Skip to content

Arthrex® AR-8943-16Recalls & reported problems

Arthrex, Inc. · 2.0mm Drill Bit, Calibrated · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867052260), received between Sep 7, 2021 and May 5, 2026 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Entrapment of Device1

Patient problems

  • Foreign Body In Patient3
  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 5, 2026Break

    IT WAS REPORTED THAT DURING A REPAIR OF A FRACTURE OF THE FOOT¿ANKLE COMPLEX THE DRILL BIT BROKE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25083183 ↗
  • MalfunctionOct 8, 2024Break; Break

    NO REVISION IS REQUIRED. THERE WAS NO PROCEDURE DELAY, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED.

    Patient problems: Foreign Body In Patient

    MAUDE report 20397518 ↗
  • InjuryJul 3, 2023Break; Break

    ON 6/5/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-8943-16 DRILL BIT BROKE DURING FIBULAR DRILLING, AND THE FRAGMENTS WERE LEFT INSIDE THE PATIENT. THIS WAS DISCOVERED DURING AN ANKLE FRACTURE PROCEDURE ON (B)(6) 2023.

    Patient problems: Foreign Body In Patient

    MAUDE report 17247211 ↗
  • MalfunctionMay 4, 2023Break; Break

    ON (B)(6) 2023, IT WAS REPORTED BY A ARTHREX EMPLOYEE VIA SEMS THAT AN AR-8943-16 DRILL BIT BROKE DURING SURGERY. THIS WAS DISCOVERED DURING A CASE WITH NO PATIENT EFFECT, DEVICE BREAKS DURING USE. MATERIAL BROKE DURING THE PROCESS. THE PATIENT HAD NO MAJOR PROBLEMS AND THE DOCTOR WAS ABLE TO FINISH THE PROCESS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16870647 ↗
  • MalfunctionJan 23, 2023Break; Entrapment of Device

    ON (B)(6) 2023 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-8943-16 DRILL BIT BROKE DURING A FIBULA ORIF PROCEDURE. THE DOCTOR WAS DRILLING THE FAR CORTEX OF THE FIBULA WITH THE 2MM DRILL WHEN THE DRILL BROKE AND BECAME LODGED IN THE FIBULAR INTER MEDULLARY CANAL. THE DRILL BIT PIECE WAS UNABLE TO BE RETRIEVED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 16223195 ↗
  • MalfunctionSep 7, 2021Break

    ON 8/19/2021 IT WAS REPORTED BY ARTHREX EMPLOYEE VIA SEMS THAT AN AR-8943-16 BROKE WHEN MAKING A HOLE WHILE USING THE FIBULA PLATE. THIS OCCURRED DURING USE ON (B)(6) 2021. THE BROKEN PIECE WAS RECOVERED FROM THE PATIENT'S BODY AND THE PROCEDURE WAS COMPLETED USING ANOTHER DRILL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12435136 ↗

Other BIT, DRILL devices

All BIT, DRILL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page