE ORTHOCORD 210712Medos International Sàrl · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE357 reports
- FAST-FIX 72201494Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE249 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page