Arthrex® AR-7250Recalls & reported problems

Arthrex, Inc. · 0 FW 38"BLU 22.2 MM W/TAPER PNT NDL · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00888867041851), received between Dec 21, 2020 and Sep 16, 2026 — classified as 1 injury, 10 malfunctions.

The most frequently reported specific device problem is “Device Damaged by Another Device” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 7 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Damaged by Another Device7
  • Material Frayed3
  • Patient-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • Erythema1
  • Itching Sensation1
  • Pain1
  • Reaction1
  • Swelling1
  • Unspecified Infection1
  • Wound Dehiscence1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2026Material Frayed

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787988 ↗
  • MalfunctionSep 16, 2026Material Frayed

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787979 ↗
  • MalfunctionMar 5, 2026Device Damaged by Another Device

    ON 13-FEB-2026, IT WAS REPORTED BY A DISTRIBUTOR VIA (B)(4) THAT A QTY. 7 OF AR-7250 FIBERWIRE #0, A QTY. 5 OF AR-7200 FIBERWIRE #2, A QTY. 1 OF AR-7535 FIBERLINK SUTURE TAPE WERE BEING CUT BY AN AR-12990N SCORPION NEEDLE DURING A MENISCAL ROOT REPAIR PROCEDURE. ANOTHER AR-12990N NEEDLE WAS OPENED BUT CONTINUED TO CUT THE SUTURES. AN AR-4068-90 SUTURE LASSO WAS THEN USED AS AN ALTERNATIVE TO TRY AND PASS THE SUTURES THROUGH TISSUE BUT WAS ALSO UNSUCCESSFUL. THE CASE WAS COMPLETED SUCCESSFULLY WITH A SHOULDER SCORPION. NO PATIENT HARM WAS REPORTED, AND NO PIECES BROKE OFF INSIDE THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24520388 ↗
  • MalfunctionMar 5, 2026Device Damaged by Another Device

    ON 13-FEB-2026, IT WAS REPORTED BY A DISTRIBUTOR VIA (B)(4) THAT A QTY. 7 OF AR-7250 FIBERWIRE #0, A QTY. 5 OF AR-7200 FIBERWIRE #2, A QTY. 1 OF AR-7535 FIBERLINK SUTURE TAPE WERE BEING CUT BY AN AR-12990N SCORPION NEEDLE DURING A MENISCAL ROOT REPAIR PROCEDURE. ANOTHER AR-12990N NEEDLE WAS OPENED BUT CONTINUED TO CUT THE SUTURES. AN AR-4068-90 SUTURE LASSO WAS THEN USED AS AN ALTERNATIVE TO TRY AND PASS THE SUTURES THROUGH TISSUE BUT WAS ALSO UNSUCCESSFUL. THE CASE WAS COMPLETED SUCCESSFULLY WITH A SHOULDER SCORPION. NO PATIENT HARM WAS REPORTED, AND NO PIECES BROKE OFF INSIDE THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24520382 ↗
  • MalfunctionMar 5, 2026Device Damaged by Another Device

    ON 13-FEB-2026, IT WAS REPORTED BY A DISTRIBUTOR VIA (B)(4) THAT A QTY. 7 OF AR-7250 FIBERWIRE #0, A QTY. 5 OF AR-7200 FIBERWIRE #2, A QTY. 1 OF AR-7535 FIBERLINK SUTURE TAPE WERE BEING CUT BY AN AR-12990N SCORPION NEEDLE DURING A MENISCAL ROOT REPAIR PROCEDURE. ANOTHER AR-12990N NEEDLE WAS OPENED BUT CONTINUED TO CUT THE SUTURES. AN AR-4068-90 SUTURE LASSO WAS THEN USED AS AN ALTERNATIVE TO TRY AND PASS THE SUTURES THROUGH TISSUE BUT WAS ALSO UNSUCCESSFUL. THE CASE WAS COMPLETED SUCCESSFULLY WITH A SHOULDER SCORPION. NO PATIENT HARM WAS REPORTED, AND NO PIECES BROKE OFF INSIDE THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24520326 ↗
  • MalfunctionMar 5, 2026Device Damaged by Another Device

    ON 13-FEB-2026, IT WAS REPORTED BY A DISTRIBUTOR VIA (B)(4) THAT A QTY. 7 OF AR-7250 FIBERWIRE #0, A QTY. 5 OF AR-7200 FIBERWIRE #2, A QTY. 1 OF AR-7535 FIBERLINK SUTURE TAPE WERE BEING CUT BY AN AR-12990N SCORPION NEEDLE DURING A MENISCAL ROOT REPAIR PROCEDURE. ANOTHER AR-12990N NEEDLE WAS OPENED BUT CONTINUED TO CUT THE SUTURES. AN AR-4068-90 SUTURE LASSO WAS THEN USED AS AN ALTERNATIVE TO TRY AND PASS THE SUTURES THROUGH TISSUE BUT WAS ALSO UNSUCCESSFUL. THE CASE WAS COMPLETED SUCCESSFULLY WITH A SHOULDER SCORPION. NO PATIENT HARM WAS REPORTED, AND NO PIECES BROKE OFF INSIDE THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24520279 ↗
  • MalfunctionMar 5, 2026Device Damaged by Another Device

    ON 13-FEB-2026, IT WAS REPORTED BY A DISTRIBUTOR VIA (B)(4) THAT A QTY. 7 OF AR-7250 FIBERWIRE #0, A QTY. 5 OF AR-7200 FIBERWIRE #2, A QTY. 1 OF AR-7535 FIBERLINK SUTURE TAPE WERE BEING CUT BY AN AR-12990N SCORPION NEEDLE DURING A MENISCAL ROOT REPAIR PROCEDURE. ANOTHER AR-12990N NEEDLE WAS OPENED BUT CONTINUED TO CUT THE SUTURES. AN AR-4068-90 SUTURE LASSO WAS THEN USED AS AN ALTERNATIVE TO TRY AND PASS THE SUTURES THROUGH TISSUE BUT WAS ALSO UNSUCCESSFUL. THE CASE WAS COMPLETED SUCCESSFULLY WITH A SHOULDER SCORPION. NO PATIENT HARM WAS REPORTED, AND NO PIECES BROKE OFF INSIDE THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24520074 ↗
  • MalfunctionMar 5, 2026Device Damaged by Another Device

    ON 13-FEB-2026, IT WAS REPORTED BY A DISTRIBUTOR VIA (B)(4) THAT A QTY. 7 OF AR-7250 FIBERWIRE #0, A QTY. 5 OF AR-7200 FIBERWIRE #2, A QTY. 1 OF AR-7535 FIBERLINK SUTURE TAPE WERE BEING CUT BY AN AR-12990N SCORPION NEEDLE DURING A MENISCAL ROOT REPAIR PROCEDURE. ANOTHER AR-12990N NEEDLE WAS OPENED BUT CONTINUED TO CUT THE SUTURES. AN AR-4068-90 SUTURE LASSO WAS THEN USED AS AN ALTERNATIVE TO TRY AND PASS THE SUTURES THROUGH TISSUE BUT WAS ALSO UNSUCCESSFUL. THE CASE WAS COMPLETED SUCCESSFULLY WITH A SHOULDER SCORPION. NO PATIENT HARM WAS REPORTED, AND NO PIECES BROKE OFF INSIDE THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24520073 ↗
  • MalfunctionMar 5, 2026Device Damaged by Another Device

    ON 13-FEB-2026, IT WAS REPORTED BY A DISTRIBUTOR VIA (B)(4) THAT A QTY. 7 OF AR-7250 FIBERWIRE #0, A QTY. 5 OF AR-7200 FIBERWIRE #2, A QTY. 1 OF AR-7535 FIBERLINK SUTURE TAPE WERE BEING CUT BY AN AR-12990N SCORPION NEEDLE DURING A MENISCAL ROOT REPAIR PROCEDURE. ANOTHER AR-12990N NEEDLE WAS OPENED BUT CONTINUED TO CUT THE SUTURES. AN AR-4068-90 SUTURE LASSO WAS THEN USED AS AN ALTERNATIVE TO TRY AND PASS THE SUTURES THROUGH TISSUE BUT WAS ALSO UNSUCCESSFUL. THE CASE WAS COMPLETED SUCCESSFULLY WITH A SHOULDER SCORPION. NO PATIENT HARM WAS REPORTED, AND NO PIECES BROKE OFF INSIDE THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24519896 ↗
  • MalfunctionSep 18, 2025Material Frayed; Material Frayed

    ON (B)(6) 2025, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-7250, FIBERWIRE® #0 BRAIDED POLYBLEND SUTURE, BLUE W/TAPERED NEEDLE, WHILST PERFORMING CERCLAGE WITH 0 FIBERWIRE ON AN ACL GRAFT, THE FIBERWIRE OUTER COATING STARTED TO SHRED. THIS WAS DETECTED DURING USE IN AN ACL RECONSTRUCTION PROCEDURE ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AS THE SURGEON CONTINUED WITH THE FRAYING SUTURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23088554 ↗
  • InjuryDec 21, 2020Patient-Device Incompatibility

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A TRAPEZIUM PROCEDURE (B)(6) 2020. DURING THE PROCEDURE THE SURGEON USED ARTHREX, AR-7250 FIBERWIRE SUTURES. SINCE THE PROCEDURE THE INCISION OPENED UP AND THE PATIENT DEVELOPED AN INFECTION WHICH REQUIRED TREATMENT. THE PATIENT HAS BEEN HAVING ONGOING ISSUES SINCE SURGERY WHICH INCLUDE PAIN, ITCHING, REDNESS AND SWELLING. PATIENT HAS HAD ISSUES IN PAST SURGERIES WITH SUTURES/ SOFT PLASTICS. THE CURRENT ISSUES ARE SIMILAR TO THOSE EXPERIENCED IN THE PAST.

    Patient problems: Erythema; Itching Sensation; Pain; Reaction; Swelling; Unspecified Infection; Wound Dehiscence

    MAUDE report 11045098 ↗

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices →