Arthrex® AR-7234Recalls & reported problems

Arthrex, Inc. · FiberLoop, #2, Blue · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867041677), received between Jan 10, 2018 and Nov 26, 2024 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn4
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Detachment Of Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Device Embedded In Tissue or Plaque1
  • Nerve Damage1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 26, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    (B)(6)2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-1370C BIOCOMPOSITE INTERFERENCE SCREW AND AN AR-4020C-07 FASTTHREAD BIOCOMPOSITE INTERFERENCE SCREW RIPPED THE SUTURE TIED TO THE GRAFT DURING THE FIXATION OF THE GRAFT TO THE FEMUR. THIS CAUSED THE GRAFT TO HAVE TO BE WHIP-STITCHED MULTIPLE TIMES. THE CASE WAS COMPLETED USING A BIOTENODESIS SCREW 8X23MM. THIS WAS DISCOVERED DURING AN MPFL RECONSTRUCTION PROCEDURE ON (B)(6)2024. ADDITIONAL INFORMATION RECEIVED ON 11/6/2024: THE SUTURE THAT BROKE WAS AN AR-7234 FIBERLOOP. THE CASE SUTURE WAS REMOVED AND REPLACED WITH ANOTHER AR-7234 FIBERLOOP. THE ANCHORS WERE ALSO REMOVED, AND AN AR-1580BC BIOCOMPOSITE-TENODESIS SCREW WAS USED. THERE WAS NO DELAY IN THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20788096 ↗
  • MalfunctionJun 27, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION WITH SEMITEDINOSUS TENDON THE DEVICE FELL APART. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19630043 ↗
  • MalfunctionJun 27, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION WITH SEMITEDINOSUS TENDON THE DEVICE FELL APART. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19629999 ↗
  • MalfunctionJun 27, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION WITH SEMITEDINOSUS TENDON THE DEVICE FELL APART. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19629951 ↗
  • InjuryAug 1, 2022Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED ON (B)(6) 2022 BY A FACILITY REPRESENTATIVE VIA PHONE THAT AN AR-7234 FIBERLOOP, AR-2261 BICEPS BUTTON AR-2262 BUTTON INSERTER AND AR-2263 DRILL WERE INVOLVED IN A DISTAL BICEPS REPAIR, THE PATIENT SUFFERED A CONTINUING NERVE ISSUE AFTER THE SURGERY. REVISION SURGERY (B)(6) 2022 TO REMOVE BUTTON AND PUT IN ANCHOR.

    Patient problems: Nerve Damage

    MAUDE report 15137723 ↗
  • InjuryJan 10, 2018Break; Detachment Of Device Component

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR THE TIP OF THE AR-13991N NEEDLE BROKE OFF. THEY WERE UNAWARE THAT THE TIP WAS STILL IN THE PATIENT UNTIL IT SHOWED UP ON X-RAY. IT WAS NOT RETRIEVED. DURING THE SAME PROCEDURE THE FIBERLOOP #2 SUTURE, AR-7234, BROKE OFF. IT IS STILL IN THE PATIENT. THERE IS A REVISION SURGERY SCHEDULED ON (B)(6) 2017 TO REMOVE THE FIBERLOOP AND POSSIBLY THE NEEDLE TIP. FOLLOW-UP INVESTIGATION: DURING THE (B)(6) 2017 PROCEDURE THE TIP OF THE SUREFIRE SCORPION NEEDLE, AR-13991N, LOT 10113253 ((B)(4)) BROKE OFF AND REMAINED IN THE PATIENT. ALSO, THE NEEDLE TIP AND SOME SUTURE BROKE OFF AND REMAINED IN THE PATIENT FROM THE AR-7234, FIBERLOOP, #2 BLUE, LOT 10026082 ((B)(4)). THE REVISION SURGERY TOOK PLACE ON (B)(6) 2017, AT A DIFFERENT FACILITY THEN THE ORIGINAL SURGERY AND WAS PERFORMED BY A DIFFERENT SURGEON THAN THE ORIGINAL PROCEDURE SURGEON. DURING THE REVISION THE SURGEON WAS ABLE TO REMOVE THE (B)(6) 2017 UNRETRIEVED SUTURE AND UNRETRIEVED NEEDLE TIP FROM THE AR-7234 FROM THE PATIENT. THE (B)(6) 2017 UNRETRIEVED NEEDLE TIP FROM THE AR-13991N SUREFIRE SCORPION NEEDLE WAS NOT FOUND AND REMAINS UNRETRIEVED. PATIENT WAS A MALE, (B)(6). DEVICES AND AN

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7178426 ↗

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices →