This Arthrex FiberTak suture anchor is a small implant used to fasten soft tissue to bone during surgery. The 30 reports mostly describe the anchor breaking, coming out of the bone or out of the drilled hole during a procedure, sticking in the aiming device, or failing to fire, with a few mentioning leftover material in the patient, inflammation and pain. Four reports involve injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
33
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 00888867101890), received between May 17, 2018 and Sep 17, 2026 — classified as 4 injuries, 29 malfunctions.

The most frequently reported specific device problem is “Break” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 6 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 5 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break12
  • Device Dislodged or Dislocated5
  • Failure to Fire5
  • Physical Resistance/Sticking5
  • Positioning Failure3
  • Adverse Event Without Identified Device or Use Problem2
  • Fitting Problem2
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions20
  • Insufficient Information7
  • Foreign Body In Patient3
  • Inflammation2
  • Pain2
  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50805592 ↗
  • MalfunctionSep 17, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50805252 ↗
  • MalfunctionSep 2, 2026Detachment of Device or Device Component

    Patient problems: Insufficient Information

    MAUDE report 50563331 ↗
  • MalfunctionAug 25, 2026Failure to Fire

    Patient problems: Insufficient Information

    MAUDE report 50430805 ↗
  • MalfunctionAug 25, 2026Failure to Fire

    Patient problems: Insufficient Information

    MAUDE report 50430667 ↗
  • MalfunctionAug 25, 2026Failure to Fire

    Patient problems: Insufficient Information

    MAUDE report 50430338 ↗
  • MalfunctionAug 25, 2026Failure to Fire

    Patient problems: Insufficient Information

    MAUDE report 50430017 ↗
  • MalfunctionAug 25, 2026Failure to Fire

    Patient problems: Insufficient Information

    MAUDE report 50429345 ↗
  • MalfunctionFeb 23, 2026Device Dislodged or Dislocated

    ON 05-FEB-2026, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT AN AR-3600 FIBERTAK SUTURE 1.6 MM ANCHOR PULLED OUT WHILE CHECKING FIXATION. THE PROCEDURE WAS COMPLETED USING ANOTHER AR-3600 FIBERTAK SUTURE 1.6 MM ANCHOR. NO SIGNIFICANT SURGICAL DELAY WAS REPORTED, AND NOTHING REMAINED IN THE PATIENT. NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. NO ADVERSE PATIENT EFFECTS WERE REPORTED DURING OR FOLLOWING THE PROCEDURE RELATED TO THIS ISSUE. THE ISSUE WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24416545 ↗
  • MalfunctionDec 8, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON 11/19/2025, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA (B)(4) THAT AN AR-3600 FIBERTAK ANCHOR CAME OUT OF THE HOLE DRILLED FOR ITS INSERTION. THIS WAS DISCOVERED DURING A PROCEDURE WITH NO REPORTED PATIENT HARM. THE CASE WAS COMPLETED USING A SUTURETAK ANCHOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23743918 ↗
  • MalfunctionOct 30, 2025Break; Break

    IT WAS REPORTED THAT DURING A BANKART REPAIR SURGERY THE AR-3600 DID NOT FIT THROUGH AIMER AR-3610CTC, AS THE RED STABILISER PEG GOT STUCK. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE AVOE 27-OCT-2025: IT WAS CONFIRMED THAT WHEN THE SURGEON USED THE DEVICE AR-3600 IN THE AR-3610CTC SPEAR, THE STABILIZATION PEG OF THE AR-3600 GOT STUCK AND BROKE INSIDE THE SPEAR WHEN THE DEVICE WAS REMOVED WITH FORCE FROM THE SPEAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23427326 ↗
  • MalfunctionOct 30, 2025Break; Break

    IT WAS REPORTED THAT DURING A BANKART REPAIR SURGERY THE AR-3600 DID NOT FIT THROUGH AIMER AR-3610CTC, AS THE RED STABILISER PEG GOT STUCK. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE AVOE 27-OCT-2025 IT WAS CONFIRMED THAT WHEN THE SURGEON USED THE DEVICE AR-3600 IN THE AR-3610CTC SPEAR, THE STABILIZATION PEG OF THE AR-3600 GOT STUCK AND BROKE INSIDE THE SPEAR WHEN THE DEVICE WAS REMOVED WITH FORCE FROM THE SPEAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23427248 ↗
  • MalfunctionOct 30, 2025Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING A BANKART REPAIR SURGERY THE AR-3600 DID NOT FIT THROUGH AIMER AR-3610CTC, AS THE RED STABILISER PEG GOT STUCK. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE AVOE 27-OCT-2025. IT WAS CONFIRMED THAT WHEN THE SURGEON USED THE DEVICE AR-3600 IN THE AR-3610CTC SPEAR, THE STABILIZATION PEG OF THE AR-3600 GOT STUCK AND BROKE INSIDE THE SPEAR WHEN THE DEVICE WAS REMOVED WITH FORCE FROM THE SPEAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23427107 ↗
  • MalfunctionOct 30, 2025Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING A BANKART REPAIR SURGERY THE AR-3600 DID NOT FIT THROUGH AIMER AR-3610CTC, AS THE RED STABILISER PEG GOT STUCK. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE (B)(6) 27-OCT-2025: IT WAS CONFIRMED THAT WHEN THE SURGEON USED THE DEVICE AR-3600 IN THE AR-3610CTC SPEAR, THE STABILIZATION PEG OF THE AR-3600 GOT STUCK AND BROKE INSIDE THE SPEAR WHEN THE DEVICE WAS REMOVED WITH FORCE FROM THE SPEAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23426949 ↗
  • MalfunctionOct 30, 2025Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING A BANKART REPAIR SURGERY THE AR-3600 DID NOT FIT THROUGH AIMER AR-3610CTC, AS THE RED STABILISER PEG GOT STUCK. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE AVOE 27-OCT-2025: IT WAS CONFIRMED THAT WHEN THE SURGEON USED THE DEVICE AR-3600 IN THE AR-3610CTC SPEAR, THE STABILIZATION PEG OF THE AR-3600 GOT STUCK AND BROKE INSIDE THE SPEAR WHEN THE DEVICE WAS REMOVED WITH FORCE FROM THE SPEAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23426767 ↗
  • MalfunctionOct 30, 2025Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING A BANKART REPAIR SURGERY THE AR-3600 DID NOT FIT THROUGH AIMER AR-3610CTC, AS THE RED STABILISER PEG GOT STUCK. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE AVOE 27-OCT-2025 IT WAS CONFIRMED THAT WHEN THE SURGEON USED THE DEVICE AR-3600 IN THE AR-3610CTC SPEAR, THE STABILIZATION PEG OF THE AR-3600 GOT STUCK AND BROKE INSIDE THE SPEAR WHEN THE DEVICE WAS REMOVED WITH FORCE FROM THE SPEAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23426668 ↗
  • MalfunctionSep 18, 2025Device Dislodged or Dislocated; Material Split, Cut or Torn; Device Dislodged or Dislocated; Material Split, Cut or Torn

    ON 08/30/2025, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT AN AR-3600 FIBERTAK SUTURE ANCHOR EXPERIENCED AN ISSUE. AFTER THE ANCHOR WAS PLACED FOLLOWING THE CORRECT STEPS, A GENTLE PULL WAS PERFORMED TO VERIFY PLACEMENT. AS A RESULT, THE ANCHOR CAME OUT. UPON INSPECTION, IT WAS FOUND TO BE TORN (SEE PHOTO ATTACHMENT). THE CASE WAS COMPLETED; HOWEVER, NO SPECIFIC DETAILS WERE PROVIDED REGARDING HOW IT WAS FINALIZED. THIS ISSUE WAS DISCOVERED DURING A SHOULDER ARTHROSCOPY PROCEDURE PERFORMED ON (B)(6)2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED ON 09/08/2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23086108 ↗
  • MalfunctionSep 2, 2025Positioning Failure; Positioning Failure

    ON 08/13/2025, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT AN AR-3600 FIBERTAK SUTURE ANCHOR FAILED TO SECURE UPON INSERTION. A SECOND ATTEMPT WAS MADE, BUT THE TIP OF THE ANCHOR WAS DAMAGED AND COULD NOT BE USED, WHICH LED TO ITS REMOVAL FROM THE PATIENT. THE CASE WAS COMPLETED USING A REPLACEMENT AR-3600 FIBERTAK SUTURE ANCHOR. THERE WAS NO SIGNIFICANT DELAY IN THE PROCEDURE, AND NO ADDITIONAL ANESTHESIA WAS ADMINISTERED. NO ADVERSE EFFECTS WERE REPORTED DURING OR FOLLOWING THE PROCEDURE. THE ISSUE WAS DISCOVERED DURING A PROCEDURE PERFORMED ON (B)(6) 2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22945933 ↗
  • MalfunctionMar 4, 2025Material Frayed; Material Frayed

    ON 02/17/2025, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT AN AR-3600 FIBERTAK SUTURE ANCHOR HAD A FRAYED SUTURE THREAD AFTER IT WAS PLACED IN THE BONE HOLE. THE SURGEON WAS CONCERNED ABOUT THE FIXATION AND REMOVED THE ANCHOR. A NEW BONE HOLE WAS DRILLED, AND A NEW AR-3600 FIBERTAK SUTURE ANCHOR WAS USED TO COMPLETE THE SURGERY. THIS WAS DISCOVERED DURING A SHOULDER DISLOCATION REPAIR PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21517597 ↗
  • MalfunctionFeb 12, 2025Fitting Problem; Physical Resistance/Sticking; Fitting Problem; Physical Resistance/Sticking

    ON 01/27/2025, IT WAS REPORTED BY A ARTHREX SUBSIDIARY EMPLOYEE VIA (B)(6) THAT AN AR-3610CTC-2 FIBNERTAK SPEAR AND AN AR-3600 FIBERTAK ANCHOR BECAME STUCK IN THE DRILL GUIDE WHEN THE SURGEON WENT TO INSERT. THIS OCCURRED DURING USE IN AN ARCR CASE WITH NO PATIENT HARM. THIS CASE WAS COMPLETED BY USING ANOTHER PRODUCT OF SAME PRODUCT NUMBER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21363336 ↗
  • MalfunctionOct 10, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT DURING A LABRUM STABILIZATION SURGERY THE ANCHORS AR-3600 (LOT: 15232907 & 15246457) DID NOT FIT THROUGH THE DRILL SLEEVE AR-3610CTC. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20415875 ↗
  • MalfunctionOct 10, 2024Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT DURING A LABRUM STABILIZATION SURGERY THE ANCHORS AR-3600 (LOT: 15232907 & 15246457) DID NOT FIT THROUGH THE DRILL SLEEVE AR-3610CTC. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20415806 ↗
  • MalfunctionOct 1, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON 09/18/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3600 FIBERTAK¿ SUTURE ANCHOR FAILED AND PULLED OUT. THIS OCCURRED DURING A SHOULDER SCOPE/LABRAL REPAIR WHEN THE FIBERTAK ANCHOR FAILED AND PULLED OUT. THERE WERE NO ADVERSE EFFECTS ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20343852 ↗
  • MalfunctionJun 2, 2024Fitting Problem; Fitting Problem

    IT WAS REPORTED THAT DURING A SHOULDER STABILIZATION SURGERY THE ANCHOR DID NOT FIT THROUGH THE ASSOCIATED SPEAR. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. ***UPDATE SWIT 24-AUG-2022: A SPEAR WITH PART NUMBER AR-2948CT, LOT. 10122774 WAS USED. THE FOLLOWING ANCHORS DID FIT THROUGH THE SPEAR. THE NURSE SAID THAT ACCORDING TO HER FEELING, THE ANCHOR IS LOADED DIFFERENTLY. THAT IS, THE ANCHOR ITSELF IS PROBABLY A BIT CROOKED ON THE INSERTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19444622 ↗
  • MalfunctionApr 29, 2024Use of Device Problem; Use of Device Problem

    ON 03/29/2022, IT WAS REPORTED BY A DISTRIBUTOR VIA EMAIL THAT AN AR-3600 FIBERTAK SUTURES WERE LOOSEN AFTER INSERTING IT. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2022. ADDITIONAL INFORMATION REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 19202667 ↗

Other FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices →